Installing the Ultrasound Localizer
Systems with serial no. < 06045
Install the probe cable inside the probe
holder (Fig. 6).
Place any excessive cable length inside
the holder.
Attach the probe holder
- front: panel screw and threaded plate.
- back: contact washer and lock nut.
Tab. 2
Move the shockwave head to the gel position.
CAUTION
When moving the W drive out, check that the probe holder does not contact board D7.
If yes, reposition the probe holder so that it clears board D7.
Raise the table top to its highest position.
Move the unit back to the end position (X direction).
Reattach the inner right-hand cover to the coupling system (1/Fig. 1).
Reattach the right side panel.
Screw the cable ties to the support bracket (Fig. 7) using washers and lock nuts.
Use cable ties to secure the cable (Fig. 7).