KN-4006 series UV Radiation Treatment System Preface Dear users, first of all, thanks for your trust and using KN-4006 series UV radiation treatment system manufactured by our company. Please read this manual and attached documentations carefully before your first installation and using this system.
KN-4006 series UV Radiation Treatment System Important Notice If you have any problem or need help in using, please contact our technical service center for help in time. We will give you technical support or arrange professional technical expert for service at the first time.
KN-4006 series UV Radiation Treatment System Content CHAPTER I SAFETY REQUIREMENTS AND PRECAUTIONS ..... 1 1.1 S ................1 AFETY EQUIREMENTS 1.2 N ....................... 2 OTES 1.3 N ................4 OTES FOR HOME USE CHAPTER II SUMMARY ................6 2.1 F ................
Page 4
KN-4006 series UV Radiation Treatment System APPENDIX C MPD / MED DETERMINATION ......... 41 APPENDIX D RADIATION DOSAGE/INTENSITY/TIME CALCULATION METHOD ................45 APPENDIX E PATIENTS RECORD ............67 APPENDIX F DECLARATION OF THE EUT ........... 68 GUIDANCE AND MANUFACTURER´S DECLARATION –...
KN-4006 series UV Radiation Treatment System Chapter I safety requirements and precautions (Please read before using equipment in this section!) Ø over exposure will hurt the eyes and skin, in serious cases may even lead to diseases such as cataracts or skin...
KN-4006 series UV Radiation Treatment System When you use the equipment for treatment, please strictly confirm the indications, contraindications and possible adverse reactions. In order to avoid circle danger, assure signal quality. The instrument is connected to only with the protection of the...
Page 7
KN-4006 series UV Radiation Treatment System Prohibit to put it in anesthetic inflammable and explosive environment, in case of fire or explosion. Prohibit open Instrument shell without authorization, or may have get an electric shock risk. To Instrument maintenance or upgrade must be after the company training or authorized service personnel.
KN-4006 series UV Radiation Treatment System harm the body. Decrease or increase radiation dosage according to patient’s erythema raising or reducing. In order to ensure the equipment running safety, please use the replacement parts, accessories provided or designated by Kernel.
Page 9
KN-4006 series UV Radiation Treatment System the treatment plans. Patients should be used on theirs own after controlling the treatment method, the relevant matters needing attention, and to ensure that can be used in right way. If any unusual circumstances occurred during treatment, promptly contact the doctor.
KN-4006 series UV Radiation Treatment System CHAPTER II SUMMARY UV phototherapy was originated from 20s of 21 century. With development of science & technology, the technology of artificial light source has been developed swiftly. And UVA, UVB methods became the most effective way in treating many kinds of dermatosis in European and American countries.
Page 11
KN-4006 series UV Radiation Treatment System Applicable Place For Hospital,clinic use and home use. Contraindication Contraindication absolutely Xeroderma pigmentosum; Salient photosensitization dermatosis; Schematized lupus erythematosus; Basal cell nevus syndrome; Lactation period woman; Pregnant period woman. Relatively Contraindication ( Can be treated but doctors should put more attention on them during treatment period )...
KN-4006 series UV Radiation Treatment System Specification l Use Philips special UV lamp as light source, with best stability and long life; l Irradiator angle can be adjusted, or operated by hand, and irradiation place is flexible . l Use simply, patients can bring and cure at home.
Page 13
KN-4006 series UV Radiation Treatment System Irradiation Irradiation device device Timer Timer control control circuit circuit Figure 2-1 KN-4006A/KN-4006B Figure 2-2 KN-4006AL/KN-4006L Lamp pipe Pic 2-3 Back effect fuse grounding Power socket Power Figure 2-4 Right port Figure 2-5 Left port...
KN-4006 series UV Radiation Treatment System 2.3 External mark declaration Attention! Please read attached catalogue Attention ultraviolet radiation radioprotection Need to wear goggle ׀ Power supply on Power supply off Iso-electric level Refer to user’s manual Recycling symbol mark 2.4 Parameters...
Page 15
KN-4006 series UV Radiation Treatment System Main Technical parameters Table 2-1 Product model comparison Spectrum scope Quantity voice Parameter (nm) alerts Model KN-4006A 350~400 KN-4006B 311~312 350~400 KN-4006AL KN-4006BL 311~312 l Working voltage : AC 230V±10%,50Hz±2% l Power rating : 60VA l Fuse protector :...
