Fluke 287C Calibration Manual page 23

True-rms digital multimeters
Table of Contents

Advertisement

Table 4. 287/289 Performance Test Steps (cont.)
Step
Function
VAC, HZ, %
29.
(Duty Cycle)
30.
VAC
VAC, Lo
31.
Pass (289)
VAC, Lo
32.
Pass (289)
33.
VAC
34.
VAC
35.
DC V
36.
DC V
37.
DC V
38.
DC V
DC V, DC,
39.
AC
DC V, AC,
40.
DC
ACV, Peak
41.
Put Meter's rotary switch to Capacitance function and Range to 10 nF
Remove Test Leads from 5520A and REL offset
Capacitance
42
43
Diode Test
Diode Test
44
Set 5520A for 2-Wire COMP and Connect in 4-Wire configuration
45
LoΩ (289)
46
LoΩ (289)
47
LoΩ (289)
Range
Input Level
5 V p-p, Sq.
5.0000 V
wave @15 %
50.000 V
15 V
500.00 V
50 V
500.00 V
50 V
500.00 V
500 V
1000.0 V
1000 V
5.0000 V
4 V
50.000 V
-40 V
500.00 V
400 V
1000.0 V
600 V
5.0000 V
200 mV
5.0000 V
2 V
4 V p-p, Sq.
5.000 V
wave, +1 V
offset
Reconnect Test Leads to 5520A
10.00 nF
[4]
5 nF
3.1000 V
3.5 kΩ
0 Ω
3.1000 V
50.000 Ω
[1]
0 Ω
REL offset
50.000 Ω
0.2 Ω
[2]
50.000 Ω
50 Ω
[2]
True-rms Digital Multimeters
Display Reading
Frequency
Lower
or Model
Limit
50 kHz
1.40
100 kHz
14.435
60 Hz
48.60
1600 Hz
[3]
0.00
10 kHz
497.75
10 kHz
993.5
- -
3.9988
- -
-40.012
- -
399.86
- -
599.6
0 Hz
0.1977
5 kHz
1.9640
2 kHz
1.863
4.90
- -
- -
2.0000
Check that
- -
Beeper is On
- -
Allow reading to stabilize
(5-10 seconds)
- -
0.180
- -
49.905
Performance Tests
[1]
Upper
Limit
28.60
15.565
51.40
8.00
502.25
1006.5
4.0012
-39.988
400.14
600.4
0.2023
2.0360
2.137
5.10
3.1000
- -
0.220
50.095
19