Appendix A Irradiation intensity
The nominal values of the irradiation intensity before leaving the
factory are recorded in the table below.
Table A-1 Irradiation intensity nominal value
Band type
Irradiation intensity (mW/cm
)
2
UVB
Note:
The working life of the light source is not less than 500 hours.
After a period of use, the irradiation intensity value will be
correspondingly attenuated. Irradiation intensity values should be
adjusted in time according to the use of the light source (for example:
after using for every 100 hours or less, when the intensity is significantly
attenuated, after the replacement of the light source, or other conditions
that may cause changes of the irradiation intensity value). The equipment
with intensity acquisition function can detect the irradiation intensity
value automatically. If the accuracy of the irradiation intensity value can
not be determined by the equipment without intensity acquisition
77