  • Page 2 9000 Series Transportable Multi Product Calibrator Operation Manual Version 4.10 : May 2010 All product names are trademarks of their respective companies...
  • Page 3 (60 Days after invoice date) AFTER THE EVALUATION PERIOD HAS EXPIRED THE OPERATION OF THE CALIBRATOR IS LOCKED AND THE DISPLAY SHOWS A NUMBER WHICH MUST BE QUOTED TO TRANSMILLE TO RECEIVE THE UNLOCK CODE THE UNLOCK CODE IS AVAILALBLE...
  • Page 4 9000 SERIES OPERATION MANUAL DECLARATION OF CONFORMITY Manufacturer’s Name: Transmille Ltd. Manufacturer’s Address: Unit 4, Select Business Centre Lodge Road Staplehurst TN12 0QW Declares, that the product Product Name: Multi-product Calibrator Model Number: 9050 / 9041 Product Options: This declaration covers all options of the above product(s)
  • Page 5: Table Of Contents

    9000 SERIES OPERATION MANUAL TABLE OF CONTENTS 9000 SERIES CALIBRATOR INTRODUCTION..................6 MAIN FEATURES ..........................6 ACCURACY AND FUNCTIONALITY ....................7 TRUE MULTIPRODUCT CALIBRATION FROM ONE INSTRUMENT..........7 PREPARING THE CALIBRATOR FOR USE..................9 ..........................9 NITIAL NSPECTION ......................9 RANSPORTING THE ALIBRATOR ................10 OWER NTERFACE...
  • Page 6 9000 SERIES OPERATION MANUAL INTERFACE TYPES ..........................49 RS232 I ..........................50 NTERFACE USB I ..........................52 NTERFACE REMOTE PROGRAMMING .........................53 ....................54 ROGRAMMING OMMANDS VERVIEW DC V ........................56 OLTAGE OMMANDS AC V ........................58 OLTAGE OMMANDS DC C ........................60 URRENT OMMANDS AC C ........................62...
  • Page 7: 9000 Series Calibrator Introduction

    9000 SERIES OPERATION MANUAL 9000 Series Calibrator Introduction The 9000 series of calibrators offer a complete transportable multi product calibration system in a rugged enclosure providing AC/DC voltage & current, resistance & capacitance as well as being expandable with add on options.
  • Page 8: Accuracy And Functionality

    Internal retro fit options allow the calibration of power meters, oscilloscopes to 600MHz, inductance and LCR meters. Designed for on use in the laboratory or portable on site calibration the 9000 series calibrator is equally suitable for use in the standards laboratory or for on site calibration work.
  • Page 9 9000 SERIES OPERATION MANUAL Serial Line RS232 Interface Available As Standard. All functions and outputs of the series 9000 calibrator are fully programmable over the RS232 interface fitted as standard. The use of the RS232 interface saves the cost of fitting GPIB cards to the PC, and also allows easy connection to portable PC’s, reducing the set up time for on site calibration.
  • Page 10: Preparing The Calibrator For Use

    9000 SERIES OPERATION MANUAL Preparing The Calibrator For Use. Initial Inspection. After shipment the calibrator should be inspected for any signs of external damage. Should external damage be found contact the carrier immediately. Do not connect a damaged instrument to the line power as this may result in internal damage. Please keep the original box which can be used when returning the calibrator for service and recalibration.
  • Page 11: Power / Interface Connections And Controls

    9000 SERIES OPERATION MANUAL Power / Interface Connections and Controls Connections at the top right of the calibrator control panel are for line power via a 3 pin IEC connector incorporating the line fuse and on-off switch - Note the mains inlet is filtered.
  • Page 12: Setting And Checking The Line Voltage

    9000 SERIES OPERATION MANUAL Setting and checking the Line Voltage. Warning : The line power cord must have an earth conductor to avoid risk of shock. This instrument must be correctly earthed. The calibrator has been designed to work from either 100-120 Volt line supply or 200 - 240 Volt line supply.
  • Page 13: Connecting To A Computer