Page 16
KN-4006 series UV Radiation Treatment System Quantity type KN-4006A KN-4006B ( mw/cm ) KN-4006AL KN-4006BL Wave band type 1~50 0.3~20 l Radiation time: 1. UVA setting scope 0~30min;testing accuracy<±1%。 2. UVB setting scope 0~30min;testing accuracy<±1%。 3. Setting accuracy:1s。...
KN-4006 series UV Radiation Treatment System CHAPTER III PRODUCT CONFIGURATION AND WORKING THEORY 3.1 Checking before installation l Take out the equipment from the box, place it in safe, stable location. Check the accessories according to the packing list. 3.2 Power supply connection...
Page 18
KN-4006 series UV Radiation Treatment System less than 0.1Ω, cut area should not be less than 2.5mm , the length should be more than 3.5m. NOTICE: —— before connecting the power supply, make sure the supplied voltage is the same as the equipment requirement.
KN-4006 series UV Radiation Treatment System CHAPTER IV OPERATION 4.1 Preparation before treatment 4.1.1 Preparation for treatment Dermatologist: make treatment plans according to the patient and determine treatment site, initial dose, treatment and interval Patient: —— Clean treatment part and remove all the cosmetics ——...
KN-4006 series UV Radiation Treatment System specific trait of the patient (skin pigment and external factors that impact body sensitivity), then adjust dose in compliance with the response after exposure. MPD or MED test: Because of individual sensitivity differences to UV rays, so make the biological dose (MPD or MED) as the unit of UV treatment.
KN-4006 series UV Radiation Treatment System ´ dosage [J/cm 1000 radiation time(s)= .Setting the radiation time intensity according to the requirement; Patient: Expose treatment part to the UV lamp in the distance of 3cm, as Figure 4-1, UV LAMP SKIN...
Page 22
KN-4006 series UV Radiation Treatment System taking treatment process as a process of learning. Treatment should be performed under the guidance of a doctor, doctor should implement tracking to monitor on a regular basis. UV protective goggles: Ensure patient wear UV protective goggles before treatment.
Page 23
KN-4006 series UV Radiation Treatment System than other parts of the body, so the dose could be increased accordingly during irradiation. Don’t stop treatment until complete remission or no further improvement. Facial skin is more sensitive to UV radiation than other parts of the body, when necessary, cover it or use an anti- light agent.
Page 24
KN-4006 series UV Radiation Treatment System avoid daily continuous treatment. Adverse reactions: adverse reactions after phototherapy are almost the same reaction as overexposure to the sun, including erythema, edema, occasional blisters. Once there is adverse reaction, adjust dose or stop...
KN-4006 series UV Radiation Treatment System 4.2 KN-4006AL/KN-4006BL Software operation Operation panel LCD screen SET/OK PAUSE/STOP START Figure 4-2 operation panel Display interface System time Irradiation time UV output mark Status mark Figure 4-3 Display interface...
Page 26
KN-4006 series UV Radiation Treatment System Status mark Speaker open Speaker off Start Pause Stop Figure 4-4 Status mark Ø UVA/UVB status Show lamp type. UVA lamp is in use showing mark; UVB shows mark. Ø Start “ ” mar k flashing shows UV is output, and mark is on.
Page 27
KN-4006 series UV Radiation Treatment System Setup Min:Sec Figure Pic 4-5irradiation time Figure 4-6 Speaker status Year-month-date Hour-Min-sec Figure 4-7 System time Press ” /OK” key to enter the setup of irradiation time, speaker on/off and system time, and the edit light is flashing.
Page 28
KN-4006 series UV Radiation Treatment System open it. Then press key to set the system time. 3. System time setting is the same as irradiation time setting way. 4. Finishing the function set, press “ /Ok” to end edit mode.
Page 29
KN-4006 series UV Radiation Treatment System In the pause mode, if press STOP button again, treatment will be finished and countdown is cleared. Saving setup In stop mode, if the last irradiation time is needed, press the button, and the last setup time will be displayed on the "Timer".
KN-4006 series UV Radiation Treatment System 4.3 KN-4006A/KN-4006B Soft operation Operation panel Function Display screen Figure 4-8 operation panel setup Press“MINUTE”key to adjust the minute,press it once, and add to one minute. Press more, more minutes. Press “SECOND” key to adjust the second,press it once,...
