/flicker emissions EN
61000-3-3
Guidance and manufacturer's declaration - electromagnetic
immunity - for all equipment and systems
Guidance and manufacturer's declaration — electromagnetic immunity
308nm Excimer System is intended for use in the electromagnetic environment
specified below. The customer or the user of 308nm Excimer System should assure that
it is used in such an environment.
Immunity
EN 60601 test
test
Electrostatic
discharge
± 8 kV contact
(ESD)
EN
± 15kV air
61000-4-2
Electrostatic
±2kV for power
transient /
supply lines
burst
±1kV for
EN
input/output lines
61000-4-4
Surge
±1 kV differential
EN
mode
Compliance level
level
± 8 kV contact
± 15kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV differential
mode
buildings used for domestic purposes.
environment-guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
90