Siemens LITHOSTAR Multiline Installation And Setting Instructions Manual page 5

Hide thumbs Also See for LITHOSTAR Multiline:

Advertisement

Installing the Ultrasound Localizer
2
Fig. 2
Mounting the probe holder
Place the template on the right side cover as shown in figure 2 and secure it with
adhesive tape.
Attach a blade suitable for wood to the jigsaw and cut out the section marked (2/Fig. 2).
Deburr the cutting line.
Systems with serial no. < 06045
Hold the mounting panel to the right
support (Fig. 3) and mark the mounting
holes. Watch for the correct placement of
the angle piece (3/Fig. 3).
Cover the drilling area to prevent drill
chips from falling into the unit.
Drill a 6 mm holes.
Remove any drill chips.
Move the unit to the starting position ;
release the button when the system stops
in X direction.
Attach the mounting panel (4/Fig. 3) with
2 screws M4 x 12,2 washers, and 2 lock
nuts.
Tab. 1
CAUTION
Siemens AG
Medical Solutions
When moving the W drive back, be careful not to damage the
cables.
RXL2-120.033.01
Rev. 03
3
4
Fig. 3
Systems with serial no. 06045
Move the unit to the starting position;
release the button when the system stops
in the X direction.
Hold the mounting panel to the support on
the right (Fig. 3). Be sure to correctly place
the angle piece (3/Fig. 3).
Attach the mounting panel (4/Fig. 3) with
2 screws M4 x 12, and 2 washers.
Page 5 of 14