Advertisement

Quick Links

JPD-FR403
0482
Infrared Thermometer
Instruction Manual
Shenzhen Jumper Medical Equipment Co., Ltd
Manual Version: 1.0
Date of Issue: 2021/05

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the JPD-FR403 and is the answer not in the manual?

Questions and answers

Summary of Contents for Jumper JPD-FR403

  • Page 1 JPD-FR403 0482 Infrared Thermometer Instruction Manual Shenzhen Jumper Medical Equipment Co., Ltd Manual Version: 1.0 Date of Issue: 2021/05...
  • Page 2 No part of this document may be changed, excerpted, copied, reproduced, or imitated in any form or by any means without the prior consent of Shenzhen Jumper Medical Equipment Co., Ltd.
  • Page 3: Table Of Contents

    Contents Unpacking Check ..................1 Package Contents ..................1 Safety Precautions ..................1 Warning ....................... 3 Symbols ....................... 5 Body Temperature Basics ................6 Product Description ..................7 Features......................8 Product Structure ..................10 Display Description ................... 10 Sounds and Backlight Color Instructions........... 11 Display and Operating Instructions ............
  • Page 4: Unpacking Check

    In case of any damage or operation problem, please contact the dealer or contact Jumper directly. You will need the following information when making your claim: device model, serial number, purchase date, and your contact information and address.
  • Page 5  - Make sure that the thermometer is not exposed to lint, dust, light (including sunlight), etc. - Please note the effects of degraded sensors that can degrade performance or cause other problems. - Make sure that the thermometer is not exposed to pets, pests. ...
  • Page 6: Warning

     Do not touch the tip of the temperature probe, on which a precise temperature sensor resides.  Keep the temperature probe clean to make sure accurate readings.  Before measuring the temperature from the ear canal, clean the earwax, if any. ...
  • Page 7 Do not charge an alkaline dry-cell battery or throw it in fire. Otherwise, the battery may explode. Do not disassemble the thermometer or attempt to repair it. Otherwise, the thermometer may be damaged permanently. Do not take temperature measurements on body parts other than forehead and ears.
  • Page 8: Symbols

    equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -- Reorient or relocate the receiving antenna. -- Increase the separation between the equipment and receiver. -- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Page 9: Body Temperature Basics

    Symbol Description Information about the manufacturer. Date of manufacture. Consult the instructions for use. This product complies with the MDD93/42/EEC requirements. Waste electrical materials should be sent to a dedicated collection point for recycling. IP22 Degree of protection against the Ingress of water. Authorized European Representative A personal injury or damage to the thermometer Warning...
  • Page 10: Product Description

    1°C (1.8°F) each day. Product Description Overview Infrared Thermometer JPD-FR403 measures the body temperature based on the infrared energy emitted from the eardrum or the forehead. Users can quickly get measurement results after positioning properly the temperature...
  • Page 11: Features

    Intended use The JPD-FR403 Dual Mode Digital Infrared Thermometer is intended for the measurement of human body temperatures. The forehead mode is indicated for use by people of all ages and the eardrum mode is indicated for use by people above three months old.
  • Page 12 ● Passive infrared receiving technology. Easy operation ● Ergonomic design ● One button measurement Quick measurement ● 1-second measurement High accuracy ● Advanced infrared temperature sensor, with high sensitivity ● High accuracy with automatic temperature calibration Diverse functions ● 20 readings recall ●...
  • Page 13: Product Structure

    Product Structure (1) LCD display screen (2) Power button / Measure button (3) Memory button / Sound switch (4) Mode button (Adult mode / Child mode/ Object) (5) Probe (take off the cover when measuring the ear temperature) (applied part) (6) Unit switch button (˚C / ˚F) (7) Probe cover (Automatic switch among object, forehead and ear modes, by removing the probe cover to measure ear temperature and placing back...
  • Page 14: Sounds And Backlight Color Instructions

