Subsystem Syntax
Comments
Example
50 Relay Multiplexer Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
subsystem stops a scan in progress when the scan is enabled

ABORt

via the interface, and the trigger modes are
TRIGger:SOURce HOLD
ABORt
ABORt
Operation invalidates the current channel list and sets
(one scanning cycles per
(no continuous scanning cycles), and sets
(continuous internal triggering).
Stopping Scans Enabled from Interface: When a scan is enabled from the
interface, use an interface
"Reset Instr" or "Clear Instr" key to stop the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS
HP E1301 front panel "Reset Instr" or "Clear Instr" keys to stop the scan.
Stopping Scans Enabled from Front Panel: When a scan is enabled from the
HP E1301 front panel, execute
"Reset Instr" or "Clear Instr" keys to stop the scan.
Related Commands:
Stopping a Scan with
ABORt
TRIG:SOUR BUS
INIT:CONT ON
SCAN (@100:115)
INIT
ABOR
TRIGger:SOURce BUS
.
INITiate
command), sets
TRIGger:SOURce IMMediate
command or the HP E1301 front panel
CLEAR
or
TRIGger:SOURce HOLD
over the interface or the front panel
*RST
,
ARM
INITiate:CONTinuous
Bus is trigger source.
Set continuous scanning.
Sets channel list.
Starts scanning cycle.
Aborts scan in progress.
ABORt
or
ARM:COUNt 1
INITiate CONTinuous OFF
, use
or the
ABORt
,
,
[ROUTe:]SCAN
TRIGger