11. Data sheet

Equipment specification
Power Supply
Mains Supply
Fuse at mains connector
At transformer secondary coils
Burning Mode
Burning DC Voltages (No load)
Short Circuit Currents
Measuring
High Voltage Pulse
High Voltage Pulse Energy
Max. Pulse Energy
Number of High Voltage Pulses
Tracing Mode
Trace Frequency
Tracing Voltages (No load)
Short Circuit Currents
Measuring
Volume control
Other
Overheat protection
Safety in Burning Mode
High Voltage Unit should be ON and Safety Button
should be pressed to proceed the burning process.
Mechanical
Cabinet dimensions
Weight
IP Class
Approvals
Environment
28 | ©Danfoss | FEC | 2022.07
DEVItector™ II V2
High Voltage Unit (HVU)
Class I
Installation category II
Pollution degree 2
220–240 V~, 50–60 Hz, 1000 VA
Voltage fluctuations may not exceed ±10% of nominal value
4TAL 250V EN/IEC 60127-2
5A, 2×2A, 0.5A
41, 82, 123, 164, 205, 246, 287, 328, 369 V⎓
Limited by Software to 3 A⎓
Readout on 2x20 character Vacuum Fluorescent Display:
Probe Voltage: 0–400 V⎓
Probe Current: 0–3.00 A⎓
Probe High Voltage Pulse: 0–10 kV
10 kV
0,38 J
70 J
1–3 pulses/sec.
1,8 kHz
5, 10, 15, 20, 25, 30, 35, 40 V RMS
80, 160, 240, 320, 400, 480, 560, 640 mA RMS
Readout on 2x20 character Vacuum Fluorescent Display:
0–1.00 A RMS
Adjustment of sound in headphones
Software controlled cut-off at 3 A or at 75 °C.
Only operable with Safety Key
250 x 260 x 380 mm
17 kg
IP20
The equipment has been tested by Intertek according to
EN/IEC 61010-1 with amendments, EN/IEC 61010-2-030
with amendments and EN/IEC 61326-1 with amendments.
For burner and tracer: indoor and outdoor use.
Temperature range: 5 ºC to 40 ºC.
Altitude: up to 2000 m above sea level.
Humidity at 31 ºC: max. 80% linear falling to 50% at 40 ºC.
BC421332622232en-010101