    Sounds and Backlight Color Instructions Range Sounds Backlight Forehead temperature (Adult / Child) 22.0℃-37.5℃/71.6℉-99.5℉ A long beep green 37.6℃-43℃/99.6℉-109.4℉ 3 short double beeps Ear temperature (Adult / Child) 34.0℃-37.5℃/93.2℉-99.5℉ A long beep green 37.6℃-43.0℃/99.6℉-109.4℉ 3 short double beeps Object temperature 0℃-100℃/32.0℉-212℉...
  • Page 15 Operating Instructions Sound and Screen Display Backlight Color Displayed State Take off the probe cover, press and release the Power button for 1 second to power on the thermometer. The symbol “ ” is displayed on the screen. See the table in the “Sounds and Switch to the adult or child backlight color...
  • Page 16 Operating Instructions Sound and Screen Display Backlight Color Displayed State Put the cover on the probe, press and release the Power button for 1 second to power on the thermometer .The “Head” symbol is displayed on the screen. See the table in the Switch to the adult or child “Sounds and measurement mode by pressing the...
  • Page 17 Operating Instructions Sound and Screen Display Backlight Color Displayed State Put the cover on the probe, switch to the object measurement mode by pressing the Mode button.The See the table in the thermometer enters the Object “Sounds and mode. The “House” symbol backlight color displayed on the screen.
  • Page 18 Operating Instructions Sound and Screen Display Backlight Color Displayed State Recall 20 memories In a power-on state, press the Memory button enter the memory mode. When the Memory button is released, 01 will be shown, followed by the recorded reading. Press the Memory button again for the next recorded data.
  • Page 19 Operating Instructions Sound and Screen Display Backlight Color Displayed State The display is as shown , when When the power is there is no more data checked while turned back on, recalling memories. Remove 2 dry there are a long batteries and after 10 seconds beep and a green re-install the power to clear all...
  • Page 20 Operating Instructions Sound and Screen Display Backlight Color Displayed State When putting on the probe cover, press the Mode button to switch between object and forehead Silent, the modes. Ear mode (adult/child) will backlight is green. be activated by removing the probe cover.
  • Page 21: Measuring Ear Temperature

    Operating Instructions Sound and Screen Display Backlight Color Displayed State Error information & low battery The ambient temperature is higher A long beeps,the than 40.0ºC (104.0ºF) or lower than backlight is red. 10.0ºC (50.0ºF). An error occurs when data is being read from or written to the memory, A long beeps,the or the temperature correction is not...
  • Page 22 ear temperature. 3. Press the Power button to power on the thermometer. 4. The “Ear” symbol is displayed on the screen. 5. Switch between “adult” and “child” modes by pressing Mode button. symbol shows up in “child” mode. 6. Insert the temperature probe into the ear canal. 7.
  • Page 23: Measuring Forehead Temperature

    back to make sure the ear canal is straight, so that the temperature probe can receive an infrared ray from the eardrum. Be careful when taking temperature on a child, whose ear canal is small. Measuring Forehead Temperature Put the cover on the probe of thermometer. Press the Power button to power on the thermometer.
  • Page 24: Measuring Object Temperature

    Measuring Object Temperature Put the cover on the probe of thermometer. Press the Power button to power on the thermometer. Press the Mode button, the thermometer enters the Object mode. The “ ” symbol is displayed on the screen. Point the thermometer probe to the center of the object, about “...
  • Page 25: Replacing Batteries

    It is dangerous to make a self-diagnosis or self-treatment based on the obtained measurement results. For such purposes, please consult a doctor. Notes: (1) The thermometer is suitable for an indoor environment without strong air convection (for example, winds from a fan, an air-conditioner, or a heater) between the thermometer and the person.
  • Page 26: Cleaning And Disinfection

