Download Print this page

Maintenance - LaMotte WaterLink SpinTouch 3580-AU Manual

Advertisement

¬ Only the Universal Disk Cover (Code 1719) can be used with the WaterLink Spin
Touch.
¬ Empty syringe of old sample before filling with next sample.
¬ Remove filled disk from meter after testing. Do not travel with filled disks in meter.
They may leak.
¬ Keep the chamber clean and dry. Gently swab LED and photodiode lenses located
around the hub with a cotton swab dampened with streak-free window cleaner.
Do not use alcohol. It will leave a thin film over the lenses when dry.
¬ Hold syringe vertically when filling disks.
¬ Store disks at 70°-80°F/21°-27°C.

MAINTENANCE

CLEANING
The optical system of the WaterLink Spin Touch must be kept clean and
dry for optimal performance. Dry the disk with a lint-free wipe before placing it into
the chamber to avoid introducing moisture. For best results, store the instrument
in an area that is dry and free from aggressive chemical vapors. Clean the exterior
housing with a damp, lint-free cloth. Do not allow water to enter the light chamber
or any other parts of the meter. To clean the light chamber and optic lenses, point a
can of compressed air into the light chamber and the lid and blow the pressurized air
into the light chamber and lid. Focus the pressurized air around the LEDs which are
the small round lenses positioned at 2:00, 4:00, 6:00, 8:00, 10:00 and 12:00 in the
lid. The photodiodes are located on the bottom of the chamber around the hub. This
area must be kept clean and dry. Use a cotton swab dampened with Windex
cleaner to gently swab the LED and photodiode lenses. Do not use alcohol; it will
leave a thin residue over the optics when dry.
Remove smudges due to routine use from the touchscreen with the Cloth Wipe
(Code 3580-WIPE). Use a cloth dampened with alcohol for more thorough cleaning
when necessary. Do not use Windex window cleaner, or similar cleaners, on the
touchscreen.
RETURNS
Should it be necessary to return the meter, pack the meter carefully in
a suitable container with adequate packing material. A return authorization number
must be obtained from LaMotte Pacific emailing support@lamottepacific.com.
Often a problem can be resolved over the phone or by email. If a return of the meter
is necessary, attach a letter with the return authorization number, meter serial
number, a brief description of problem and contact
information including phone and FAX numbers to the shipping carton.
METER DISPOSAL
Waste Electrical and Electronic Equipment (WEEE). Natural
resources were used in the production of this equipment. This equipment may
contain materials that are hazardous to health and the environment. To avoid harm
to the environment and natural resources, the use of appropriate take-back systems
is recommended. The crossed out wheeled bin symbol on the meter encourages the
use of these systems when disposing of this equipment.
Take-back systems will allow the materials to be reused or recycled in a
way that will not harm the environment. For more information on approved
collection, reuse, and recycling systems contact local or regional waste
administration or recycling services. Do not incinerate the equipment
DISK DISPOSAL
The disks cannot be reused. Over time, the water in reacted disks
will eva3porate. Disks can be recycled. Warning: Recyclers should check with the
local authorities. Some states may require that no chemical residue remains on the
plastic or may not be able to accept plastic waste with stainless steel mixing beads.
Used disks may be returned, at the customer's expense, to LaMotte for recycling.
GENERAL INFORMATION
PACKAGING AND RETURNS
Experienced packaging personnel at LaMotte
Company assure adequate protection against normal hazards encountered
in transportation of shipments. After the product leaves the manufacturer, all
responsibility for its safe delivery is assured by the transportation company. Damage
claims must be filed immediately with the transportation company to receive
compensation for damaged goods. Should it be necessary to return the instrument,
pack the instrument carefully in a suitable container with adequate packing material.
A return authorization number must be obtained from LaMotte Company by calling
1-800-344-3100 or 1-410-778-3100, ext. 3 or emailing
lamotte.com. Attach a letter with the authorization number to the shipping carton
which describes the kind of trouble experienced.
GENERAL PRECAUTIONS
Read the instruction manual before attempting to set
up or use the instrument. Failure to do so could result in personal injury or damage
to the meter. The WaterLink Spin Touch should not be stored or used in a damp
or excessively corrosive environment. Care should be taken to prevent water or
reagents from entering the photometer chamber. Wet disks should never be put into
the photometer chamber.
SAFETY PRECAUTIONS
Read the safety precautions on the labels of all reagent
containers and packaging prior to use. Safety Data Sheets (SDS) can be found at
