ABLE OF ONTENTS Symbols ................................1 Model Description .............................. 2 Features ..................................2 Intended Use ................................2 Warranty ..................................2 Caution and Warning Statements ........................3 Initial Setup ................................. 4 Operation ................................4 Quick Start ................................5 Settings ................................6 Quick Adjust Time and Speed ............................
Symbols Symbol Definition Caution to safety hazard. Caution Potential risk of personal injury or damage to the instrument if improperly handled. Consult the manual before proceeding. Manufacturer Manufacturer of record. Electrical and Recycle only as electronic waste. Do not dispose in normal waste. electronic products recycling symbol RoHS Compliant...
ODEL ESCRIPTION DASH is a versatile line of centrifuges designed for STAT sample processing. Minimize lab error, reduce turnaround time, and streamline training with the simple controls, powerful performance, and unique LED lid lighting system. As a general purpose laboratory centrifuge, it is designed to also run other approved containers filled with chemicals (non-flammable, non-explosive, non-volatile, and non-highly reactive only), environmental samples, and other non-human body samples.
AUTION AND ARNING TATEMENTS This device is intended to be operated by properly trained personnel who have carefully read the operating manual and are familiar with the function of the device. [Refer to the clinical laboratory method specified by the specimen receptacle manufacturer or established by the medical technology for the products applications.] For the safety of both the operator and service personnel, care should be taken when using this centrifuge if handling substances that are known to be toxic, radioactive or contaminated with pathogenic...
NITIAL ETUP o Unpack and verify that all the following are included: Centrifuge Power cord Tube holders Quick Start Guide o Setup the centrifuge on flat and level surface. A bench top clearance height of 21” (54 cm) is required to open the lid.
UICK TART The top screen display alternates between the name of the currently selected speed. The bottom screen displays the time setting. Begins running the cycle displayed on the screen. The lid must be Start closed. Allows access into the rotor chamber by engaging the unlocking Unlock mechanism.
ETTINGS UICK DJUST IME AND PEED Change time, speed (RPM) or g-force (RCF) for a single cycle. To change the speed (RPM) shown on the top display, use the up and down buttons next to that screen. The CYCLE number is Setting Speed replaced with a “--“in the display, and the top screen displays the speed.
HANGING THE UDIBLE EEPER Enter the Press the GEAR button to enter the advanced menu. Advanced Menu While in the advanced menu, navigate to “Beeper”. Switch ON or Turn Beeper OFF with the UP and DOWN buttons next to the TIME display. This On or Off setting will apply to all cycles.
ODIFYING A RESET Save up to 10 custom cycles. The top screen alternates between cycle name and speed. Select Preset Press the CYCLE button to select the Preset Setting you would like Setting to modify Hold the CYCLE button until you hear a double beep. The cycle Access the Menu number should begin flashing.
ELETING A YCLE Enter the With the desired cycle selected, access the menu and enter the Advanced Menu advanced menu. Using the UP and DOWN buttons, navigate to DELETE. Exit the Navigate to Delete menu. WARNING: CYCLE WILL BE DELETED IF MENU IS EXITED WITH DELETE SELECTED Confirm Deletion Press the CYCLE button to Delete the cycle...
RESET YCLES Mem-1 Mem-2 Mem-3 4,300 4,300 3,500 5 min 6 min 7 min Time Brake ALANCING OADS Your centrifuge must contain a balanced load to work properly. Spinning balanced loads will extend the life of the centrifuge and produce better results. Use the following rules when loading the rotor. f an odd number of samples is to be spun, fill a tube with water to match the weight of the unpaired sample and place it across from this sample.
ARE AND REVENTATIVE AINTENANCE With proper care and maintenance, your centrifuge will provide years of laboratory service. For proper care, the following steps should be taken: Always Spin Balanced Loads: Make certain that you are always spinning a balanced load, as shown in the previous section.
ROUBLESHOOTING NOTE: The latch must be turned completely clockwise to its stop position for the centrifuge to operate. The centrifuge does not run o Verify that the centrifuge is powered. One of the LED lights should be on. o If “Lid not closed” message is displayed, make sure the lid latch is turned completely clockwise to its stop position.
Page 15
o If cycle selection is locked on one cycle, press the UNLOCK Cycle parameters cannot be changed button for 5 seconds. Then, press the GEAR button and follow the instructions elsewhere in this manual. o If different saved cycles can be selected but not modified, the centrifuge is in Preset Lock mode.
ENERAL PECIFICATIONS The rotor and accessories are rated for the maximum rotation frequency shown below. Tube Capacity 24 tubes – 3 to 10 mL 12 tubes – 3 to 15 mL 6 tubes – 50 mL Dimensions (H x W x D) 9 in x 14.5 in x 17 in (23 cm x 37 cm x 43 cm) Weight...
02-004-0-0012K 4pl Carrier Assembly, 30% CF 02-006-1-0066 Lid LED PCBA Assembly, Red 03-0-0003-0817 Dash Flex 24 Front Panel Label, Boekel 00-100-100-010 12/24 Series Soft Button & Spacer Replacement Kit 00-100-100-005 Replacement Grommet & Bushing Kit Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 15...
Page 18
Protected by U.S. Patents #6,811,531, # 7,422,554, #D718,463, & #D734,489. Other Patents Pending To access previous manuals, please click the relevant link below: DASH Flex 24 Operator’s Manual, Rev B FDA LISTED INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION This product must not be disposed of with other waste.