61000-4-5
±2 kV common
mode
< 5 % UT
Voltage dips,
(>95 % dip in UT )
short
for 0.5 cycle
interruptions
40 % UT
and voltage
(60 % dip in UT )
variations on
for 5 cycles
power
70 % UT
supply input
(30 % dip in UT )
lines
for 25 cycles
< 5 % UT
EN
(>95 % dip in UT )
61000-4-11
for 5 sec
Power
frequency
(50/60 Hz)
magnetic
30A/m
field
EN
61000-4-8
NOTE:UT is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic
immunity for equipment and system that are not
± 2 kV common
mode
< 5 % UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
30A/m
91
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
308nm Excimer System
requires continued
operation during power
mains interruptions, it is
recommended that 308nm
Excimer System be powered
from an uninterruptible
power supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.