7. Error Messages and Troubleshooting

In case of any of the below problems occur during measurement, first check that no
other electrical device is within 12 inches (30 cm). If the problem persists, please
refer to the table below.
Display/
Problem
appears or the
wrist cuff does
not inflate.
appears
appears
Possible Cause
The wrist cuff is not
applied correctly.
Air is leaking from the
wrist cuff.
The wrist cuff is
overinflated exceeding
300 mmHg.
You move or talk during a
measurement. Vibrations
disrupt a measurement.
Apply the wrist cuff correctly, then take
another measurement. Refer to
section 2.4.
Contact customer service. Refer to
chapter 11.
Do not touch the wrist cuff while taking
a measurement.
Remain still and do not talk during a
measurement.
32
Solution