8
Attaching the corrugated hose to the unit
Fig. 8
Cable type
1 x 4 mm2
3 x 0.75 mm2
shielded
Canon, 9 pole
Fiber optic cable
Video
•
Connect the Sonoline protective conductor to the stud on the cable holder (Fig. 9).
•
Attach the screw mount for the power cable:
- Loosen the screw (A/Fig. 9) and then retighten the clamp.
•
Connect the SONOLINE power supply cable:
- for 220 V:
- for 115 V: (The parts are included in the bag for Power supply 115 V)
•
Route the video cable and the fiber optic cable to the control cabinet.
•
Remove the EMC cover from N11 in front of boards D7 d/c, if existing.
LITHOSTAR Multiline
System
Installing the Ultrasound Localizer
6
Function
Protective con-
ductor
Power supply
Control cable
Communication Back panel
- Connect the wires to the line voltage terminal NK 1, 2, 3 (Fig. 9)
For shielding, refer to (A/Fig. 9).
- Attach the adapter plug (female part) und cable conduit supplied to the
115 V power supply cable routed in the base plate on-site.
- Cut off the Faston connector of the line voltage cable protruding from
the corrugated hose. Replace the Faston connector with the the pin-
shaped (male) connector and the cable conduit.
- Connect both adapter plugs together (from the corrugated hose - from
the power supply)
•
Select urology mode or insert the uro-panel.
•
Remove both covers from the base plate
(6/Fig. 8).
•
Turn the unit on and move the system in the
Y direction to the left until D31 is accessible.
•
Open the cable entry hole at the right rear
portion of the base frame.
•
Attach the support plate for the corrugated
hose with screws underneath the opening.
•
Feed the Sonoline corrugated hose through
the opening and secure it with the hose clamp
to the support plate.
Ultrasound
device
Back panel
Back panel
VBF US
No function
RXL2-120.033.01
Rev. 03
11.01
Multiline
Unit
Cabinet A12
Protective con-
ductor terminal
refer to text
below
D31.X28