    9000 SERIES OPERATION MANUAL Connecting to a computer If required, a standard serial 9 pin cable may be used to connect the calibrator to a COM port on a PC. A Null modem cable is not needed. RS232 Interface Connection...
  • Page 14: Output Connections

    Output sockets are all 4mm safety type, the voltage pairs contacts are low thermal gold plated for minimum thermal EMF. The 9000 series calibrators outputs have been designed to allow most multimeters to be calibrated with minimal lead changing. There are 3 separate pairs of outputs : 1) Voltage, Resistance, Capacitance, frequency &...
  • Page 15: Output Overloads

    9000 SERIES OPERATION MANUAL A second example is where the meter has separate voltage and current inputs, often using four wire ohms on both pairs. In this case simply connect the voltage and current outputs to the meter’s inputs, the calibrator will use both the voltage and current pair on four wire ohms.
  • Page 16: Operation

    Introduction to Operation All functions of the 9000 Series Calibrator can be controlled from the front panel. or controlled remotely by a computer over the interface. The front panel controls are ‘locked out’, but local control may be resumed by selecting a soft key - it must be remembered that this action may disrupt the computer program.
  • Page 17: Front Panel Controls And Indicators

    9000 SERIES OPERATION MANUAL Front Panel Controls and Indicators LCD Display with Menu buttons integral backlight Digital Control (Soft Keys) Active Terminal StandBy Button LEDs Output On Button Adapter Safety Output Numeric Keypad Interface Terminals Multiplier Keys Function Keys TRANSMILLE LTD.
  • Page 18: Front Panel Keyboard

    9000 SERIES OPERATION MANUAL Front Panel Keyboard The front panel of the 9000 Series Calibrator utilises a high quality custom rubber keyboard with tactile feel buttons and integral display window. The front panel can easily be wiped clean with a soft cloth. Care should be taken not scratch the display window.
  • Page 19: Front Panel Keyboard - Control Sections

    9000 SERIES OPERATION MANUAL Front Panel Keyboard – Control Sections The Keyboard is divided into sections to allow easy operation. Numeric section Allows numeric values to be entered Multiplier section Mega (M), Kilo (K), milli (m), micro(u) or nano (n) Function section Volts (V), Amps(A), Ohms (2 &...
  • Page 20: Graphic Lcd Display

    9000 SERIES OPERATION MANUAL Graphic LCD Display A back lit graphic LCD display shows the present output, instrument status, % or ppm change from the entered value, and also the new value being entered. The bottom line of the display is used to assign the function of the four ‘soft (menu) keys’...
  • Page 21: Digital Control

    9000 SERIES OPERATION MANUAL Digital Control A digital potentiometer allows ‘highlighted numbers’ on the display to be incremented (turning clockwise) or decrement (turning anti-clockwise). As an output is changed the deviation from the original value entered on the keyboard is shown in either % or ppm.
  • Page 22: Terminal Status Led's

    9000 SERIES OPERATION MANUAL Terminal status LED’s LED’s above the terminals indicate which pair is active. When terminals are not active they are electrically isolated from each other, this enables terminals to be linked together if required. Active terminals indicated by illuminated LED Voltage Output Terminal Pair (Black &...
  • Page 23 9000 SERIES OPERATION MANUAL Current Output Terminals (Black & Red) 4mm safety terminals for all current outputs up to 2 Amps, capacitance, TTL frequency and for sense connection for 4 wire ohms. Note the low ‘black’ terminal can be internally switched to line earth by a soft key function.
  • Page 24: Pin Adapter Interface Connector

    9000 SERIES OPERATION MANUAL Scope BNC Connector Output Isolated BNC Output for oscilloscope calibration. Incorporates green LED indication if BNC is the active terminal. WARNING : Dangerous voltages up to 400V may be present on these terminals. 9 Pin Adapter Interface Connector.
  • Page 25: Setting A Voltage Or Current Output

    9000 SERIES OPERATION MANUAL Setting a Voltage or Current Output To set an output follow the example below to which will set 12.345mA note that the value will appear in the display as it is entered, use ‘backspace’ and clear ’C’ to edit an incorrect entry.
  • Page 26: Automatic Display Of % Or Ppm Error And Ref. Key

