Replacing The Poc1-Connect - Invacare Platinum Mobile POC1-100 Series Service Manual

Oxygen concentrator
Table of Contents

Advertisement

Invacare® Platinum® Mobile
5. Slide the new pull ring into the mount.
6. Continue sliding the new pull ring into the mount until
it is fully engaged.
7. Ensure pull ring is seated correctly.
22

4.7 Replacing the POC1-CONNECT

The POC1-CONNECT is not available in all regions.
The POC1-CONNECT dongle is only compatible with
Invacare Platinum Mobile oxygen concentrators
beginning with serial number 17JF029716.
1. Remove the Platinum Mobile oxygen concentrator from
the carry bag. Refer to the user manual.
Manuals are available online at
www.invacare.com or call 800-333-6900.
2. Remove the existing POC1-CONNECT dongle A and
insert the new one in its place.
3. Register the new POC1-CONNECT dongle. Visit
www.invacare.com/BreatheSmartBreatheEasy for
instructions on how to register the new POC1-CONNECT
and Platinum Mobile oxygen concentrator.
1187375-E