2100 2100 Electrical Test Equipment Calibrator Operation Manual Version 4.00 December 2004 All product names are trademarks of their respective companies...
Page 2
FROM TRANSMILLE ONLYAFTER PAYMENT HAS BEEN RECEIVED. (This code is only entered once in the life of the instrument.) Please contact Transmille or use the form in the back of the manual to obtain the code. Transmille Ltd. Staplehurst , Kent.
Page 3
2100 Operation Manual DECLARATION OF CONFORMITY Manufacturer’s Name: Transmille Ltd. Manufacturer’s Address: Unit 4, Select Business Centre Lodge Road Staplehurst TN12 0QW Declares, that the product Product Name: Electrical Test Calibrator Model Number: 2100 Product Options: This declaration covers all options of the above...
2100 Operation Manual Table of Contents 2100 ELECTRICAL TEST EQUIPMENT CALIBRATOR INTRODUCTION ......6 INSTALLATION & POWER REQUIREMENTS FOR THE 2100 ..........7 PREPARING THE CALIBRATOR FOR USE................... 9 ..........................9 NITIAL NSPECTION ..........................9 HIPPING HECKLIST .................... 9 IFTING AND CARRYING THE CALIBRATOR ......................
Page 5
CALIBRATION AND MAINTENANCE................... 50 ............................50 ENERAL ........................ 50 LECTRICAL AFETY ESTS ......................50 LEANING THE EXTERNAL CASE ........................50 ALIBRATION VERVIEW GUARANTEE AND SERVICE......................52 APPENDIX A....................2100 SPECIFICATIONS APPENDIX B....................APPLICATION NOTES Transmille Ltd. Issue 4.00 Page 5...
A complete electrical test equipment calibration solution from one instrument. Designed to provide an accurate cost effective portable instrument for the calibration of Insulation testers, RCD testers, LOOP testers and PATs the 2100 calibrator can be ProCal combined with the Calibration System to allow automated calibration.
There should only be a small voltage between Phase (Neutral) and Phase (Earth). To keep the LOOP impedance value to a minimum, the mains input to the 2100 is of a direct cable type. This avoids any introduction of unnecessary impedance.
Page 8
Designed for on use in the laboratory or portable on site calibration. The 2100 calibrator is equally suitable for use in the standards laboratory or for on site calibration work. The Small footprint and low weight makes the 2100 calibrator ideal for on site calibration.
2100 Operation Manual Preparing the calibrator for use. Initial Inspection. After shipment the calibrator should be inspected for any signs of external damage. Should external damage be found contact the carrier immediately. Do not connect a damaged instrument to the line power as this may result in internal damage. Please keep the original box which can be used when returning the calibrator for service and re-calibration.
2100 Operation Manual Positioning the Calibrator. The calibrator can be used free standing on a bench or mounted in a standard 19” rack enclosure. The calibrator can be operated at any angle, the two front feet have tilt legs for bench operation. For all installations care must be taken not to cover the ventilation slots underneath.
2100 Operation Manual Setting and checking the Line Voltage. Warning The line power cord must have an earth conductor to avoid risk of shock. This instrument must be correctly earthed. The calibrator has been designed to work from either 100-120 Volt line supply or 200 - 240 Volt line supply.
2100 CALIBRATOR Ver x.x TRANSMILLE LTD. The calibrator has been designed to be powered up continuously, automatically switching the display to power save mode after a pre-set period of time from the last command.
A carry-case is available for regular transportation of the calibrator. Introduction to Operation All functions of the 2100 Calibrator can be controlled from the front panel or controlled remotely by a computer over the RS232 interface. Front Panel Controls and Indicators...
2100 Operation Manual Front panel Keyboard The front panel of the 2100 Calibrator utilises a high quality membrane switch panel with tactile feel buttons and integral display window. The front panel is therefore sealed against the ingress of moisture and dirt enabling the calibrator to be used in working environments without risk of dirt causing early failure of the operating buttons.
2100 Operation Manual The Keyboard is divided into sections to allow rapid operation. The Numeric section allowing values to be entered, Functions keys for RCD, LOOP, PAT, Insulation Resistance, Continuity Resistance and ACV Output Range up and range down keys allows range changing for the currently selected function Left/right arrow keys select the digit to be controlled by the rotary control knob.
2100 Operation Manual Using the Digital Potentiometer A digital potentiometer allows selected digits on the display to be incremented (turning clockwise) or decrement (turning anti-clockwise). Selected digit marker. Cursor Keys can be used to move the position of the digit marker, and increment / decrement the digit.
2100 Operation Manual Terminal status LEDs LED’s above the terminals indicate the active terminal. Active terminals indicated by illuminated LED PAT Test IEC Socket WARNING Dangerous voltage may be present on these terminals. LOOP & RCD TEST IEC Plug WARNING Dangerous voltage may be present on these terminals.