    9000 SERIES OPERATION MANUAL Automatic Display of % or ppm Error and Ref. Key When the output value is changed by the methods above, the display will show the change in ppm or % from the original reference value entered from the keyboard. If needed, the reference value can be reset to the present value on the display by the REF key.
  • Page 27: Selecting Ac And Setting A Frequency

    9000 SERIES OPERATION MANUAL Selecting AC and Setting a Frequency. To set the calibrator to either AC volts or Current follow the example below which sets 1234Hz. Note that the value will appear in the display as it is entered, use ‘backspace’...
  • Page 28: Returning The Calibrator To Dc

    9000 SERIES OPERATION MANUAL Returning the calibrator to DC The output can be returned to DC by following the sequence below: 1: Press ‘DC’ Key 2: Enter. Select DC Press Enter The display will show DC in the bottom right hand corner of the display. For safety, AC/DC changes will set the output to zero.
  • Page 29: Setting 2 Wire Resistance Output

    9000 SERIES OPERATION MANUAL Setting 2 Wire Resistance Output. Note : The calibrator uses standard resistors of fixed decade values. The nearest available resistance to the entered value will be automatically selected. The example below selects 100kOhm in 2 wire mode.
  • Page 30 9000 SERIES OPERATION MANUAL 2 Wire Ohms Operation Two wire ohms is output on the voltage terminals as indicated by the terminal LED’s. 2 Wire Ohms Connection Connect the measuring instrument 2-Wire input to the calibrator 2-Wire resistance output. Use shielded connection cables – 50 Ohm co-axial recommended Connect Guard / Earth if required –...
  • Page 31: Setting 4 Wire Resistance Output

    9000 SERIES OPERATION MANUAL Setting 4 Wire Resistance Output. Note : The calibrator uses standard resistors of fixed decade values. The nearest available resistance to the entered value will be automatically selected. The example below selects 100mOhms in 4 wire mode.
  • Page 32 9000 SERIES OPERATION MANUAL 4 Wire Ohms Operation Four wire ohms is output on the voltage and current terminals as indicated by the terminal LED’s. Connect the ‘sense’ from the DMM to the voltage output on the calibrator, and connect the ‘current’ from the DMM to the calibrator current terminals.
  • Page 33: Setting Capacitance Output

    9000 SERIES OPERATION MANUAL Setting Capacitance Output. Note : The calibrator uses standard capacitors of fixed values. The nearest available capacitance to the entered value will be automatically selected. The example below selects 100nF. 1) Key in the required value, e.g. ‘100’...
  • Page 34: Setting Inductance Output (Option)

    9000 SERIES OPERATION MANUAL Setting Inductance Output (Option) Note : The calibrator uses standard inductors of fixed values. The nearest available inductance to the entered value will be automatically selected. The example below selects 10mH. 1) Key in the required value, e.g. ‘10’...
  • Page 35: Thermocouple Simulation (Option)

    Using the supplied adapter connection lead (9 way ‘D’ type to 9 way ‘D’ type), connect the thermocouple adapter to the adapter interface connection on the front panel of the 9000 series calibrator. 9 way ‘D’ to 9 way ‘D’...
  • Page 36 9000 SERIES OPERATION MANUAL Thermocouple Simulation Option Operation The thermocouple simulation option allows the user to enter a temperature in °C and set the calibrator to the specific voltage output which corresponds to the thermocouple type selected. On entering thermocouple simulation mode, type K will be the default function.
  • Page 37 9000 SERIES OPERATION MANUAL Connect the thermocouple simulation adapter to the 9000 Series calibrator via the adapter interface connector using the supplied 9 way ‘D’ type to 9 way ‘D’ type lead. Connect the UUT to the thermocouple plug extending from the opposite end of the thermocouple simulation adapter.
  • Page 38 9000 SERIES OPERATION MANUAL To deviate the temperature output from the nominal value, use the deviation function. This is controlled by using the left and right arrow keys to select the digit to be increased or decreased Use the left right arrows to highlight the digit to be adjusted (indicated by the underlined digit) To increase or decrease the digit, simply use the up down keys on the calibrator’s...
  • Page 39: Pecial Unctions Vailable Using The Soft Keys

