  • Page 1 English Operating manual Weather station data logger HD33[L]MT.4 Companies / Brands of GHM www.deltaohm.com Keep for future reference.
  • Page 2: Table Of Contents

    INDEX INTRODUCTION ....................3 DESCRIPTION ....................... 4 TECHNICAL CHARACTERISTICS ................5 INTERNAL CONNECTIONS ..................8 FRONT PANEL ..................... 11 CONFIGURATION MENU ..................12 SIM CARD ......................19 USB CONNECTION ....................20 HD35AP-S SOFTWARE ..................20 10 SMS COMMANDS ....................21 11 TCP/IP CONNECTION VIA MOBILE NETWORK.............
  • Page 3: Introduction

    1 INTRODUCTION The HD33[L]MT.4 data logger is specifically designed for use in weather stations. Many types of sensors can be connected thanks to its standard terminal header inputs. The data logger is equipped with: 4 analog independently configurable inputs (0…50 mV, -50…+50 mV, 0…1 V, 0…10 V, •...
  • Page 4: Description

    2 DESCRIPTION Cable glands, connectors and USB port with mini-USB connector Housing closing hooks Antenna Housing cover LCD (optional) Push-buttons Battery (optional) SIM holder Internal terminal header HD33[L]MT.4 V4.1...
  • Page 5: Technical Characteristics

    3 TECHNICAL CHARACTERISTICS Power supply If the rechargeable battery is used: 18…30 Vdc If the rechargeable battery is not used: 7…30 Vdc without ETHERNET module 12…30 Vdc with ETHERNET module Power consumption @ 12 Vdc < 4 mA without ETHERNET module and with no mobile network activity ∼...
  • Page 6 Measurement characteristics: Temperature (instrument in line with the probe HP3517WTC… or TP350NTC…) Sensor NTC 10 kΩ @ 25 °C Measuring range -40…+105 °C Resolution (of instrument) 0.1 °C Accuracy ± 0.3 °C in the range 0…+70 °C / ± 0.4 °C outside Stability 0.1 °C / year Relative Humidity (instrument in line with the probe HP3517WTC …)
  • Page 7 0/4…20 mA input Shunt resistance Internal (50 Ω Resolution 16 bit Accuracy ± 2 µA Inputs 0…50 mV / -50…50 mV / 0…1 V / 0…10 V Input Resistance 100 M Ω Resolution 16 bit Accuracy ± 0.01% f.s. Inputs for counting the switchings of a voltage-free contact Switching frequency 50 Hz max.
  • Page 8: Internal Connections

    4 INTERNAL CONNECTIONS Analog Analog input 3 input 2 Analog input 1 Analog input 4 Contact input SWIN2 Contact input SWIN1 External power supply RS485 line Battery termination SDI-12 port Alarm output 1 Alarm Serial port output 2 Power supply: To power the data logger with a solar panel, connect the panel to the +VPanel and GND terminals.
  • Page 9 Configuration of inputs is done with the HD35AP-S software. Below are the connections in the various configurations. Potentiometer 2-wire Pt100 / Pt1000 3-wire Pt100 / Pt1000 4-wire Pt100 / Pt1000 Thermocouple 0…50 mV / -50…50 mV 0…1 V 0…10 V Pyrgeometer 0…20 mA / 4…20 mA 4…20 mA...
  • Page 10 SWIN2 contact input: connect the output contact of the sensor between the terminals SWIN2 and GND. The default contact state can be configured: Normally Open (NO) or Normally Closed (NC). The open state is logged as 1, while the closed state is logged as 0. The logged contact state depends on how long the contact remains in the non-default state during the logging interval.
  • Page 11: Front Panel

    5 FRONT PANEL Bicolor LED: red blinking indicates that the instrument is powered, blinks green to signal the mobile network activity. ESC button: exits the selected function. FUNC/ button: in normal operation, it displays the maximum (MAX), the minimum (MIN) and the average (AVG) of the measurements; it scrolls downwards the available options or decreases the set value in the menu.
  • Page 12: Configuration Menu

