3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set Table Of Contents SAFETY WARNINGS ..............................3 GENERAL ..................................4 INSTALLING THE 3200 / 2100 CALIBRATION CONTROL PANEL ...............4 CALIBRATING 3200 / 2100 CALIBRATOR - INTRODUCTION................5 INSULATION RESISTANCE FUNCTION........................6 ................................6 ONNECTIONS ..............................6...
Page 3
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set PAT EARTH BOND CURRENT FUNCTION ......................22 ................................22 ONNECTIONS .............................23 ALIBRATION ETHOD .............................23 ERIFICATION OINTS PAT LOAD FUNCTION............................24 ................................24 ONNECTIONS ............................25 ERIFICATION ETHOD .............................25...
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set SAFETY WARNINGS HIGH VOLTAGES CONNECTIONS TO LINE THIS VERIFICATION AND CALIBRATION GUIDE INVOLVES CONNECTIONS TO LINE AND MEASUREMENT OF HIGH VOLTAGES FOR THIS REASON VERIFICATION AND...
The steps detailed below will allow adjustment of a 3200 / 2100 Electrical Test Equipment Calibrator. This requires the use of the 3200 / 2100/3200 Calibration control Panel program supplied and the 3200 / 2100 calibration lead set kit supplied by Transmille.
For Use With Calibration Control Panel & Calibration Lead Set Calibrating 3200 / 2100 calibrator - Introduction The 3200 / 2100 Calibration kit comprises of a set of test leads as follows : In addition a performance verification sheet is provided –...
1. Connect 3200 / 2100 to meter as per diagram using LEAD SET A 2. Select Insulation resistance mode and test points as per verification sheet using the 3200 / 2100 front panel and record measurement obtained using 2-wire connection.
1. Connect 3200 / 2100 to meter as per diagram using LEAD SET A and LEAD SET B 2. Select test point as per verification sheet using the 3200 / 2100 front panel and record measurement obtained using 4-wire connection.
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set Continuity Current Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
For Use With Calibration Control Panel & Calibration Lead Set Calibration Method 1. Connect 3200 / 2100 to meter as per diagram using LEAD SET A 2. Select continuity test current mode on 3200 / 2100 CCP this will set the 3200 / 2100 to 1 Ohm output 3.
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set AC Voltage Source Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
1. Connect 3200 / 2100 to meter as per diagram using LEAD SET A 2. Select ACV Output mode on 3200 / 2100 CCP - this will set the 3200 / 2100 to 100V Output initially 3. Adjust cal factor using CCP software adjustment control...
COMMON SET OF TEST LEADS (RED & BLACK LEADS) IS A COMMON SET FOR USE WITH ANY TYPE OF 2100 / 3200 CALIBRATOR. DEPENDING ON THE CONFIGURATION OF THE 2100 / 3200 CALIBRATOR THE SECOND LEAD SHOULD BE EITHER : Set C2 LEAD MARKED FOR ‘NORMAL’...
Page 14
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set 3200 Verification & Calibration Guide.doc Page 13...
1. Connect 3200 / 2100 to meter as per diagram using LEAD SET C 2. Select LOOP function on 3200 / 2100 and select MANUAL – type in 0.001 then press enter to reset the line impedance compensation to ZERO before calibrating the loop resistance values.
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set RCD (Residual Current Device) Current Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
1. Connect 3200 / 2100 to meter as per diagram using LEAD SET D 2. Select RCD Current mode on 3200 / 2100 CCP this will set the 3200 / 2100 to 10mA Range in calibration mode (continuous reading) 3. Set the current source to the current range full scale 4.
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set RCD (Residual Current Device) Trip Time Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
1. Connect 3200 / 2100 to meter as per diagram using RCD & OSCILLOSCOPE TRIGGER BOX 2. Select RCD Function on 3200 / 2100 – set 10mA / 40ms trip current / time 3. Adjust setting on oscilloscope for storage mode and to use external trigger input Set 5V/Div Amplitude, 5ms Timebase.
1. Connect 3200 / 2100 to meter as per diagram using LEAD SET G 2. Select PAT Insulation Resistance mode and test points as per verification sheet using the 3200 / 2100 front panel and record measurement obtained using 2-wire connection.