    9000 SERIES OPERATION MANUAL Special Functions Available using the ‘soft’ keys The ‘soft’ keys are positioned directly under the display and the function of these keys will change depending on the function of the calibrator. To enable scrolling down through menus NEXT will take you down one level and BACK return to the previous level.
  • Page 40: Selecting Front Panel Control

    9000 SERIES OPERATION MANUAL Selecting front panel control After the calibrator has been controlled from the interface, the front panel controls are disabled. To regain front panel control use the ‘Local’ soft key Setting TTL Logic Frequency Output. Note : The calibrator uses a precision Temperature Compensated Crystal Oscillator (TCXO) and divider chain for this output with exact values available only.
  • Page 41: Setting Pwm (Mark Space Ratio)

    9000 SERIES OPERATION MANUAL Setting PWM (Mark Space Ratio) The calibrator can produce an accurate mark space ratio output at approx. 1.23kHz at 20%, 40%, 60% and 80% intervals. These may be selected using the keyboard or the digital rotary control.
  • Page 42: Selecting Prt (Pt100) Resistance Output (Option)

    9000 SERIES OPERATION MANUAL Selecting PRT (PT100) Resistance Output (Option) Note : The calibrator uses standard resistors of fixed values. The nearest available temperature to the entered value will be automatically selected. 1) Select PRT from the soft menu’s 2) Use up / down arrow keys or the rotary control to select the required temperature.
  • Page 43: Selecting Ac Power Calibration Output (Option)

    9000 SERIES OPERATION MANUAL Selecting AC Power Calibration Output (Option) The Calibrator can simulate power by simultaneously outputting AC voltage and AC current with a phase relationship. Phase & Current Display Selected Waveform (9041 PowerDDS Option Only) Return to DCV mode...
  • Page 44 9000 SERIES OPERATION MANUAL Note : Current output in power mode follows the same convention as normal current operation – Up to 2A is output from the low current terminals 2.02 – 30A (20A for 9050) output from the high current terminals 5) To change phase, select the phase menu item as shown previously Phase &...
  • Page 45: Selecting Dc Power Calibration Output (Option)

    9000 SERIES OPERATION MANUAL Selecting DC Power Calibration Output (Option) The Calibrator can simulate DC power by simultaneously outputting DC voltage and DC current. Current Display Return to DCV mode To configure power calibration mode : 7) Select ‘POWER’ function from the soft menu’s 8) Enter a voltage by keying in a number followed by V, then press ENTER, eg.
  • Page 46: Selecting Oscilloscope Calibration Output (Option)

    9000 SERIES OPERATION MANUAL Selecting Oscilloscope Calibration Output (Option) The Calibrator can provide Amplitude, Timebase and Bandwidth outputs for calibration of oscilloscopes. To configure oscilloscope calibration mode : 1) Select SCOPE function from the soft menu’s 2) Calibrator will select TIMEBASE mode.
  • Page 47 9000 SERIES OPERATION MANUAL 4. For BW SWEEP output, select the BW SWEEP. Mode using the softkey. To select a digit, use the left/right arrow buttons. To change the output, use the up / down arrow buttons or use the digital potentiometer.
  • Page 48: Warning And Output Overload Indications

    9000 SERIES OPERATION MANUAL Warning and output overload indications. The self test function of the 9000 series calibrator also continuously monitors the output of the calibrator of overload or fault conditions. In the event of the calibrator not being able to drive the load, it will automatically trip into standby and the display will show STBY !.
  • Page 49: Dry Block Temperature Calibration Module

    9000 SERIES OPERATION MANUAL Dry Block Temperature Calibration Module Range : Ambient to 200°C Built-In Digital Temperature Controller Multiple Drilled Insert – Supports 5 Probe Sizes Supports °C and °F Integrated temperature Sensor Using the Module 1. Switch on the module using the red power switch.
  • Page 50: Interface Types

    9000 SERIES OPERATION MANUAL Interface Types Connection to the 9000 Series calibrators is achieved by the following interfaces : RS232 9 Way ‘D’ type female connection USB (optional) USB connection using COM to USB conversion via a dedicated adapter TRANSMILLE LTD.
  • Page 51: Rs232 Interface