www.lamotte.com. Additional emergency information for all LaMotte reagents is
available 24 hours a day from the National Poison Control Center at 1-800-222-
1222 or by contacting the 24 hour emergency line for ChemTel at 1-800-255-3924
(USA, Canada, Puerto Rico). For locations outside of the North American continent
call 813-248-0585 collect.
Disk US Patent No. 8.734.734
FCl US Patent No. 8.987.000
FCl EU Patent No. EP2784503 A1
TCl US Patent No. 8.993.337
LaMotte, WaterLink, Spin Touch are ® registered
trademarks of LaMotte Company © 2023 LaMotte
Company, All Rights Reserved
Ensure that the protection provided by this equipment is not impaired. Do not install
or use this equipment in a manner that is not indicated in this manual.
LIMITS OF LIABILITY
loss of life, property, profits, or other damages incurred through the use or misuse of
its products.
CE MARK
earned the European CE Mark of compliance for electromagnetic compatibility and
safety. To view certificates, go to the LaMotte website at www.lamotte.com.
This device complies with Part 15 of the FCC rules. Operation is subject to the
following two conditions (1) This device may not cause harmful interference, and (2)
this device must accept any interference received, including interference that may
cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment
is operated in a residential environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the user
is encouraged to try to correct the interference by one or more of the following
window
®
measures:
¬ Reorient or relocate the receiving antenna.
¬ Increase the separation between the equipment and receiver.
¬ Connect the equipment into an outlet on a circuit different from that to which the
¬ Consult the dealer or an experienced radio/TV technician for help.
NSF CERTIFICATION
and operating range. Certification levels of L1, L2, and L3 are possible with L1 being
the highest rating. Go to
For the most recent WaterLink Connect 2 update and the most recent Spin Touch
firmware update, customers who intend to use a WaterLink Spin Touch Meter for
NSF certification compliance purposes should:
1. Visit
2. Plug the Spin Touch into the computer using the included USB cable.
3. Open the WaterLink Connect 2 desktop application and wait for the update to
Test Factor
pH
Free Chlorine
Chlorine, Combined
Cyanuric Acid
Effective April 2022. Based on a 12 month shelf life. Other test factors not evaluated
by NSF.
Go to
softwaresupport@
supported, and a current listing of LaMotte NSF 50 certified products.
WARRANTY
in parts and workmanship for 2 years from the date of shipment. Keep the proof
of purchase for warranty verification. If it should become necessary to return
the instrument during or the warranty period, contact LaMotte Pacific emailing
support@lamottepacific.com. for a return authorization number or visit
www.lamotte.com
charges, freight, insurance, and proper packaging to prevent damage in transit. This
warranty does not apply to defects resulting from action of the user such as misuse,
improper wiring, operation outside of specification, improper maintenance or repair,
or unauthorized modification. LaMotte Company specifically disclaims any implied
warranties or merchantability or fitness for a specific purpose and will not be liable
for any direct, indirect, incidental, or consequential damages. LaMotte Company's
total liability is limited to repair or replacement of the product with a new or
refurbished meter as determined by LaMotte Company. The warranty set forth above
is inclusive and no other warranty, whether written or oral, is expressed or implied.
Water Testing Leader Since 1919
Australia and New Zealand Support
1300-723-488 • support@lamottepacific.com.au
LaMotte Pacific
2/6 Bonz Place, Seven Hills, NSW 2147
PO Box 345, Seven Hills, NSW 1730
Under no circumstances shall LaMotte Company be liable for
The WaterLink Spin Touch meter has been independently tested and has
receiver is connected.
The NSF/ANSI/CAN 50 certification mark on a water quality
testing device (WQTD) used in recreational waters, such as pools and spas,
means that the product was reviewed and certified by NSF International
to meet applicable American National Standards for product design
and performance. NSF tests and re-tests products to confirm that they
comply with all requirements of certification for performance, accuracy
www.nsf.org
softwarecenter.lamotte.com
application for Windows.
complete.
Application
Pool
Spa/Hot Tub
Pool
Spa/Hot Tub
Pool
Spa/Hot Tub
Pool
www.NSF.org
to review the latest levels, ranges, shelf life, firmware versions
LaMotte Company warrants this instrument to be free of defects
for troubleshooting help. The sender is responsible for shipping
for more information.
and download the WaterLink Connect 2
NSF Level
NSF Certified Range Accuracy/Overall
6.4–8.6
L1
6.4–8.6
L1
0–15 ppm
L1
0–7 ppm
L1
0-1 ppm
L2
0-1 ppm
L2
5–50 ppm
L1
05.09.2023
3580-QG-AU-A4 | EN |
Certified to
NSF/ANSI/CAN 5
0

Advertisement

loading

This manual is also suitable for:

Waterlink spintouch 3581-au