Setup And Calibration - Detecto 750EU Owner's Manual

Weight indicator with body mass index features
Table of Contents

Advertisement

Your Model 750EU indicator has been thoroughly tested and calibrated before being shipped
to you. If you received the indicator with a scale, calibration is not necessary. If the indicator
is being connected to a scale for the first time or recalibration is necessary for other reasons,
proceed as indicated.
Calibration of the indicator is accomplished entirely by the keypad. To enter setup and
calibration, the calibration switch must be pushed in and held while turning on the indicator.
Begin Setup and Calibration:
1. With power off, remove the 2 screws from the left
end cap (as viewed from indicator front) and
remove the end cap.
2. Locate the calibration switch on the circuit board
and push it in. See Figure No. 8.
3. With the switch held in, press the ON / OFF key.
4. When display shows SETUP, release the
calibration switch. The indicator is now ready for

setup and calibration.

During setup and calibration it is necessary to enter
values using the indicator's keypad. When a prompt
is displayed on the indicator, press the BMI / ENTER
key to view the current setting. To retain the current
setting and proceed to the next prompt, press the
BMI / ENTER key again. To change a setting, press
the  or  keys (up or down arrow) to scroll through
and select a new value. After a new value has been
selected, press the BMI / ENTER key to save it and
advance to the next prompt. Note that some setup
prompts have values with 2 or more digits. The
blinking character is the position to be changed and
can be advanced to the next position by pressing the
 or  (left or right arrow) key.
DO NOT operate the keypad with pointed objects (pencils, pens, etc).
Damage to keypad resulting from this practice is NOT covered under warranty.
USA= (DOMESTIC or INTERNATIONAL)
This is the prompt to select whether the indicator is used in the USA (domestic) or outside
the USA (international). Press the BMI / ENTER key to view the current setting. If the value
displayed is acceptable, press the BMI / ENTER key again. Otherwise, press the  or 
keys to scroll through and select a new value and then press the BMI / ENTER key to save
it and proceed to the next prompt. Allowable values are: 0 or 1.
USA=1 (Domestic)
No Zero Limit
CAP +4% to OC
int= (INTERVAL SETTING)
Press the BMI / ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI / ENTER key again. Otherwise, press the  or  keys to scroll through and
select a new value and then press the BMI / ENTER key to save it and proceed to the next
prompt. Allowable values are: 1, 2 or 5.
SETUP AND CALIBRATION
USA=0 (International)
+/- 4% Zero Limit
CAP +9 grads to OC
8555-M260-O1 Rev J • 750EU Owner's
Calibration
Switch
Figure No. 8