    6 CONFIGURATION MENU The menu allows displaying the data logger information and changing operation parameters. The menu is structured in levels, with main categories and submenus. To access the menu you need to enter the user password (configurable through the appropriate menu item) or the administrator password (supplied with the system and not editable).
  • Page 13 3) LOW_BATT_PROT: enables or disables the battery protection against excessive discharge. Selecting ON, when the battery voltage falls below the configured threshold (see the following menu item LOW_BATT_THLD_V) the high-consumption activities of the data logger are disabled: mobile communication, communication with digital buses (RS485 and SDI-12), power supply of active sensors, relays.
  • Page 14 5) EML_DATA_TX_MODE: format of the data sent via e-mail. 0 (LOG): only internal LOG format (for database) 1 (CSV): only standard CSV format (for Excel) 2 (LOG+CSV): both internal LOG and standard CSV formats 6) FTP_DATA_TX: enables or disables the periodic sending of data via FTP. Select On to enable the sending of data via FTP.
  • Page 15 18) CURR_RAIN_UP_THLD_unit: higher alarm threshold of the rainfall quantity in the set unit of measurement. 19) THLD_ALRM: enables or disables the buzzer when measurement thresholds are exceeded. 20) EXIT: returns to the main menu. 6) HYST_MENU (hysteresis of the alarm thresholds) Note: the menu items depend on the data logger configuration.
  • Page 16 4) CH4_Input Type_ALRM_DELY: delay for alarm activation of the quantity measured by the analog input 4. 5) ATM_PRES_ALRM_DELY: delay for alarm activation of the atmospheric pressure (optional). 6) BATT_ALRM_DELY: delay for alarm activation of the battery voltage. 7) VOLT_PWR_SPLY_ALRM_DELY: delay for alarm activation of the external power supply.
  • Page 17 9) LOG_MENU (logging) 1) LOG_STAT: enables or disables the logging. 2) LOG_CYCL: choice between cyclical management (the new data overwrite the old ones when the memory is full) or non-cyclical management (logging stops when the memory is full) of the data logger memory. Select YES for the cyclical management. 3) LOG_TIME: choice of logging interval.
  • Page 18 13) CAL_MENU (calibration) – Only available with administrator password Note: the availability of the menu items depends on the data logger configuration. 1) RH_75%_CAL: relative humidity sensor calibration at 75%RH. 2) RH_33%_CAL: relative humidity sensor calibration at 33%RH. 3) RAIN_TIP_mm: tipping bucket rain gauge resolution in mm. 4) RST_ALL_RAIN_CNTS: reset of all the rainfall counters.
  • Page 19: Sim Card

    7 SIM CARD In order to use the mobile network functionalities, a SIM card enabled for data transmission must be inserted into the data logger. The card should be requested to an operator that has an adequate coverage of the mobile network in the place where the data logger will be installed. To insert the card, proceed as follows.
  • Page 20: Usb Connection

    8 USB CONNECTION The data logger can be connected to a PC through the mini-USB connector located at the bottom of the housing. Remove the connector protective cap and connect the CP23 cable. USB connection doesn’t require the installation of drivers: when the data logger is connected to a PC, Windows®...
  • Page 21: Sms Commands

    10 SMS COMMANDS SMS messages containing commands can be sent by a mobile phone to the data logger. The SMS must be sent to the number of the SIM card inserted into the data logger. The following table lists the available commands: Command Description RESET...
  • Page 22 Command Description SMS-DEVICE-ALARM-REPORT Transmits via SMS a report of the measurements selected for SMS alarms SMS-DEVICE-REPORT Transmits via SMS a report of the measurements of the device SMS-HELP Transmits an SMS containing the list of all SMS commands TCP-SERVER-ON Activates a TCP connection with AP acting as a TCP server TCP-SERVER-OFF Deactivates the TCP connection with the device acting as a TCP server TCP-CLIENT-ON...
  • Page 23 recipients " and setting the field " SMS keyword " (see chapter " GSM settings " of the software online help). Example: supposing you entered the string “>>>” in the SMS keyword field and you wish to activate periodic download via e-mail of the measured data with an interval of 1 hour, you will have to send the following text message: >>>...
  • Page 24: Tcp/Ip Connection Via Mobile Network

