Installing the Ultrasound Localizer
Systems with serial no. < 06045
•
Install the probe cable inside the probe
holder (Fig. 6).
•
Place any excessive cable length inside
the holder.
•
Attach the probe holder
- front: panel screw and threaded plate.
- back: contact washer and lock nut.
Tab. 2
•
Move the shockwave head to the gel position.
CAUTION
•
When moving the W drive out, check that the probe holder does not contact board D7.
If yes, reposition the probe holder so that it clears board D7.
•
Raise the table top to its highest position.
•
Move the unit back to the end position (X direction).
•
Reattach the inner right-hand cover to the coupling system (1/Fig. 1).
•
Reattach the right side panel.
•
Screw the cable ties to the support bracket (Fig. 7) using washers and lock nuts.
•
Use cable ties to secure the cable (Fig. 7).
Siemens AG
Medical Solutions
Avoid damaging the cables when moving the W drive forward.
RXL2-120.033.01
Rev. 03
Systems with serial no. 06045
•
Attach the probe holder
- front: panel screw.
- back: contact washer and lock nut.
Page 7 of 14
11.01
CS PS 24
7
LITHOSTAR Multiline
System