Appendix A: Calibrating The Wavelength Of The Maya2000Pro Series Spectrometers; Overview; About Wavelength Calibration - Halma Ocean Optics Maya2000Pro Series Installation And Operation Manual

Spectrometers
Table of Contents

Advertisement

Calibrating the Wavelength of

Overview

This appendix describes how to calibrate the wavelength of your spectrometer. Though each spectrometer
is calibrated before it leaves Ocean Optics, the wavelength for all spectrometers will drift slightly as a
function of time and environmental conditions. Ocean Optics recommends periodically recalibrating the
Maya2000Pro Series.

About Wavelength Calibration

You are going to be solving the following equation, which shows that the relationship between pixel
number and wavelength is a third-order polynomial:
Where:
= the wavelength of pixel p
I = the wavelength of pixel 0
C
= the first coefficient (nm/pixel)
1
C
= the second coefficient (nm/pixel
2
C
= the third coefficient (nm/pixel
3
You will be calculating the value for I and the three Cs.
020-00000-000-02-201603b
the Maya2000Pro Series
= I + C
p + C
p
1
2
)
3
)
Appendix A
Spectrometers
2
3
p
+ C
p
2
3
15