W&H proxeo ULTRA PB-5 L Instructions For Use Manual page 11

Handpiece
Table of Contents

Advertisement

Tips
> Only use tips that have been approved by W&H and the associated tip changers or spanners.
> An overview of the correct power settings is included with every tip.
> With periodontal tips, the medical device is suitable for the removal of concretions in the subgingival
region, but not for applications which demand sterile conditions. Choose the lower performance range
when carrying out periodontal treatments on hypersensitive patients in order to guarantee optimum
pain-free treatment.
> Ensure that the original shape of the tips is not affected (e.g. by being dropped).
> The tips must not be bent back into shape or resharpened.
> Locate and secure the tip only with the medical device switched off.
> Never touch the tips when vibrating.
> Insert the tip changer onto the inserted tip of the stationary medical device after every treatment
(protection against injury and infection, tip protection). Tips that are changed using a spanner must be
removed from the medical device immediately after treatment.
> Do not touch into the tip changer (with tip inserted).
> Check for the effect of wear on the tips using the accompanying tip card.
> Change tips if there are visible signs of wear.