Fluke 287C Calibration Manual page 32

True-rms digital multimeters
Table of Contents

Advertisement

287/289
Calibration Manual
Rotary Switch
Step
20
VDC
21
VDC
22
VDC
23
VDC
24
VDC
25
VDC
26
VDC
27
VDC
28
VDC
29
VDC
30
mV DC
Ω
31
Ω
32
Ω
33
Ω
34
Ω
35
Ω
36
37
Ω
Ω
38
39
Diode Test
40
Diode Test
μA DC
41
42
μA AC
43
mA DC
44
mA AC
45
ADC
46
AAC
28
Table 8. Calibration Adjustment Steps for Firmware Version 1.10 (cont.)
Source Value
Position
0.0 V, 0 Hz
5.0 V, 0 Hz
50.0 V, 0 Hz
500.0 V, 0 Hz
1000.0 V, 0 Hz
5.0 V, 55 Hz
5.0 V, 20.0 kHz
50.0 V, 55 Hz
50.0 V, 20.0 kHz
1000.0 V, 55 Hz
500.0 mV, 55 Hz
Connect the 5520A in a 4-Wire configuration and select COMP 2-Wire
0 Ω, zcomp wire2
500 Ω, zcomp wire2
5 kΩ, zcomp wire2
50 kΩ, zcomp wire2
Disconnect 5520A 4-Wire configuration and turn COMP off
500 kΩ
5.0 MΩ
50.0 MΩ
500.0 MΩ
Connect the 5520A in a 4-Wire configuration and select COMP 2-Wire
0.0 Ω, zcomp wire2
3.1 V, 0 Hz
5000 μA, 0 Hz
5000 μA, 55 Hz
400.0 mA, 0 Hz
400.0 mA, 55 Hz
10.0 A, 0 Hz
10.0 A, 55 Hz
Settling Time
Calibrating Time
(seconds)
2
2
2
2
3
3
3
5
5
5
5
2
3
3
2
2
2
2
2
2
2
2
2
2
2
2
2
[1]
(seconds)
8
1
1
1
1
12
13
16
32
7
8
15
1
1
1
13
9
9
2
1
1
1
1
1
1
1
2