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set PAT Earth Bond Resistance Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
For Use With Calibration Control Panel & Calibration Lead Set Calibration Method 1. Connect 3200 / 2100 to meter as per diagram using LEAD SET G 2. Select PAT Earth Bond Resistance mode on 3200 / 2100 CCP this will set the 3200 / 2100 to 0Ohms Output initially 3.
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set PAT Earth Bond Current Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
5s test. Note : To abort the test at any time press 5. When the test has completed check the measured value displayed by the 3200 / 2100. If adjustment is required use the button to automatically calculate the cal factor.
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set PAT Load Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
For Use With Calibration Control Panel & Calibration Lead Set Verification Method 1. Connect 3200 / 2100 to meter as per diagram using LEAD SET H 2. Select PAT mode on 3200 / 2100 using front panel 3. Select the NEXT button to move to the next menu...
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set PAT Leakage Current Function VERIFICATION AND CALIBRATION PROCEDURES ARE FOR USE BY QUALIFIED TECHNICIANS ONLY – OBSERVE ALL PRECAUTIONS DURING CONNECTION AND MEASUREMENT OF LINE HIGH VOLTAGES...
CONNECT PAT TO IEC INLET SOCKET LOAD FLASH BACK LEAKAGE 4. Compare current measured on battery powered meter with value displayed on 3200 / 2100 calibrator Verification Test Points See Performance Verifications Sheet Page 8 3200 Verification & Calibration Guide.doc Page 27...
Page 29
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set APPENDIX A 3200 Verification & Calibration Guide.doc Page 28...
Page 46
3200 / 2100 Verification & Calibration Guide For Use With Calibration Control Panel & Calibration Lead Set APPENDIX B 3200 Verification & Calibration Guide.doc Page 29...
Page 47
EXTENDED SPECIFICATIONS 3200 SERIES ELECTRICAL TEST EQUIPMENT CALIBRATOR...
Page 48
DECLARATION OF CONFORMITY Manufacturer's Name: Transmille Ltd. Manufacturer's Address: Unit 4, Select Business Centre Lodge Road Staplehurst TN12 0QW. United Kingdom. Declares, that the product Product Name: Electrical Test Calibrator Model Number: 3200 Product Options: This declaration covers all options of the above product(s)
Page 49
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: GENERAL SPECIFICATIONS TRANSMILLE LTD. Warm Up Time Double the time since last used up to 20 minutes maximum Standard Interfaces RS232 Optional Interfaces USB (Universal Serial Bus) Temperature Performance Storage : -5°C to +60°C Operation : 0°C to +50°C...
Page 50
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: CONTINUITY SPECIFICATIONS TRANSMILLE LTD. Continuity Resistance* Standard Accuracy - Relative to Calibration Standards Specifications Range Type Resolution Accuracy mΩ ± 0.2Ω to 20Ω Continuously 10mΩ Variable ± 100Ω 10mΩ Fixed ± 1kΩ 10mΩ Fixed ±...
Page 51
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: INSULATION SPECIFICATIONS TRANSMILLE LTD. Insulation Resistance Standard Accuracy - Relative to Calibration Standards Specifications Range Type Resolution Maximum Accuracy Voltage/Power 0Ω to 5MΩ Continuously 10kΩ Variable 1.1kV or 1 Watt 5MΩ to 2GΩ Continuously 10kΩ...
Page 52
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: INSULATION SPECIFICATIONS TRANSMILLE LTD. RCD Time Accuracy - Relative to Calibration Standards Specifications Resolution Range 20ms to 5s 10ms RCD (Residual Current Device) Current Trip Current Range 0.5 to 3000mA Current Multiplier 0.5, 1, 2, 5...
Page 54
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: PAT SPECIFICATIONS TRANSMILLE LTD. All PAT functions are isolated from mains earth to enable calibration of PAT testers which cannot function with connections to ground. PAT Earth Bond Resistance Accuracy - Relative to Calibration Standards Specifications...
Page 55
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: PAT SPECIFICATIONS TRANSMILLE LTD. PAT Leakage Test Voltage Resolution Range Accuracy (RMS) ± Counts 100V to 300V 0.1V ± PAT Load Testing Range Accuracy Ohms ± Short Circuit Open Circuit 440Ω 0.13kW ± PAT Flash Voltage Measurement...