    11 TCP/IP CONNECTION VIA MOBILE NETWORK It is possible to interact with the data logger by establishing a direct TCP/IP connection via mobile network with a remote PC connected to the Internet. The connection can be of two types: 1) Data Logger = Client , PC = Server The data logger acts as TCP client and requests the connection to the PC, the PC acts as TCP server and waits for the connection request.
  • Page 25: Modbus

    12 MODBUS The complete list of MODBUS registers for “Slave” mode is shown below. Depending on the system configuration, some of the listed registers could not be present if not significant for that particular system (for ex., atmospheric pressure measurement will not be available if it is not measured by the data logger).
  • Page 26 Address Type Coil description Wind speed when the measurement is below the minimum threshold of the sensors: 0=0 m/s, 1= threshold value in m/s Scalar average: the average intensity is calculated as average of intensities. For the calculation of the average direction, also called “prevailing direction”, the velocity versor (unit vector having the same direction of the velocity vector) is considered for each measurement, and the versor coordinates along the measurement axes are calculated, then the average of the coordinates along each axis is calculated.
  • Page 27 Address Type Input Register description alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm. (VWC) in % (x10). OIL VOLUMETRIC WATER CONTENT Soil volumetric water content alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm. in mV (x10). PROBE OUTPUT VWC probe output alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
  • Page 28 Address Type Input Register description (HD51.3D…/HD52.3D… anemometer) in degrees (x10). IND DIRECTION Wind direction (HD51.3D…/HD52.3D… anemometer) alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm. TATE OF THE CONTACT INPUT Contact input alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm. in l/s. Flow (l/s) alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
  • Page 29 Address Type Input Register description Daily evapotranspiration alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm. in W/m ET RADIATION Net radiation alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm. in hPa. ELATIVE PRESSURE Relative pressure alarm: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm. in m (x100).
  • Page 30 Address Type Input Register description 1008 + with 3-wire Pt1000 sensor of channel N in the set measurement unit EMPERATURE 200 x (N-1) (x10). 1009 + Alarm for temperature with 3-wire Pt1000 sensor of channel N: 0=OFF, 1=lower 200 x (N-1) threshold alarm, 2=higher threshold alarm.
  • Page 31 Address Type Input Register description 1042 + Value of quantity associated to channel N if the channel is configured as potentiometric 200 x (N-1) input. 1043 + Alarm for quantity associated to channel N if the channel is configured as potentiometric 200 x (N-1) input: 0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
  • Page 32 Address Type Input Register description 5074 Unit of measurement for : 0=°C, 1=°F. GLOBE THERMOMETER TEMPERATURE 5076 Unit of measurement for WBGT : 0=°C, 1=°F. INDOOR INDEX 5078 Unit of measurement for WBGT : 0=°C, 1=°F. OUTDOOR INDEX 6000 + Unit of measurement for with 2-wire Pt100 sensor of channel N: 0=°C, TEMPERATURE...
  • Page 33 Address Type Input Register description 9004 Measurement unit of the (HD52.3DT…). See TAB 12.1 RAINFALL QUANTITY Resolution of the (HD52.3DT…): RAINFALL QUANTITY 9005 …, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, … 9006 Measurement unit of the . See TAB 12.1 DAILY RAINFALL QUANTITY Resolution of the DAILY RAINFALL QUANTITY 9007...
  • Page 34 Address Type Holding Register description Lower alarm threshold for (VWC) in % (x10). SOIL VOLUMETRIC WATER CONTENT Higher alarm threshold for soil volumetric water content (VWC) in % (x10). Lower alarm threshold for VWC in mV (x10). PROBE OUTPUT Higher alarm threshold for VWC probe output in mV (x10). (cup anemometer) lower alarm threshold in the set measurement unit (the IND SPEED multiplier depends on the set unit).
  • Page 35 Address Type Holding Register description Higher alarm threshold for soil volumetric water content in % (x10) – channel 2. Lower alarm threshold for VWC in mV (x10) – channel 2. PROBE OUTPUT Higher alarm threshold for VWC probe output in mV (x10) – channel 2. Lower alarm threshold for (VWC) in % (x10) –...
  • Page 36 Address Type Holding Register description UVA irradiance higher alarm threshold in W/m (x100). lower alarm threshold in % (x100). ELATIVE UMIDITY Relative Humidity higher alarm threshold in % (x100). lower alarm threshold in mm/h (x10). AXIMUM RAIN RATE Maximum rain rate higher alarm threshold in mm/h (x10). lower alarm threshold in % (x10).
  • Page 37 Address Type Holding Register description 1030 + Channel N lower alarm threshold in mV (x100). Only if channel N is configured as 200 x (N-1) 0…50 mV input. 1031 + Channel N higher alarm threshold in mV (x100). Only if channel N is configured as 200 x (N-1) 0…50 mV input.
  • Page 38 Address Type Holding Register description 4010 & Rainfall quantity (HD52.3DT…) higher alarm threshold in the set measurement unit. 4011 Register 4009 contains the most significant bits. 4012 & lower alarm threshold in the set measurement unit. Register AILY RAINFALL QUANTITY 4013 4013 contains the most significant bits.
  • Page 39 Address Type Holding Register description 20072 User code with ASCII codification of measurement #7. to 20083 Available for models with more measurements of the same type. 20084 User code with ASCII codification of measurement #8. to 20095 Available for models with more measurements of the same type. 20096 User code with ASCII codification of measurement #9.
  • Page 40: Storage Of Instruments

