"HI" and "LO" Messages

Your meter can accurately measure blood glucose concentrations
between 1.1-33.3 mmol/L (20-600 mg/dL). "HI" and "LO" messages
indicate that the test measurement was outside of this range.
If "HI" is displayed, it indicates the test result is above 33.3 mmol/L (600
mg/dL). You should test again to ensure that no mistake was made in the
procedure. If you are sure your glucose meter is functioning properly, no
mistakes were made, and the test results still show "HI", then you may be
in a state of severe hyperglycemia, please contact your healthcare
professional immediately.
If "LO" is displayed, it indicates the test result is below 1.1 mmol/L (20
mg/dL). You should test again to ensure that no mistake was made in the
procedure. If you are sure your glucose meter is functioning properly, no
mistakes were made, and the test results still show "LO", then you may be
in a state of severe hypoglycemia, please contact your healthcare
professional immediately.
"

"HYPO" and "HYPER" Messages

If "HYPO" is displayed, it indicates the test result is below the
hypoglycemia limit that you entered in the settings.
If "HYPER" is displayed, it indicates the test result is above the
hyperglycemia limit that you entered in the settings.
24
Blood Ketone Test Reminder: if "KETONE?" is
displayed, this indicates a blood ketone test is
necessary because your blood glucose test results
are higher than 16.7mmol/L (300 mg/dL).
Please contact your healthcare provider.
Notes:
This meter sytem is only for in vitro diagnostic use, and can only be
used with
Blood Glucose Test Strips. Use of other test
strips can lead to wrong test results.
This meter can only be used to determine blood glucose levels with
whole blood samples. Do not use serum or plasma samples.
This meter can be used within a hematocrit (HCT) range of 10% ~
70%. Please do not use this product for testing if your hematocrit
value is out of this range.
Ascorbic acid >3 mg/dL and Xylose >9 mg/dL will cause over
estimation of blood glucose results.
Fatty substances, such as triglycerides less than 3,000 mg/dL or
cholesterol less than 500 mg/dL, have no major effect on blood
glucose test results.
Patients who are severely ill, suffer from severe dehydration, or are in
a hyperosmolar state (with or without ketosis) should not use the
Blood Glucose Monitoring system.
This product is only suitable for clinical screening tests or family
self-monitoring. The test results can't be referred to as confirmed
cases. In order to ensure the accuracy of the results, the test results
can be further confirmed by other methods, such as biochemical
methods.
As with all diagnostic reagents, test results must be linked with a
professional doctor's diagnosis from the other clinical symptoms.
Carefully process waste caused by blood glucose tests according to
the relevant local laws and regulations, because blood samples are
considered a biohazard.

Meter Memory

The glucose meter can store up to 500 test results with the corresponding
time and date. If 500 test results already exist, a new result will overwrite
the oldest one. The glucose meter is also capable of calculating 7, 14, 30,
60 and 90 day averages from the stored results.