Maintenance and Servicing

6
Maintenance and Servicing
6.1

General Requirements

• Check the pump daily for unusual running noises and heat building up on the surface of
the pump.
• We recommend changing the diaphragm after 10,000 operating hours. The user may
specify that the exchange be made earlier, depending upon the application process.
• Check the electrical and vacuum connections daily.
6.2

Maintenance Performed by the User

Only perform the work that is described here, and that which is permitted to be done
by the user.
All other maintenance and service work may only be performed by the manufacturer
or a dealer authorized by him.
Beware of the pump parts being possibly contaminated by hazardous substances.
Wear protective clothing if there is contamination.
Scope of permissible work:
• Inspect the pump chambers, diaphragms and valves,
• Deposits in the inside of the pump must be cleaned out,
• Change the diaphragms, valves and seals.
Before opening the pump unplug it from the mains.
Tools required:
The tool kit, order No. 402107
consists of:
• 826801-4 Cross-head screwdriver, size 2,
• 826801-2 Open spanner, size SW 14,
• 826801-9 Open spanner, size SW 7.