Scan:mode; Scan:port - HP E1343A User Manual

Relay multiplexer
Table of Contents

Advertisement

Example

SCAN:MODE?

Example

SCAN:PORT

Parameters
Comments
Example
62 Relay Multiplexer Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Selecting the 4-Wire Ohms Mode
TRIG:SOUR EXT
SCAN:MODE FRES
SCAN (@100:107)
INIT
[ROUTe:]SCAN:MODE?
returns
NONE
,
VOLT
,
RES
or
mode, respectively.
FRES
Query Scan Mode
Since this example selects the
returns
FRES
.
SCAN:MODE FRES
SCAN:MODE?
[ROUTe:]SCAN:PORT <
port
tree switches during scanning.
switches (for Analog Bus connections). For correct operation, set the measurement
modes by the
[ROUTe:]SCAN:MODE
The
[ROUTe:]SCAN:PORT NONE
Parameter Name
port
<
>
Order of Command Execution: The
be executed before the
*RST Condition:
[ROUTe:]SCAN:PORT NONE
Selecting the ABUS Port
TRIG:SOUR EXT
SCAN:MODE FRES
SCAN:PORT ABUS
SCAN (@100:107)
INIT
Selects external trigger source.
Selects the 4-wire ohms scanning mode.
Sets channel list.
Starts scanning cycle.
returns the current state of the scan mode. The command
, or
FRES
if the scan mode is in the
FRES
(4-wire ohms) mode, the query command
Selects the 4-wire ohms scanning mode.
Query the scanning mode.
>
enables/disables the closing of the AT, BT, and AT2
SCAN:PORT ABUS
closes the appropriate tree
command.
command prevents closing the tree switches.
Parameter Type
discrete
[ROUTe:]SCAN:PORT
[ROUTe:]SCAN <
channel_list
Selects external trigger source.
Selects the 4-wire ohms scanning mode.
Selects the ABUS port.
Sets channel list.
Starts scanning cycle.
NONE
,
VOLT
,
RES
,
Range of Values
ABUS|NONE
command must
>
command.

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1344aE1345aE1347a

Table of Contents