Page 31
KN-4006 series UV Radiation Treatment System key, ”beep” sound will be heart before working, then the count down time starts. At the last 10 seconds will have a warning sound。 Ø pause treatment press“START/STOP” to pause treatment,lamp puts out, countdown pause,press“START/STOP”key to recover the treatment process,lamp lighten,continue the count down time。...
KN-4006 series UV Radiation Treatment System CHAPTER V MAINTENACE&MAINTAIN In order to ensure equipment’s normal use, extend the equipment life, users should notice Maintenance & Maintain. 5.1 Maintenance of equipment and accessories In order to ensure normal and safe work, every 6 months,...
KN-4006 series UV Radiation Treatment System hospital in accordance with the provisions of calibration cycle, when the calibration cycle ends, please contact the manufacturer. Recommend that users inspect equipment and accessories in their daily use in order to receive appropriate treatment.
KN-4006 series UV Radiation Treatment System agent abrasive materials. 5. When the plug is occasional wet, please use distilled or deionized water rinse, and then put it at 40 ℃ to 80 ℃ in the dry environment for at least one hour.
KN-4006 series UV Radiation Treatment System Transport and storage environmental conditions: Ambient temperature: -40 ~ 55 ℃ Atmospheric pressure: 500 ~ 1060hpa Relative humidity: ≤ 95% 5.5 Lamp Replacement UVA, UVB lamp was produced by Philips, the power is 9W.
Page 36
KN-4006 series UV Radiation Treatment System Warning: Please cut off the power when replace the lamps. Remove the irradiation device. Remove the fixed light buckle fastening screws, remove the lamp buckle. screw Lamp Buckle Figure 5-1 Figure 5-2 Slide the lamp to the left to separate it from the lamp holder, hold the end of the lamp, uplift and pull it out gently.
KN-4006 series UV Radiation Treatment System Figure 5-4 5.6 Fuse replacement WARNING: Turn off the power and disconnect the power cord before Replacement the fuse. Figure 5-5 Figure 5-6 Figure 5-7 Figure 5-5, demount the fuse using a screwdriver counterclockwise, and remove the fuse, as shown in Figure 5-6.
KN-4006 series UV Radiation Treatment System Chapter VI Common fault analysis and elimination Instrument common failure analysis and elimination method see table 6-1 content. If you can't tell or resolve instrument malfunction, please do not hesitate to contact my company after-sales service center.
Fault from incorrect operation which are shown in the manual. e. Fault from man-made cause which damnify equipment and accessories. f. Fault from self-repair without Kernel company’s approval. g. Fault from that local voltage is not stable. h. Fault from fire, disaster and earthquake.
Page 40
KN-4006 series UV Radiation Treatment System Contact information: Company Name: Xuzhou Kernel Medical Equipment Co., Ltd. Company Address: Kernel Mansion, Economic Development District, Xuzhou City, Jiangsu Province, China Zip Code: 221004 Tel:+86(516)87732209 Fax: +86(516)87732210 Web: www.kernelmed.com www.kerneluvb.com Email: sales1@kernelmed.com EU representative:...
KN-4006 series UV Radiation Treatment System Appendix A radiation intensity The radiation intensity will be recorded in Table A-1. Table A-1 irradiation intensity setting Wave band type Radiation Intensity ( mW/cm² ) Attention: After a period of the usage of lamp, the value of exposure...
KN-4006 series UV Radiation Treatment System Appendix B UV Phototherapy Doses Reference Table 1. Skin Type According to regional differences and population differences, human skin can be broadly divided into six kinds (I ~ VI). I ~ IV by asking patients’ skin reaction after 30 minutes sun exposure at noon of early summer to determine the skin type, V and VI skin types depend on the skin color (see Table B-1).
Page 43
KN-4006 series UV Radiation Treatment System PUVA treatment: UVA Radiation photosensitizer treatment. Table B-2 PUVA Dose The Initial Dose Increasing Dose Maximum Dose Skin Type ( J/cm² ) ( J/cm² ) ( J/cm² ) UVA treatment: only UVA Radiation Treatment.
Page 44
KN-4006 series UV Radiation Treatment System Table B-4 Narrow Spectrum UVB Dose The Initial Dose Increasing Dose Maximum Dose Skin Type ( J/cm² ) ( J/cm² ) ( J/cm² ) 0.05 0.05 0.10 0.10 0.15 0.15 Notice: - Although the method that determine radiation dosage by skin type is simple, but not as accurate as MPD or MED.
