Installing the Ultrasound Localizer
A
Clamp for power cable
Ground lead connector
Corru-
gated
Clamp for corrugated hose
hose
Fig. 9
•
Install the additional board D7 in the control frame. Position D7 E (motor output stage for
cone rotation).
•
Remount the EMC cover.
•
Remove the lower shockwave head cover.
•
Remove the jumper from switch S42 (cone drive), if existing, or remove jumper X42 on
board W2 in the cone drive (Fig. 10).
•
Reattach the cover.
•
Connect the fiber optic cable to board D6 (observe the color code).
•
Attach the adhesive clamps to positions 22 and 6 on the right side of the control cabinet.
•
Secure thefiber optic cable to the adhesive clamps.
•
Connect the plug from the table-side operating panel to the ultrasound system.
•
Insert the inline probe in the storage compartment and connect it.
•
Connect the corrugated tubing to the SONOLINE, if required.
If the corrugated tubing has a plug, a wall support bracket is required.