SETUP AND CALIBRATION, CONT.
F= (FILTER LEVEL)
Press the BMI / ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI / ENTER key again. Otherwise, press the or keys to scroll through and
select a new value and then press the BMI / ENTER key to save it and proceed to the next
prompt. Allowable values are 1 (least amount of filtering) to 99 (greatest amount of filtering).
BR= (BREAK RANGE)
Press the BMI / ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI / ENTER key again. Otherwise, press the or keys to scroll through and
select a new value and then press the BMI / ENTER key to save it and proceed to the next
prompt. Allowable values are 1 to 99 which correspond to the number of division changes
to break out of filtering.
SR= (SAMPLE RATE)
Press the BMI / ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI / ENTER key again. Otherwise, press the or keys to scroll through and
select a new value and then press the BMI / ENTER key to save it and proceed to the next
prompt. Allowable values are a minimum of 1 sample per second to a maximum of 16
samples per second in one sample per second intervals.
CAL= (CALIBRATION)
With the display showing CAL=, press the BMI / ENTER key. The display will change to
show the current setting 0 (0=NO). If the scale has been previously calibrated and you wish
to skip calibration and proceed to the HOLD= (Hold Mode) prompt, press the BMI / ENTER
key and the previous calibration will be retained.
To begin calibration, press the key to select 1 (yes) and then press the BMI / ENTER key.
After pressing the BMI / ENTER key, the display will change to LOAd=.
LoAd= (LOAD CALIBRATION WEIGHT)
With the display showing LoAd= perform the following steps:
1. Make certain the scale platform is empty and free of debris.
2. Place the desired amount of calibrated test weights on the scale platform. A minimum
of 50% of scale's capacity is required. However 70% to 100% is recommended.
3. Press the BMI / ENTER key.
4. If the value displayed is acceptable, press the BMI / ENTER key again. Otherwise,
determine the exact amount of test weight placed on the scale platform and then using
the or keys and the or keys scroll through and select the test weight amount.
5. Verify that the numbers selected are the same as the amount of the test weight and
then press the BMI / ENTER key.
6. Starting at the left and preceding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the display
will proceed to the next prompt.
8555-M260-O1 Rev J • 750EU Owner's
13