    9000 SERIES OPERATION MANUAL RS232 Interface The calibrator can be fully controlled and calibrated via the bi-directional RS232 interface. The interface uses the standard 9 pin PC connector and a standard serial lead. The interface is fully optically isolated from the rest of the calibrator circuitry.
  • Page 52 DO NOT TRY TO USE A NULL MODEM TYPE CABLE AS THIS HAS PINS 2 & 3 REVERSED AND WILL NOT WORK. Connect the RS232 Interface from the 9 Way ‘D’ type connector on the rear of the PC to the 9 Way ‘D’ type connector on the 9000 Series calibrator. TRANSMILLE LTD.
  • Page 53: Usb Interface

    RS232 COM port has been replaced by one or more USB ports. Connect the USB Interface from the USB type connector on the rear of the PC to the 9 Way ‘D’ type connector on the 9000 Series calibrator via the RS232 to USB converter. TRANSMILLE LTD.
  • Page 54: Remote Programming

    9000 SERIES OPERATION MANUAL Remote Programming WARNING The 9000 series calibrators can produce high voltages up to 1025V and must be programmed with due caution to prevent dangerous voltages from being output without warning to the operator. Any programs should be extensively tested to maintain safe operation and include safeguard’s such as error catchments and...
  • Page 55: Programming Commands Overview

    9000 SERIES OPERATION MANUAL Programming Commands Overview The 9000 series is controlled by a set of simple high level commands which can be used either individually or as part of a command sequence to setup the 9000 Series calibrator to required state.
  • Page 56 9000 SERIES OPERATION MANUAL RESPONSE CODES The 9000 Series calibrators will respond to any command with a fixed code beginning with an star (*) - the codes are listed below Response Code Description ERROR IN COMMAND LINE ERROR IN RANGE COMMAND...
  • Page 57: Dc Voltage Commands

    9000 SERIES OPERATION MANUAL DC Voltage Commands Function Range Command DC Voltage 200mV 200V 1000V Standby Mode Standby ON Standby OFF Output Set Output O (not zero) The DC voltage section consists of a set of range commands which are used in conjunction with the standby and output command.
  • Page 58 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 59: Ac Voltage Commands

    9000 SERIES OPERATION MANUAL AC Voltage Commands Function Range Command AC Voltage 200mV 200V 1000V Standby Mode Standby ON Standby OFF Output Set Output O (not zero) AC Frequency Fxxxxx E.G. 10kHz = F10000 The AC voltage section consists of a set of range commands which are used in conjunction with the standby and output command.
  • Page 60 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 61: Dc Current Commands

    9000 SERIES OPERATION MANUAL DC Current Commands Function Range Command DC Current 200uA 20mA 200mA Standby Mode Standby ON Standby OFF Output Set Output O (not zero) The DC current section consists of a set of range commands which are used in conjunction with the standby and output command.
  • Page 62 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 63: Ac Current Commands

    9000 SERIES OPERATION MANUAL AC Current Commands Function Range Command AC Current 200uA 20mA 200mA Standby Mode Standby ON Standby OFF Output Set Output O (not zero) AC Frequency Fxxxxx E.G. 10kHz = F10000 The AC current section consists of a set of range commands which are used in conjunction with the standby and output command.
  • Page 64 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 65: Resistance Commands

    9000 SERIES OPERATION MANUAL Resistance Commands Function Range Command Resistance 0 Ohms 0.1 Ohms 1 Ohms 10 Ohms 100 Ohms 1 kOhm 10 kOhms 100 kOhms 1 MOhms 10 MOhms 100 MOhms 1G Ohms Standby Mode Standby ON Standby OFF...
  • Page 66 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 67: Capacitance Commands

    9000 SERIES OPERATION MANUAL Capacitance Commands Function Range Command Capacitance 10nF 20nF 50nF 100nF 10uF 100uF Standby Mode Standby ON Standby OFF The capacitance section consists of a set of range commands which are used in conjunction with the standby command. To enable a capacitance to be set, the following command sequence should be used : <RANGE>/<STANDBY CONDITION><CR>...
  • Page 68: Simulated Resistance Commands (Option)

