RapidFISH
Models 240200 and 240200-2
Operating Instructions
N2400362
Rev. 1
04/16

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RapidFISH 240200 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Boekel RapidFISH 240200

  • Page 1 RapidFISH Models 240200 and 240200-2 Operating Instructions N2400362 Rev. 1 04/16...
  • Page 2: Table Of Contents

    Table of Contents Safety ..................3 Intended Use of Product ............4 Installation ................4 Unpacking the Unit ................. 4 Set up ....................4 Control Panel Overview ............4 Operation ................5 Turning the Unit On ................ 5 Modes of Operation ................ 6 Calibration ................
  • Page 3: Safety

    Safety The following symbols marked on the equipment mean: Caution: Read these operating instructions fully before use and pay particular attention to sections containing this symbol. Attention: Suivre attentivement les instructions avant l’usage et prêtez une attention particulière aux sections comportant ce symbole.
  • Page 4: Intended Use Of Product

    Intended Use of Product The Boekel Scientific RapidFISH is designed for use in slide hybridization and to simplify Flourescent In-Situ Hybridization (FISH) procedures for FFPE samples, cell suspensions and other tissue preparation procedures.
  • Page 5: Control Panel Overview

    Control Panel Overview A. Program – Cycles through the different programmed steps, “S1”, “S2”, “S3”, and “HT”. B. Set – Allows changing of time and temperature values. Used to toggle between time adjustment and temperature adjustment. C. Start/Stop/Reset – Starts, stops, and resets the programmed runs. Single press of the button will start, stop, or resume the programs.
  • Page 6: Modes Of Operation

    5.1.2 Turn unit on by switching the power switch to the on position (I). The unit will go through a brief startup test, which includes displaying the software revision level. 5.1.3 After the start up test is completed, the unit will switch to normal operation mode.
  • Page 7: Calibration

    Place a certified reference thermometer in the center of the tray. Ensure the thermometer is not touching any parts of the tray. Boekel calibration kit part number 240550 can be used to assist in the setup. Page 7 of 13...
  • Page 8 Turn the unit on and simultaneously press the Up and Down arrow keys and the Program key to enter calibration mode. Allow the unit 60 minutes to achieve calibration setpoint of 40.0°C. Once the 60 minutes have elapsed, the unit will beep once and flash the 40.0°C text on the display.
  • Page 9: Technical Specifications

    Technical Specifications This equipment is for indoor use and will meet its performance specifications within the ambient temperature range of 10°C to 30°C, with maximum relative humidity of 80%. Installation category II (transient voltages). Pollution degree 2 in accordance with IEC 664.
  • Page 10: Additional Accessories Available For Purchase

    Additional Accessories Available for Purchase Item # Description 240550 Calibration Kit Fault Diagnosis Symptom Possible Cause Action Required Unit does not operate a. Unit is not switched a. Switch the unit on b. Unit is not plugged b. Plug in, switch on into a power supply c.
  • Page 11: Maintenance And Service

    10.0 Maintenance and Service All Boekel laboratory products are designed to comply with IEC61010-1. No routine maintenance is required. 10.1 Cleaning Prior to cleaning, turn power off, disengage power cord and allow unit to return to ambient temperature, if still warm. Both the inner chamber and outer housing can be cleaned with a cloth dampened with water and mild soap.
  • Page 12: Contrast Adjustment

    Boekel warrants this product to be free of defective material and workmanship for a period of two years from the date of manufacture. The liability of Boekel for any defective equipment during the warranty period shall be limited to the repair of such equipment or replacement thereof without charge for parts or labor.
  • Page 13: Service

    13.0 Service It is required to obtain a Returned Material Authorization (RMA) number before any Boekel products are returned for any reason. A Decontamination Certificate must be completed, signed by the user, and returned to Boekel Scientific prior to receiving the RMA number.