Hazardous Substances; High Temperatures; Dimensioning The Fine-Wire Fuse - Ilmvac MPC 104 Tp Operation Manual

Diaphragm pump
Table of Contents

Advertisement

2.4

Hazardous Substances

The operating company bears the responsibility for the use of the diaphragm pump.
Hazardous substances in the gases to be pumped can cause personal injuries and property
damage. Pay attention to the warning notices for handling hazardous substances.
The local regulations apply in foreign countries.
Combustible Gases
Examine before switching on whether that can form gas combustible gas/air mixtures which
can be promoted! Consider the regulations of the guideline 1999/92/EC.
Explosive gases
The diaphragm pumps is not certified according to ATEX guidelines 94/9/EC.
Aggressive gases
Extracting aggressive gases is not permitted.
Poisonous gases
Use a separator when pumping poisonous or harmful gases. Prevent such substances from
leaking out of the appliance or pump. Treat these substances according to the applicable
environmental protection regulations.
Test the strength and leak-tightness of the connecting lines and the connected apparatus.
Prevent environmental poisons, e.g. mercury, getting into the diaphragm pumps.
Fulfil the requirements, for example:
German Hazardous Substances Regulation (GefStoffV) of 23. December 2004
Regulations 2006/121/EC (classification, packaging and identification of hazardous substances),
Manufacturer's safety data sheets on hazardous substances.
2.5

High Temperatures

The diaphragm pump may heat up as a result of the temperature of the gas being pumped
and through compression heat. Prevent the following maximum permissible temperatures
from being exceeded.
+ 40 °C for the environment, and
+ 60 °C for the gas to be pumped.
The motor for single phase alternating current is protected against overload by a fine-wire
fuse.
2.6

Dimensioning the fine-wire fuse

Diaphragm pump model
MPC 104 Tp - 230 V
MPC 104 Tp - 115/230 V
Open the base plate to change the fuse. First, pull out the power plug.
Basic Safety Instructions
Fine-wire fuse
T 1 A
T 3.15 A
400082
Order no.
825366
825391
CAUTION !
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

825366825391

Table of Contents