    9000 SERIES OPERATION MANUAL Simulated Resistance Commands (Option) Function Mode Activation Command Simulated Resistance Range Command Output 10 Ohms – 99.9 Ohms Ohms 100 Ohms – 999 Ohms kOhms 1kOhms – 9.99kOhms kOhms 10kOhms – 99.9kOhms kOhms 100kOhms – 999kOhms MOhms 1MOhms –...
  • Page 69: Simulated Capacitance Commands (Option)

    9000 SERIES OPERATION MANUAL Simulated Capacitance Commands (Option) Function Range Command Capacitance 100uF 10mF Standby Mode Standby ON Standby OFF The capacitance section consists of a set of range commands which are used in conjunction with the standby command. To enable a capacitance to be set, the following command sequence should be used : <RANGE>/<STANDBY CONDITION><CR>...
  • Page 70 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 71: Ttl Frequency Commands (Option )

    9000 SERIES OPERATION MANUAL TTL Frequency Commands (Option) Function Mode Activation Command TTL Frequency Range Command 10Hz 100Hz 1kHz 10kHz 20kHz 50kHz 100kHz 1MHz 10MHz Standby Mode Standby ON Standby OFF The TTL frequency section consists of a mode activation command followed by a set of range commands which are used in conjunction with the standby command.
  • Page 72 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 73: Pulse Width Modulation Commands (Option)

    9000 SERIES OPERATION MANUAL Pulse Width Modulation Commands (Option) Function Mode Activation Command Pulse Width Modulation Range Command Standby Mode Standby ON Standby OFF The pulse width modulation section consists of a mode activation command followed by a set of range commands which are used in conjunction with the standby command.
  • Page 74 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 75: Inductance Commands (Option)

    9000 SERIES OPERATION MANUAL Inductance Commands (Option) Function Range Command Inductance 10mH 19mH 29mH 50mH 100mH Standby Mode Standby ON Standby OFF The inductance section consists of a set of range commands which are used in conjunction with the standby command. To enable a inductance to be set, the following command sequence should be used : <RANGE>/<STANDBY CONDITION><CR>...
  • Page 76 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 77: Prt Commands (Option)

    9000 SERIES OPERATION MANUAL PRT Commands (Option) Function Range Command PRT Value -100°C 0°C 30°C 60°C 100°C 200°C 300°C 400°C Standby Mode Standby ON Standby OFF The PRT section consists of a set of range commands which are used in conjunction with the standby command.
  • Page 78 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 79: Thermocouple Simulation Commands (Option)

    9000 SERIES OPERATION MANUAL Thermocouple Simulation Commands (Option) This function requires the use of the optional Thermocouple simulation adapter. This is used in conjunction with the feature connector on the 9000 Series to provide the thermocouple simulation. Function Mode Activation Command...
  • Page 80 9000 SERIES OPERATION MANUAL To enable thermocouple simulation to be set up, the following command sequence should be used : <MODE>/<CJC TYPE>/<THERMO TYPE>/<TEMP VALUE>/ <STANDBY CONDITION><CR> For example, to set the following configuration : • AUTOMATIC COLD JUNCTION COMPENSATION • TYPE R •...
  • Page 81 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 82: Miscellaneous Commands

    9000 SERIES OPERATION MANUAL Miscellaneous Commands Earth Relay (Floating or Grounded negative terminals) Earth Relay Earth Relay On Earth Relay Off The earth relay command allows the negative terminals to either floating or grounded, depending on the command set. This is also indicated in the display of the calibrator by the -VE symbol (top left).
  • Page 83: Oscilloscope Calibration Commands (Option)