    13 STORAGE OF INSTRUMENTS Storage conditions of the instruments: Temperature: -40…+70 °C. • Humidity: less than 90 %RH no condensation. • For storage, avoid places where: • There is a high level of humidity; • Instruments are exposed to direct sun radiation; •...
  • Page 41: Ordering Codes

    15 ORDERING CODES HD33[L]MT…. Data logger for weather station with mobile communication module. Stores measurements in the internal memory. Transmits the acquired data via FTP, via e-mail or to an HTTP server (Cloud). Optional LCD Display. SDI-12 and Master or Slave RS485 MODBUS-RTU connection. Connection to ETHERNET network with MODBUS TCP/IP protocol through optional module.
  • Page 42 Approvals HD33[L]MT.4 contains LTE module FCC ID: XMR201903EG25G IC ID: 10224A-201903EG25G ANATEL: 02828-19-07968 TELEC certified RF module: [R] 201-190133 201-190133 AD190040201 HD33[L]MT.4 V4.1...
  • Page 44 G U A R A N T E E TERMS OF GUARANTEE All DELTA OHM instruments are subject to accurate testing, and are guaranteed for 24 months from the date of purchase. DELTA OHM will repair or replace free of charge the parts that, within the warranty period, shall be deemed non efficient according to its own judgement.
  • Page 45 GHM GROUP – Delta OHM | Delta Ohm S.r.l. a socio unico Via Marconi 5 | 35030 Caselle di Selvazzano | Padova | ITALY Phone +39 049 8977150 | Fax +39 049 635596 www.deltaohm.com | info@deltaohm.com The quality level of our instruments is the result of the constant development of the product. This may produce some differences between the information written in this manual and the instrument you have purchased.
  • Page 46 GHM GROUP – Delta OHM | Delta Ohm S.r.l. a socio unico Via Marconi 5 | 35030 Caselle di Selvazzano | Padova | ITALY Phone +39 049 8977150 | Fax +39 049 635596 www.deltaohm.com | info@deltaohm.com V4.1 09/12/2020...