Presys T-500PIR Technical Manual page 12

Calibrator for infrared thermometer
Table of Contents

Advertisement

PRESYS
 Instruments
BZ: Menu that configures the piezoelectric buzzer.
NO: Disables the buzzer.
YES: The buzzer beeps when the cavity reaches the set point and stabilizes.
ENDCAL: The buzzer beeps only at the end of a calibration in the automatic operating
mode
TU: Menu that configures the PID control parameters for the heating side. More details
in section 6 - PID Control Parameters.
K: Proportional gain
I: Integral gain
D: Derivative gain
FACT: Restore the control parameters to the factory values.
PRG: Menu that programs the calibrator.
DEC_IN: Selects the number of decimals of the thermo-element reading. The default
value depends on the input signal.
DEC_PRB: Selects the number of decimals of the cavity temperature and
point value. The default number is 1.
SET POINT: Enables the calibrator Programmable Operating Mode, and allows the
configuration of programmed values. The selected program is indicated by the
selection arrow. Choose one of the 6 temperature programs or NO to disable this
Programmable Mode.
Select any of the 6 programs and confirm with the key ENTER. Following, it is
shown the menu of temperature set point configuration.
NO
YES
K
I
DEC_IN
SET POINT
DEFAULT
0
1
2
0
1
ENDCAL
D
FACT
DEC_PRB
3
4