TESTERS for all insulation meter tests. High Value Resistance for Insulation Testing Step 1 Select ‘INS RES’ from the function key section of the 2100 front panel Step 2 Connect the insulation tester to the Black & Red 4mm terminals. Step 3 Enter the required resistance in MΩ...
1mA at the test voltage. Older insulation testers (15 Edition Reg.) only had to produce 0.5mA current at the test voltage. The 2100 should be set to 0.5mA current for these instruments. Very old tester’s may only produce a very small current and the voltage will collapse under any load - these cannot be tested with the 2100.
Page 20
Step 4 Press the test button on the insulation tester to apply the insulation test voltage and read the voltage and current on the 2100 display. To return from the Voltage measurement screen to the insulation resistance select the Back menu item using the Softkey.
2100 is derived from transformer tappings. The output is unregulated and will vary with line voltage and loading. The actual output voltage at any time is measured and displayed by the 2100 which can be compared to the displayed value on the UUT.
The insulation tester must have a guard terminal for this adapter to function correctly Step 1 Select ‘INS RES’ from the function key section of the 2100 front panel Step 2 Connect the 5kV Adapter flying leads to the 2100 Insulation tester terminals...
It is important to understand that an RCD tester does not generate the current, but acts as a resistor allowing current to flow from the live to earth, simulating a fault. The 2100 measures the current flowing back to earth. Transmille Ltd. Issue 4.00...
2100 Operation Manual Calibrating RCD Testers using the 2100 The 2100 has three functions which can be used for complete testing of an RCD tester: 1. RCD Current measurement 2. RCD Trip Time measurement 3. RCD Duration Measurement Use the IEC socket labelled LOOP &...
Some RCD testers may require a settling delay between the application of mains from the 2100 calibrator and pressing of the TEST button on the RCD tester. This may be indicated by a symbol on the display of the RCD tester.
Page 26
SET TIME SET I(X) Step 6 Press the test button on the RCD tester to allow the 2100 to measure the current. Step 7 Read the measured current from the display of the 2100. The polarity and measurement technique will also be indicated on the 2100 screen.
Some RCD testers may require a settling delay between the application of mains from the 2100 calibrator and pressing of the TEST button on the RCD tester. This may be indicated by a symbol on the display of the RCD tester.
Page 28
2100 Operation Manual Step 6 Press the ‘TEST’ button on the RCD tester to allow the 2100 to output the current. Step 7 Read the measured time from the display of the RCD tester. The polarity and measurement technique will also be indicated on the 2100 screen.
Step 4 Press the TEST button on the RCD tester to begin the test. RCD : 100mA 200ms ABORT Step 5 On completion of the measurement, read the value displayed on the 2100 display Transmille Ltd. Issue 4.00 Page 29...
2100 Operation Manual Calibrating Portable Appliance Testers (PATs) The 2100 has four functions which can be used for testing of Portable Appliance Testers (PATs) : 1. Earth Bond Resistance measurement 2. Earth Bond Current measurements 3. Insulation testing 4. Load testing 5.
Read the measured resistance value from the display of the PAT. Notes on the PAT Earth Bond Function The 2100 uses a fixed set of resistors to produce the earth bond resistances. The value displayed by the 2100 is the measured value as determined during the calibration of the 2100.
EARTH BOND CURRENT MEASURED 0.00A CURRENT RANGE - 30A TEST BACK Step 5 Press the ‘SoftKey’ beneath the TEST menu item to allow the 2100 to begin detecting the PAT output current. EARTH BOND CURRENT MEASURED 0.00A CURRENT RANGE - 30A BACK...
Connect the PAT tester 3-pin plug to the PAT TEST IEC socket using using the input adapter lead supplied. Step 2 Select ‘PAT’ from the function key section of the 2100 front panel Step 3 Using the ‘SoftKeys’ beneath the on-screen PAT menu, select ‘NEXT’...
Step 1 Connect the Flash adapter pod to the 2100 back panel socket. Step 2 Select ‘PAT’ from the function key section of the 2100 front panel Step 3 Using the ‘SoftKeys’ beneath the on-screen PAT menu, select ‘NEXT’ PAT TESTING...
Connect the PAT tester 3-pin plug to the PAT TEST IEC socket using using the input adapter lead supplied. Step 2 Select ‘PAT’ from the function key section of the 2100 front panel Step 3 Using the ‘SoftKeys’ beneath the on-screen PAT menu, select ‘NEXT’...
2100 Operation Manual Introduction to LOOP Testers What is the loop ? When an appliance is connected to the mains supply a circuit is made. It completes the loop current flow round the circuit loop from the power station in the live wire through the appliance and then back to the power station in the neutral wire.