Page 56
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: MAINS V MEASURE SPECIFICATIONS TRANSMILLE LTD. Line Voltage Measurement Resolution Accuracy Range 200V to 260V 0.1V 0.8% ± 6 Counts Due to continuous development specifications may be subject to change. 3200 Extended Specifications Page 8...
Page 57
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR :: BREAKDOWN ADAPTER SPECIFICATIONS TRANSMILLE LTD. AC/DC VOLTAGE MEASUREMENT Resolution Accuracy (1 Year Rel.) Range ± Counts ± 12kV ± AC/DC CURRENT MEASUREMENT Resolution Accuracy (1 Year Rel.) Range ± Counts 200uA 100nA ± ±...
Page 58
EXTENDED SPECIFICATIONS EXTENDED SPECIFICATIONS TRANSMILLE LTD., UNIT 4 SELECT BUSINESS CENTRE, LODGE ROAD, STAPLEHURST, KENT. TN12 0QW. UK. www.transmille.com : sales@transmille.com : Tel : +44 (0) 1580 890700 : Fax : +44 (0) 1580 890711 Case Colour Matt Dark Grey (RAL7016)
Page 59
DECLARATION OF CONFORMITY Manufacturer’s Name: Transmille Ltd. Manufacturer’s Address: Unit 4, Select Business Centre Lodge Road Staplehurst TN12 0QW Declares, that the product Product Name: Electrical Test Calibrator Model Number: 2100 Product Options: This declaration covers all options of the above product(s)
Page 60
2100 General Specifications 2100 General Specifications TRANSMILLE Solutions In Calibration Warm Up Time Double the time since last used up to 20 minutes maximum Standard Interfaces RS232 Optional Interfaces USB (Universal Serial Bus) Temperature Performance Storage : -5°C to +60°C Operation : 0°C to +50°C...
Page 61
2100 Continuity Testing Specifications 2100 Continuity Testing Specifications TRANSMILLE Solutions In Calibration Continuity Resistance Continuity Resistance Standard Accuracy - Relative to Calibration Standards Specifications Range Type Resolution Accuracy mΩ Ω ± 0.1Ω to 20Ω Continuously 10mΩ Variable ± 100Ω 10mΩ...
Page 62
2100 Insulation Testing Specifications 2100 Insulation Testing Specifications TRANSMILLE Solutions In Calibration Insulation Resistance Insulation Resistance Standard Accuracy - Relative to Calibration Standards Specifications Range Type Resolution Maximum Accuracy Voltage/Power 0Ω to 5MΩ Continuously 10kΩ Variable 1.1kV or 1 Watt 5MΩ...
Page 63
2100 RCD Testing Specifications 2100 RCD Testing Specifications TRANSMILLE Solutions In Calibration RCD Time RCD Time Accuracy - Relative to Calibration Standards Specifications Resolution Range Accuracy 20ms to 5s 10ms ± 0.7ms RCD Current RCD Current Accuracy - Relative to Calibration Standards Specifications...
Page 65
2100 PAT Testing Specifications 2100 PAT Testing Specifications TRANSMILLE Solutions In Calibration PAT Earth Bond Resistance PAT Earth Bond Resistance Standard Accuracy - Relative to Calibration Standards Specifications Nominal Resolution Accuracy mΩ Ω Resistance Values ± 0.05Ω 0.1mΩ ± 0.1Ω...
Page 66
2100 PAT Testing Specifications (Cont'd) 2100 PAT Testing Specifications (Cont'd) TRANSMILLE Solutions In Calibration PAT Leakage Test Voltage PAT Leakage Test Voltage Resolution Range Accuracy (RMS) ± Counts 100V to 300V 0.1V ± PAT Load Testing PAT Load Testing Range...
Page 67
2100 Loop Testing Specifications 2100 Loop Testing Specifications TRANSMILLE Solutions In Calibration Line Voltage Measurement Line Voltage Measurement Resolution Accuracy Range 200V to 260V 0.1V 0.8% ± 6 Counts Due to continuous development specifications may be subject to change. 2100 Extended Specifications Page 9 Mains Voltage Measure : V4.00...