    9000 SERIES OPERATION MANUAL Oscilloscope Calibration Commands (Option) This function requires the oscilloscope calibration option to be installed in the 9000 series calibrator - the output will appear on the BNC connector indicated by the green LED. Amplitude Function Mode Activation Command...
  • Page 84 9000 SERIES OPERATION MANUAL To enable an amplitude range to be set up, the following command sequence should be used : <MODE>/<RANGE>/<WAVEFORM>/<STANDBY CONDITION><CR> For example, to set the following configuration : • 1V/Div • Square Wave • Output ON Send the following command sequence : A1/H8/G0/S0<CR>...
  • Page 85 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 86 9000 SERIES OPERATION MANUAL Timebase Function Mode Activation Command Timebase Mode Range Command 5s/DIV 2s/DIV 1s/DIV 500ms/DIV 200ms/DIV 100ms/DIV 50ms/DIV 20ms/DIV 10ms/DIV 5ms/DIV 2ms/DIV 1ms/DIV 500us/DIV 200us/DIV 100us/DIV 50us/DIV 20us/DIV 10us/DIV 5us/DIV 2us/DIV 1us/DIV 500ns/DIV 200ns/DIV 100ns/DIV 50ns/DIV 20ns/DIV The Timebase function consists of the following commands : •...
  • Page 87 9000 SERIES OPERATION MANUAL To enable a timebase range to be set up, the following command sequence should be used : <MODE>/<RANGE>/<STANDBY CONDITION><CR> For example, to set the following configuration : • 1ms/Div • Output ON Send the following command sequence : R61/H11/S0<CR>...
  • Page 88 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 89 9000 SERIES OPERATION MANUAL Bandwidth (Levelled) Sweep Function Mode Activation Command BW Sweep Mode Output Set Output O (not zero) The bandwidth function consists of the following commands : • Bandwidth mode activation • Output value (MHz) • Standby mode command...
  • Page 90 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 91 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 92: Power Calibration Commands (Option)

    9000 SERIES OPERATION MANUAL Power Calibration Commands (Option) This function requires the power calibration option to be installed in the 9000 series calibrator - the output will appear on the voltage and low current terminals and be indicated by the combined illumination of the voltage and low current terminal LEDs.
  • Page 93 9000 SERIES OPERATION MANUAL The power function consists of the following commands : • Power mode activation • Voltage output setting • Current output setting • Phase relationship in degrees • Standby mode command To enable a power output to be set up, the following command sequence should be used : <MODE>/<VOLTAGE>/<CURRENT>/<PHASE>/<STANDBY CONDITION><CR>...
  • Page 94 This functionality is employed within the ProCal calibration software from Transmille to allow safe operation of the calibrator and to ensure the calibrator is returned to a safe state in between test points and at the completion of a test sequence.
  • Page 95: Technical Description

    9000 SERIES OPERATION MANUAL Technical Description General The series 9000 calibrators use the latest in reference, resistor and processor technology designed to minimise cost and size yet maximise performance. The micro processor controls and monitors all functions of the calibrator. Calibration constants are held in non volatile memory allowing the calibration to be performed without removing the covers.
  • Page 96: Construction

    9000 SERIES OPERATION MANUAL Construction The calibrator is constructed in a custom inner enclosure built into a rugged transportable outer enclosure with fan cooling used for the high voltage and high current amplifiers. The calibrator is construction is modular to allow ease of servicing.
  • Page 97: Internal Fuses

    9000 SERIES OPERATION MANUAL Internal Fuses. In normal operation these fuses should never need to be replaced. Only under fault conditions will they require changing. NOTE : To access these fuses it is necessary to dismantle the case which should only be carried out by an engineer.
  • Page 98: Opening The Enclosure

    9000 SERIES OPERATION MANUAL Opening The Enclosure Warning risk of shock. The line power cord must be disconnected before opening the instrument To gain access to the inside remove the six screws from the bottom of the transportable enclosure. Once removed, unscrew the two top bolts to provide a grip for extracting the entire calibrator assembly from the transportable enclosure (this is one self contained assembly).
  • Page 99: Power Supply And Output Switching Board

    9000 SERIES OPERATION MANUAL Power supply and output switching board On this board is the power supply for the calibrator which is a linear design having the inherent advantage of being low in noise. Supply voltages are ± 5 Volts, ±15 Volts, ±35 Volts and unregulated ±9 Volts at 30Amps.
  • Page 100: Main Analogue Amplifier And Feedback Board