    Make sure that the batteries are installed correctly. Otherwise, the thermometer may be damaged. If the low-battery symbol is displayed on the screen, replace the batteries. Batteries of a same type should be used. Dispose the used batteries in accordance with the local environmental policies. The thermometer is shipped with batteries.
  • Page 27 soft cloth. Clean the lens of the temperature probe with a cotton swab. (3) Wipe the thermometer body with a slightly damp soft cloth. Keep water out off the lens during the cleaning process. Otherwise, the lens may be damaged. The lens may be scratched if it is cleaned with a hard object, which might result in inaccurate readings.
  • Page 28: Maintenance

    of repetitions per disinfection is not more than 2 times. Clean and disinfect the thermometer under the temperature of +10℃~+40℃(50℉-104℉), the relative humidity of 15%~85%RH (no condensation) and the barometric pressure of 86kPa~106kPa. Maintenance Preventive inspection & maintenance period Ensure the safety of thermometer, and check whether it has potential safety hazards in normal use each week, e.g.
  • Page 29: Troubleshooting

    Troubleshooting Problem Possible Cause Solution Low battery Change the batteries. Polarities of the Make sure that the The thermometer batteries are reversed. batteries are installed fails to power on. correctly. The thermometer is Contact the damaged. manufacturer. The ambient Take a measurement temperature is lower under an ambient "Er1"...
  • Page 30: Specifications

    Specifications Product Name Infrared Thermometer Product Model JPD-FR403 Power Supply Mode Internal power supply Operating Voltage DC 3V Battery Model AAA x 2 Operating Mode Continuous operating Display Segment LCD Measure time About 1 second Latency Time About 3 seconds Forehead mode: 22.0°C–43.0°C (71.6°F–109.4°F)
  • Page 31: Security Class

    Outer dimensions 156.5*38.6*43.4mm (mm) Thermometer (with batteries): 90g Weight (g) Manufacturing date see the label Service life 2 years Alkaline dry battery for around 20000 Battery life measurements Temperature: 10°C~ 40°C (50°F–104°F) Operating Humidity: 15%–95% RH, non-condensing Environment Atmospheric pressure: 86–106 kPa The infrared thermometer has been tested and conforms to the standard ASTM E1965-98.
  • Page 32: Storage And Transportation

    EMC information provided for in the ACCOMPANYING DOCUMENTS. ● Portable and mobile RF communications equipment can affect Infrared Thermometer JPD-FR403. ● The Infrared Thermometer JPD-FR403 should not be used adjacent to or stacked with other equipment. Guidance and manufacturer´s declaration – Electromagnetic emission...
  • Page 33 Guidance and manufacturer´s declaration – Electromagnetic emission The Infrared Thermometer JPD-FR403 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer JPD-FR403 should assure that it is used in such an environment. Emissions...
  • Page 34 Guidance and manufacturer´s declaration – Electromagnetic immunity The Infrared Thermometer JPD-FR403 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer JPD-FR403 should assure that it is used in such an environment. Immunity...
  • Page 35 transmitter. Recommended separation distance where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed transmitters, determined by an Electromagnetic site survey,a should be less than...
  • Page 36 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the JPD-FR403 is used exceeds the applicable RF compliance level above, the JPD-FR403 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the JPD-FR403.
  • Page 37 Thermometer JPD-FR403 as recommended below, according to the maximum output power of the communications equipment. Rated Separation distance according to frequency of maximum transmitter output power 80 MHz to 800 MHz 800 MHz to 2.5 GHz transmitter 0.01 0.12 0.23 0.38...
  • Page 38: Warranty And After-Sale Service

    2. Failure resulting from an unexpected dropping during application or transportation. 3. Failure resulting from not following the instructions in the user’s manual. After-sale service unit: Shenzhen Jumper Medical Equipment Co., Ltd. Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen,Guangdong,China Tel: +86-755-26696279 Fax: +86-755-26852025 E-mail:info@jumper-medical.com...
  • Page 39 Shenzhen Jumper Medical Equipment Co., Ltd. Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, China Tel:+86-755-26696279 Fax:+86-755-26852025 E-mail: info@jumper-medical.com Website: http://www.jumper-medical.com...

Table of Contents