Page 37
The value displayed on the tester can then be compared to that of the known value of the resistor. plus the loop resistance of the supply. The 2100 has 8 calibrated resistance values which are non inductive and are able to withstand the 23 Amps needed.
Page 38
(Note this is not the value of the socket the 2100 is plugged in to but the value at the test socket, which is always greater due to the resistance added by the 2100 wiring.) How can I calibrate a Loop Tester at Zero It is desirable when calibrating loop testers to calibrate at as near zero as possible.
Page 39
Remember every switch, fuse socket, even the 2100 will add resistance. If it is required to calibrate at values below that available from of the 2100 then a short length of extension cable of which the resistance of the live + earth conductors have been measured by a 4 wite DMM can be used.
2100 Operation Manual Calibrating LOOP Testers using the 2100 Loop testing using the 2100 is performed using a fixed set of resistance ranges, with auto (optional extra) and manual loop measurement. n Auto Loop Measurement NOTE : Auto loop measurement is an optional extra specified when a 2100 is ordered.
Page 41
If the 2100 is moved to another location or plugged into a different socket, this will have an impact on the loop impedance – the auto loop function will need to be used or the specific socket’s impedance measured and entered manually.
This function can be calibrated with the 2100 by using the LOOP function The standard lead supplied with the 2100 will enable you to simulate Phase (Live) to Earth PSCC. If Phase (Live) to Phase (Neutral) PSCC testing is required, the lead...
Windows HYPER Terminal program, any basic or high level program or from the ProCal98 Calibration System. Programming Commands The 2100 is controlled by a set of simple high level commands. Function Command...
Page 44
2100 Operation Manual PAT : EARTH BOND Select the insulation tester voltage RESISTANCE measurement function Set 0Ω Ω range Set 0.05Ω Ω range Set 0.1Ω Ω range Set 0.22Ω Ω range Set 0.33Ω Ω range Set 0.5Ω Ω range Set 1Ω Ω range Set 5Ω...
Page 45
Transmit output value (internal resistor value + loop resistance) in 1uΩ Ω units (12 Chars) Transmit Mains Voltage in mV (12 Chars) Command 2100 to measure its mains resistance Select manual mains resistance menu Sx.xxx Send manual mains resistance in ohms For use in conjunction with F84 command Set calibration factor (uΩ...
Example command sequences Note : A forward slash can be used to separate commands on a single line. To set up the 2100 to Insulation mode, 1GΩ Ω the following commands would be used : Set to 1GΩ Ω Insulation...
Technical Description General The 2100 calibrator uses the latest in reference, resistor and processor technology designed to minimise cost and size yet maximise performance. The micro processor controls and monitors all functions of the calibrator. Calibration constants are held in non volatile memory allowing the calibration to be performed without removing the covers.
2100 Operation Manual Internal Fuses. In normal operation these fuses should never need to be replaced. Only under fault conditions will they require changing. NOTE : To access these fuses it is necessary to dismantle the case which should only be carried out by an engineer.
2100 Operation Manual Processor Board Plugs into the main PCB and controls all functions within the calibrator. The processor board is a complete working board containing RAM, PROM, Clock, Cal Ram, I/O and RS232. The processor also applies all calibration factors held in RAM.
Calibration Overview The calibration of the 2100 calibrator can be performed covers on. Calibration factors for positive, negative and zero are store in non volatile memory for each range. Values for resistance will only need adjusting after repair.
Page 51
5: To have available the Correct Test leads required 6: To understand the required test specifications. 7: To operate the calibrator at all times within its load and voltage capabilities COMPREHENSIVE CALIBRATION INSTRUCTIONS ARE AVAILABLE IN THE 2100 SERVICE MANUAL. Transmille Ltd. Issue 4.00 Page 51...
To obtain repair under this guarantee the purchaser must send the instrument (carriage prepaid) and a description of the fault to Transmille at the address shown below. The instrument will be repaired at the factory and returned to the purchaser, carriage prepaid.
Page 53
+44 (0) 1580 890711. Please Fax This Form To : On receipt of this fax Transmille will, on receipt of payment for the calibrator, send details of the security code with details on how to enter this code. Transmille Ltd. Issue 4.00...
