Druck DPI 841 User Manual

Druck DPI 841 User Manual

Frequency calibrator and frequency loop calibrator

Advertisement

Quick Links

Druck DPI 841/842
Frequency calibrator and
Frequency loop calibrator
User manual - K395
Visit us at www.TestEquipmentDepot.com
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DPI 841 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Druck DPI 841

  • Page 1 Druck DPI 841/842 Frequency calibrator and Frequency loop calibrator User manual - K395 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com...
  • Page 2 DPI 842 K395 Issue 3...
  • Page 3: Table Of Contents

    Table of Contents Introduction The DPI 841 Frequency Calibrator and DPI 842 Frequency Introduction ................. 1 Loop Calibrator are part of the Druck DPI 800 series of Safety ..................1 hand held instruments. Marks and symbols on the instrument ......... 2 The DPI 800 series uses Intelligent Digital Output Sensor To start ..................
  • Page 4: Marks And Symbols On The Instrument

    Safety (Continued) To start - Items on the display Item Description • To prevent electrical shocks or damage to the DPI 842 only. Task indication for the switch test. instrument, do not connect more than 30V between = switch closed the terminals, or between the terminals and the = switch open ground (earth).
  • Page 5: Power On Or Off

    To start - Power on or off To start - Select a task (Measure and/or supply) To turn the instrument on or off, press ❍ (A1 - item [1]). The When the instrument is set up (Table 1), use the Select Task instrument does a self test and then shows the applicable menu to select the applicable task.
  • Page 6: Set Up The Settings

    To start - Set up the settings Table 4: (Part of table) Menu options - Settings (Input) When the task is set up (Table 2/3), use the Settings menu Options Description to adjust the input and/or output operation. (If applicable) To select and set up a tare value for the Display: Task Settings selection...
  • Page 7 Example 2) Set up values for x:y Scaling = 0 to 100%. Table 5: (Part of table) Menu options - Settings (Output) Settings Settings Options Description Edit To select and set up values for the “Defined Step” ... Step ▲ output.
  • Page 8: Operation

    Operation Operation - Change the output values When the output operation is set up (Table 5), use one of This section gives examples of how to connect and use the these procedures to change the output values: instrument. Before you start: •...
  • Page 9: Supply Hz Or Pulses

    Operation - Supply Hz or pulses To supply Hz or pulses: Maximum: 30V 1. Connect the instrument (Figure 2) and, if necessary, adjust the Set Up (Table 1). 2. Select a Hz or Pulses output task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 5).
  • Page 10: Ma Measurements

    Operation - mA measurements 3. Supply the output values to the system (Table 7). DPI 842 only. To measure a current: • Example - “Nudge” output. 1. Connect the instrument (Figure 5) and, if necessary, a. Use Edit (■ ■) to set a value less than the switch adjust the Set Up (Table 1).
  • Page 11: Upm Pressure Measurements

    Operation - UPM Pressure measurements Read all the instructions supplied with the UPM and then use the specified procedures to connect it (Figure 7/8). Settings Start Settings a) Pressure b) Leak test Settings Figure 8: Example configuration - To measure pressure Figure 7: Example configuration - Pressure measurement and frequency with a UPM...
  • Page 12: Maintenance

    Total error: 7 ppm or better Resolution: 8 digits (minimum) Frequency = 990 Hz 2) Signal generator DPI 841/842: Input units = Hz (Table 4) Pressure UPM only. Refer to the user manual for the IDOS Input trigger level = 5V (Table 4) UPM.
  • Page 13: Procedures (Ma Input)

    Hz then kHz. Make sure that the error is in the specified 2. Let the equipment get to a stable temperature limits. (minimum: 5 minutes since the last power on). Table 10: Hz error limits (Measure/Supply) 3. Set up the DPI 841/842 with these conditions: Measure/ Calibrator Permitted Supply...
  • Page 14: Procedures (Idos Umm)

    7. To make sure that the calibration is correct, set up the Specification - Frequency DPI 841/842 with these conditions: cpm = Pulses/minute, cph = Pulses/hour Range (Measure): Accuracy: Output Hz = 0 (For direct current output) 0 ... 999.999 Hz...