Installing the Ultrasound Localizer
A
Clamp for power cable
Ground lead connector
Corru-
gated
Clamp for corrugated hose
hose
Fig. 9
Install the additional board D7 in the control frame. Position D7 E (motor output stage for
cone rotation).
Remount the EMC cover.
Remove the lower shockwave head cover.
Remove the jumper from switch S42 (cone drive), if existing, or remove jumper X42 on
board W2 in the cone drive (Fig. 10).
Reattach the cover.
Connect the fiber optic cable to board D6 (observe the color code).
Attach the adhesive clamps to positions 22 and 6 on the right side of the control cabinet.
Secure thefiber optic cable to the adhesive clamps.
Connect the plug from the table-side operating panel to the ultrasound system.
Insert the inline probe in the storage compartment and connect it.
Connect the corrugated tubing to the SONOLINE, if required.
If the corrugated tubing has a plug, a wall support bracket is required.
Siemens AG
Medical Solutions
Power cable connector
D3
Cable duct to cabinets
W2
Fig. 10
RXL2-120.033.01
Rev. 03
11.01
X5
Page 9 of 14
LITHOSTAR Multiline
CS PS 24
9