Linde HiQ R200 Instructions For Use Manual page 19

4-/ 6-port line regulator
Table of Contents

Advertisement

Available languages

Available languages

Description
Reason
Increase of
Leakage in
outlet pressure
pressure regu-
& relief valve
lator seat
opens
Slight increase
Inlet pressure
of outlet pres-
drop
sure & relief
valve does not
open
Slight increase
Difference
of outlet pres-
between flow
sure
pressure (with
gas flow) and
Zero Pressure
(without gas
flow)
Slight drop of
Difference
outlet pressure
between flow
pressure (with
gas flow) and
Zero Pressure
(without gas
flow)
Heavy drop of
Required
outlet pressure
flow to high
for pressure
regulator
Heavy drop of
Leakage in
outlet pressure
process gas
tubing
Outlet Pres-
Wrong set-
sure to high
tings for pres-
sure regulator
Outlet pres-
Pollution or
sure increase
damage of
immediately
pressure
all the way to
regulator seat
the stop or
"Creap"
outlet pressure
increases and
gets slowly
higher than set
value and relief
valve opens
Noise during
Required
operation
flow too high
for pressure
regulator
Solution
Valve
Repair by
manufacturer
No failure, nor-
mal operating
state
No failure, nor-
mal operating
state
No failure, nor-
mal operating
state
Choose new
pressure
regulator for re-
quired pressure
and flow
Check process
gas tubing
Adjust pressure
regulator set-
tings, readjust
pressure
regulator, main-
tenance
Repair by
manufacturer
Choose new
pressure
regulator for re-
quired pressure
and flow
19/44
Cylinder
Line
Gas
Regulator
Regulator
Supply Panel
EN