Siemens MAMMOMAT 1000 Nova Opdima Maintenance Instruction page 11

Hide thumbs Also See for MAMMOMAT 1000 Nova Opdima:
Table of Contents

Advertisement

System
NOTE
PMP
General cable information
Check the connectors and cables between the biopsy controller and CCD camera.
Check the connectors and cables between the biopsy controller and workstation.
Check that all cables are undamaged, especially the monitor cables.
Check that the connectors are properly attached.
SW
Disk cache settings
For instructions on how to check the disk cache settings, see the Service Instructions for
Opdima®, chapter "Disk cache settings".
SI
Biopsy unit
This unit contains the following safety-related parts which should be checked every year
and replaced as necessary:
Biopsy unit connectors (25-pin DSUB connector and 9-pin object table connector). Vi-
sual inspection
Sliding pads on the underside of the biopsy unit object table. Check that the sliding pads
are free of apparent and visible damage and that the biopsy unit object table runs
smoothly on the base plate.
Check that the biopsy unit is safely locked to the MAMMOMAT.
Siemens AG
Medical Solutions
It is important not to interrupt the system when it is saving images
on the MO disk because the MO disk can be destroyed.
SPB7-230.101.03.05.02
02.06
Page 11 of 18
CS PS SP
11
MAMMOMAT Nova - Opdima

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mammomat 3000 nova opdima

Table of Contents