Hi Testing - ReSound RT961-DRW Technical Manual

Cambridge rie c6
Table of Contents

Advertisement

HI Test - DSA6000
Please also see GTO info doc "GI-01-10-002 – Guide to "WL Only test" - to be found on GTO web
Add the correct bat pill size 312 or 13 and close the bat door (no bat pill for RHI models as they have integrated bat). Attach the test rec to the test jig. The
test rec mic is to be placed on the jig as shown (see also pics below). Connect the test rec to the HI
Mount the 2cc coupler to the TM12 coupl mic. Attach the test rec to the 2cc coupler. Attach the HI to the test jig/fixture in the
correct position. Attach the jig/fixt to the black test frame and place it all in the test chamber. Make sure to place the test rec mic
correct on the test jigs before testing. Select the correct test prog and make sure to follow instructions coming up on the screen
Size 13
& RHI
Not subject to issue control when printed
Doc 0371130 rev. F
https://gto.gnresound.com
Size 312
Page 21 of 22