KN-4006 series UV Radiation Treatment System Appendix C MPD / MED Determination Please confirm the patient's skin type according to appendix B B-1. Select the test area. Test area can be flexor side arm, abdomen, back, thighs (see photo) and other non-lesions non-exposed parts, the surrounding skin should be no deformity, ulceration and pigmentation.
Page 46
KN-4006 series UV Radiation Treatment System upper arm extensor side of the forearm extensor side of these special parts of . when testing, each hole tested dose values in Table C-2 and C-4 table doses given in the remaining parts of the test in accordance with the normal dose values in Table C-1 and C-3 in the table given in the test.
Page 47
KN-4006 series UV Radiation Treatment System Table C-2 special parts MPD test dose values corresponding to each hole Dose Skin Type ( J/cm ) Ⅰ Ⅱ Ⅲ Ⅳ Ⅴ Ⅵ Sequence 14.5 2.88 11.6 2.16 1.44 0.72 Table C-3 normal part of MED test dose values corresponding to each hole...
Page 48
KN-4006 series UV Radiation Treatment System Example: to test MED, if the patient's skin type is Ⅳ , Test site is the upper arm flexor side, look for it from C-3: the first hole's radiate dosage is 1.2J/cm , The second is 0.96J/cm , the third is 0.72J/cm...
KN-4006 series UV Radiation Treatment System Appendix D Radiation Dosage/Intensity/Time calculation method Method 1 : According to the table Users can find the radiation time directly from Appendix D.Common dosage/intensity/time calculation value are in below table.The first line is radiation intensity ,the unit is mW/cm²;the first column is radiation dosage,the unit is J/cm ;the values in...
Page 51
KN-4006 series UV Radiation Treatment System Table D-2 General Parameter List 2 (Radiation Intensity 4.0~6.8mW/ cm²) Table D-3 General Parameter List 3 (Radiation Intensity 7.0~9.8mW/ cm²) Table D-4 General Parameter List 4 (Radiation Intensity 10.0~12.8mW/ cm²) Table D-5 General Parameter List 5 (Radiation Intensity 13.0~15.8mW/ cm²)
Page 52
KN-4006 series UV Radiation Treatment System Table D-1 General Parameter List 1 (Radiation Intensity 1.0 ~ 3.8mW/ cm²) Radiation Intensity ( mW/cm² )↓ Radiation time 01.00 01.20 01.40 01.60 01.80 02.00 02.20 02.40 02.60 02.80 03.00 03.20 03.40 03.60 03.80 (Min:Sec) ↘...
KN-4006 series UV Radiation Treatment System Appendix E Patients Record No.: Name: Age: Prescriptio Radiation time doctor Irradiatio Examinatio Cumulative Skin n dose exposure time Date Reviews (minutes / comment eaction ( J/cm2 ) seconds) site dates (time / min) Record the log will help patients to help doctors keep abreast of the patient attending the course of treatment, to provide reference for the next treatment.
KN-4006 series UV Radiation Treatment System Appendix F Declaration of the EUT Guidance and manufacturer´s declaration – electromagnetic emission –for all EQUIPMENT AND SYSTEMS Guidance and manufacturer´s declaration – electromagnetic emission The KN-4000 series UV Radiation Treatment System is intended for use in the electromagnetic environment specified below.
Page 73
KN-4006 series UV Radiation Treatment System EN 61000-3-3 Guidance and manufacturer's declaration – electromagnetic immunity –for all EQUIPMENT and SYSTEMS Guidance and manufacturer´s declaration – electromagnetic immunity The KN-4000 series UV Radiation Treatment System is intended for use in the electromagnetic environment specified below.
Page 74
KN-4006 series UV Radiation Treatment System continued table : Immunity test EN 60601 test level Compliance level Electromagnetic environment-guidance < 5 % U < 5 % U (>95 % dip in U (>95 % dip in U for 0.5 cycle for 0.5 cycle...
Page 75
KN-4006 series UV Radiation Treatment System Guidance and manufacturer´s declaration – electromagnetic immunity –for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING Guidance and manufacturer´s declaration – electromagnetic immunity The KN-4000 series UV Radiation Treatment System is intended for use in the electromagnetic environment specified below.
Page 76
KN-4006 series UV Radiation Treatment System continued table : Immunity test EN 60601 test level Compliance level Electromagnetic environment - guidance where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Page 77
KN-4006 series UV Radiation Treatment System Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM - for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and The KN-4000 series UV Radiation Treatment System The KN-4000 series UV Radiation Treatment System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.