Testing Blood Glucose

When you insert the test strip, the meter will automatically turn on.
Insert the test strip into the test strip port, and the meter will turn on
1.
automatically. If the audio is on, you will hear a "beep" sound, and
all of the icons on the display will turn on at the same time. If the strip
is inserted upside down, the meter will not turn on.
The test strip has been inserted correctly when you see the "Apply
2.
Blood" icon and drop icon flashing. If the test strip is not correctly
inserted, the meter will not turn on – repeat the steps above.
Apply blood to the sample area of the test strip. If the audio is on, the
3.
meter will beep, telling you that enough blood has been applied and
the meter will start to measure.
DO NOT:
Apply blood to the front or back of the test strip.
Smear the blood drop onto the test strip.
Press your finger against the test strip.
20
The display will count down 5 seconds during
4.
the measurement process. If the audio is on, the
test will end with a beep. If you applied a blood
sample but the meter does not begin a
countdown, you may reapply a second drop of
blood within 3 seconds.
If you press the ">" button, the test results will be marked with the "#"
icon. Results marked with "#" symbol will not be used in the 7, 14, 30,
60 or 90 day average calculations. If you accidentally mark your result
with the "#" icon, simply press the ">" button again to cancel the mark.
After marking a result with "#", please test again with a new strip.
If the before/after meal marker setting is on, you can press the "<"
button to tag the reading before or after meal, as shown in the figure:
Press the "M" button to confirm.
If an error code is displayed, please check the troubleshoot-
ing instructions (page 32). If "HI" or "LO" symbol is displayed, refer
to the "HI" and "LO" Messages section (page 25).