Page 55
EXTENDED SPECIFICATIONS EXTENDED SPECIFICATIONS TRANSMILLE LTD., UNIT 4 SELECT BUSINESS CENTRE, LODGE ROAD, STAPLEHURST, KENT. TN12 0QW. UK. www.transmille.com : sales@transmille.com : Tel : +44 (0) 1580 890700 : Fax : +44 (0) 1580 890711 Case Colour Matt Dark Grey (RAL7016)
Page 56
DECLARATION OF CONFORMITY Manufacturer’s Name: Transmille Ltd. Manufacturer’s Address: Unit 4, Select Business Centre Lodge Road Staplehurst TN12 0QW Declares, that the product Product Name: Electrical Test Calibrator Model Number: 2100 Product Options: This declaration covers all options of the above product(s)
Page 57
2100 General Specifications 2100 General Specifications TRANSMILLE Solutions In Calibration Warm Up Time Double the time since last used up to 20 minutes maximum Standard Interfaces RS232 Optional Interfaces USB (Universal Serial Bus) Temperature Performance Storage : -5°C to +60°C Operation : 0°C to +50°C...
Page 58
2100 Continuity Testing Specifications 2100 Continuity Testing Specifications TRANSMILLE Solutions In Calibration Continuity Resistance Continuity Resistance Standard Accuracy - Relative to Calibration Standards Specifications Range Type Resolution Accuracy mΩ Ω ± 0.1Ω to 20Ω Continuously 10mΩ Variable ± 100Ω 10mΩ...
Page 59
2100 Insulation Testing Specifications 2100 Insulation Testing Specifications TRANSMILLE Solutions In Calibration Insulation Resistance Insulation Resistance Standard Accuracy - Relative to Calibration Standards Specifications Range Type Resolution Maximum Accuracy Voltage/Power 0Ω to 5MΩ Continuously 10kΩ Variable 1.1kV or 1 Watt 5MΩ...
Page 60
2100 RCD Testing Specifications 2100 RCD Testing Specifications TRANSMILLE Solutions In Calibration RCD Time RCD Time Accuracy - Relative to Calibration Standards Specifications Resolution Range Accuracy 20ms to 5s 10ms ± 0.7ms RCD Current RCD Current Accuracy - Relative to Calibration Standards Specifications...
Page 62
2100 PAT Testing Specifications 2100 PAT Testing Specifications TRANSMILLE Solutions In Calibration PAT Earth Bond Resistance PAT Earth Bond Resistance Standard Accuracy - Relative to Calibration Standards Specifications Nominal Resolution Accuracy mΩ Ω Resistance Values ± 0.05Ω 0.1mΩ ± 0.1Ω...
Page 63
2100 PAT Testing Specifications (Cont'd) 2100 PAT Testing Specifications (Cont'd) TRANSMILLE Solutions In Calibration PAT Leakage Test Voltage PAT Leakage Test Voltage Resolution Range Accuracy (RMS) ± Counts 100V to 300V 0.1V ± PAT Load Testing PAT Load Testing Range...
Page 64
2100 Loop Testing Specifications 2100 Loop Testing Specifications TRANSMILLE Solutions In Calibration Line Voltage Measurement Line Voltage Measurement Resolution Accuracy Range 200V to 260V 0.1V 0.8% ± 6 Counts Due to continuous development specifications may be subject to change. 2100 Extended Specifications Page 9 Mains Voltage Measure : V4.00...
Page 66
Use the Range Up button or turn the digital values (Nominal) potentiometer clockwise to range up to the highest internal resistance value (nominal value 1kΩ 0.05Ω Range up again – the 2100 display will indicate EXT. 0.1Ω RESISTOR 1 is selected, as shown below : 0.22Ω 0.33Ω...
Page 67
2100 through to the front panel. For this reason it is important to use a resistance meter to verify the actual resistance value measured on the 2100 front panel connector from the external resistors used.
Page 68
-there is no ‘peak hold’ of the ground terminal. reading. Adjust the hipot tester output to the required calibration point and record the actual output voltage displayed on the 2100. Application Note Calibrating Breakdown & Hipot Testers Using the 2100 and 2102 V1.00...
Page 69
Phase (Earth). Rewiring Earth to neutral 2100 Automatic Supply Test If no other option is available the 2100 can be re- The 2100 automatically tests the supply when it is wired (internally) to take the earth return current from turned on and will display an error message and the tester back to the neutral.
Page 70
Alternatively if no other option the live to earth loop and have only one load current, exists it is possible to rewire the 2100 to take the usually 26 Amps, others can also measure the Live to earth return to the neutral.
Page 71
Automating Calibration With ProCal known which can then be added to the additional resistance values used for calibration. ProCal can fully control all operations of the 2100 via Correcting For Supply Loop an optically isolated RS232 connection. Using the built in procedure wizards, procedures can be rapidly One of the advantages of the 2100 is the created.
Page 72
The same problem exists for the 2100, which has to detect the start of the fault current. Most testers will start their timing at the next...
Page 73
RCB tester which started at T1. To do this the 2100 automatically corrects its internal counter using the time interval from the FULL trip point to the zero crossing point, This formula can be used in ProCal to automatically i.e.