PREMASGARD
The pressure measuring transducer PREMASGARD
It converts the measurand pressure into standard signals of 0 -10 V or 4...20 mA. Process connection is G ½ " straight external pipe thread.
SHD-LC is used for pressure measurement in gaseous and liquid media. Applications of this pressure transmitter are in hydraulics,
pneumatics, process technology, in mechanical and plant engineering.
Not applicable for ammonia and Freon!
TECHNICAL DATA:
Power supply: ........................ 24 V AC ⁄ DC for output 0 -10 V
Measuring ranges: .................. see table
Output signal: ........................ 0 -10 V, 3-wire, (working resistance > 10 kOhm) or
Electrical connection: .............. plug connector DIN EN 175301-803-A
Pressure connection: ............... G ½ " sealing at the back,
Type of pressure: .................... relative
Measuring principle: ................ ceramic measuring cell
Temperature of medium: .......... –15...+ 125 °C
Mounting: .............................. directly on pressure line
Enclosure: ............................. stainless steel 1.4305
Connecting head: .................... plastic, approx. 98 x 50 x 34 mm
Medium contacting parts:......... stainless steel type 1.4305, AISI 303;
Response time: ...................... 2 ms (1 ms typical)
Load changes: ........................ < 100 Hz
Characteristic line: ................. ± 0.3 %
Overload range/
Bursting pressure: .................. < 4 bar: 3 x FS
Protection class: .................... III (according to EN 60 730)
Protection type: ..................... IP 65 (according to EN 60 529)
Standards: ............................ CE conformity, electromagnetic compatibility
ACCESSORIES: ....................... incl. receptacle DIN EN 175301-803-A
Connecting diagram
2
Output pressure 0-10V
3
GND
1
Supply voltage
UB+ 24V AC/DC
Type ⁄ WG1
Measuring
SHD -LC - U
S H D - L C - U 6
0 ... 6 bar
S H D - L C - U 10
0 ...10 bar
S H D - L C - U 16
0 ...16 bar
SHD - LC
®
SHD-LC measures relative pressures in the bar range.
7 - 33 V DC for output 4 ... 20 mA
(other ranges upon request)
4... 20 mA, 2-wire, (working resistance < (UB (V) -7 V) ⁄ 0,02 A ;
R
depending on working resistance
L
and manometer (combined) with profile gasket FPM,
special WW G ¼ " DIN 3852
measuring element ceramic Al
sealing material FPM (Viton)
> 4 bar: 2.5 x FS
according to EN 61 326, EMC directive 2004 ⁄ 108 ⁄ EC
SHD - LC - U
OUT
2
3
1
+
IN
Item No.
Range
U - variant
1 3 0 1 - 2 1 2 1 - 0 5 5 0 - 1 2 0
1 3 0 1 - 2 1 2 1 - 0 5 6 0 - 1 2 0
1 3 0 1 - 2 1 2 1 - 0 5 7 0 - 1 2 0
0
(96%);
2
3
Connecting diagram
2
Output pressure 4...20mA
GND
3
Free
1
Supply voltage
UB+ 24V DC
Type ⁄ WG1
SHD -LC - I
S H D - L C - I 6
S H D - L C - I 10
S H D - L C - I 16
Rev. 2015 - 1.0 GB
OUT
2
3
1
+
IN
Measuring
Item No.
Range
I - variant
0 ... 6 bar
1301-2122- 0550-000
0 ...10 bar
1301-2122- 0560-000
0 ...16 bar
1301-2122- 0570-000
SHD - LC - I