    9000 SERIES OPERATION MANUAL Main analogue amplifier and feedback board This board generates AC and DC voltages and currents which are derived from the D/A boards 0 to 10Volt reference. This reference voltage is compared against the output after it is scaled from either a precision resistive divider of switchable gain for voltage ranges or from a set of precision current shunts for the current ranges.
  • Page 101: High Voltage Amplifier And Output

    9000 SERIES OPERATION MANUAL High Voltage Amplifier and Output All outputs above 20 volts use this amplifier. A high power 150W IC amplifier running from 25 volt supplies with thermal and output overload protection output is switched into either a high frequency ferrite transformers or a LF iron laminate transformer depending on the frequency to produce all high voltage outputs.
  • Page 102: Output Currents Sensing & Shunts

    9000 SERIES OPERATION MANUAL Output Currents Sensing & Shunts Six high stability current shunts with low temperature coefficients from 4kOhms to 0.01ohm provide feedback for the current ranges. The lower values are switched using a four wire method for optimum accuracy, there is no provision for the adjustment of the value of the shunts, calibration is performed by the calibration factors.
  • Page 103: Calibration Tutorial

    Calibration Tutorial Getting the best out of the calibrator. The 9000 series are very accurate calibrators producing a very wide range of output signals. To make the best possible use of the range of outputs and to eliminate errors this section details some common sources of errors and offers some techniques to reduce them.
  • Page 104 9000 SERIES OPERATION MANUAL Power line and low frequency Pick up and noise These effects are most noticeable when using high resistance (100kohms and above) and low current. All constant current sources have a very high output impedance which will pick up noise just like the high value resistance . To reduce pickup, use screened leads and try earthing the low side of the calibrator output.
  • Page 105: Calibration And Maintenance

    The user must at all times be adequately protected from electric shock. General The 9000 series calibrators maintenance requirements are listed below. Please note that the calibrator does not require any regular internal servicing or adjustment. 1) Electrical Safety Checks on Line power lead and case...
  • Page 106: Cleaning Of The Fan Duct

    9000 SERIES OPERATION MANUAL Cleaning of the Fan Duct WARNING : Risk of Shock Ensure calibrator is disconnected from line power before proceeding. Fan ducts may be cleaned with brush and vacuum cleaner Cleaning the external case Use a damp cloth with a mild water based cleaner for the outside case and front panel.
  • Page 107: Calibration

    9000 SERIES OPERATION MANUAL Calibration To adjust the 9000 Series calibrator the calibrator can either be connected to a computer via the RS232 Serial interface. Calibration constants stored within the calibrator can then be adjusted using the 9000 Series Virtual Front Panel software.
  • Page 108: Adjustment Overview Using Pc Virtual Front Panel Software

    9000 SERIES OPERATION MANUAL Equipment Required Precision 8 ½ Digital Multimeter. Eg. Transmille 8081, Hewlett Packard HP3458A or Wavetek 1281. Capacitance / Inductance bridge. Eg. Wayne Kerr B905. Frequency counter. Shunt resistors for measurement of 2A and 30A (Not required for Transmille 8081) Low thermal test leads with 4mm plug terminations.
  • Page 109 10) Press the ‘abort’ button to abandon changes for the range being adjusted. 11) Select next range to be adjusted. 12) Close calibration control panel and exit virtual front panel program Comprehensive details of the calibration sequence are available in the 9000 Series Service Manual. TRANSMILLE LTD. Page 108...
  • Page 110 9000 SERIES OPERATION MANUAL Adjustment Overview Using Calibrator Front Panel Controls 1) Allow all equipment to stabilise for at least 4 hours. 2) To select front panel calibration mode on the calibrator Enter Then press Enter The calibrator will produce a 2 second beep to confirm front panel calibration mode is selected 3) Select range &...
  • Page 111: Guarantee And Service

    To obtain repair under this guarantee the purchaser must send the instrument in its original packaging (carriage prepaid) and a description of the fault to Transmille at the address shown below. The instrument will be repaired at the factory and returned to the purchaser, carriage prepaid.
  • Page 112 +44 (0) 1580 890711 Please Fax This Form To : On receipt of this fax Transmille will, on receipt of payment for the calibrator, send details of the security code with details on how to enter this code. TRANSMILLE LTD.