Download Print this page

Hioki 9694 Instruction Manual page 2

Clamp on sensor

Advertisement

• Accurate measurement may be impossible in the presence of strong
magnetic fields, such as near transformers and high-current conductors, or
in the presence of strong electromagnetic fields such as near radio
transmitters.
• This product may cause interference if used in residential areas. Such use
must be avoided unless the user takes special measures to reduce
electromagnetic emissions to prevent interference to the reception of radio
and television broadcasts.
Specifications
Guaranteed accuracy period:1 year
Guaranteed accuracy period after adjustment
Conditions of guaran-
made by Hioki:1 year
teed accuracy
Opening and closing of the Jaw :10000 times or less
Accuracy guarantee for temperature and humidity
:23 °C±5° C (73 °F±9 °F), 80% RH or less
Rated primary current 5 A AC
Output rate
10 mV AC/A
±0.3% rdg.±0.02% f.s.
Amplitude accuracy
(f.s.:5 A, 45 Hz - 66 Hz, at jaws center)
2
Amplitude
characteristics
1.5
1
0.5
0
-0.5
-1
-1.5
-2
0.01
Phase accuracy
Within ±2° (45 Hz - 5 kHz)
Amplitude frequency
Within ±1% at 40 Hz - 5 kHz (deviation from accuracy)
characteristics
Effect of conductor
Within ±0.5% (deviation from center)
position
Effect of external
0.1 A equivalent or lower
electromagnetic field
(in an AC electromagnetic field of 400 A/m)
Maximum input current 50 A continuous (at 45 - 66 Hz, ambient temperature 50°C)
Temperature coefficient 0.02%rdg./°C
4290 V rms 50 Hz/60 Hz,1minute,
Dielectric strength
(jaw - hand heldportion,jaw - outputconnector)
Maximum rated voltage
300 V AC rms or lower
to earth
Operating Temperature
0 to 50°C (32 - 122°F), 80%RH or lower
&Humidity
(non-condensating)
Storage Temperature
-10 to 60°C(14 - 140°F), 80%RH or lower
&Humidity
(non-condensating)
Operating Environment Indoors, altitude up to 2000 m (6562-ft.)
Safety
Applicable Standards
EMC
Measurable conductor
φ15 mm (0.59") or less
diameter
Cable length
Approx. 3 m (118.11")
Approx. 46W x 135H x 21D mm (1.81W" x 5.31H" x
Dimensions
0.83D") (excluding protrusions)
Mass
Approx. 230 g (8.1 oz.)
Accessory
Instruction Manual
f.s. : maximum display value or scale length
rdg.: reading value (The value currently being measured and indicated on the
measuring product)
0.1
1
10
Input current AC (A)
入力電流AC(A)
EN61010
Measurement Category III, Pollution Degree 2
(Anticipated Transient Overvoltage: 4000 V)
EN61326 (Class A)
Parts Names
The 9694 is a voltage-output-type clamp-on sensor.
BNC connector
Ferrite core
Cable
Measurement Procedures
• When disconnecting the BNC connector, be sure to release the
100
lock before pulling off the connector. Forcibly pulling the connector
周波数50Hz
without releasing the lock, or pulling on the cable, can damage the
connector.
• To prevent damage to the connected instruments, never connect
or disconnect a sensor while the power is on.
Attach the clamp around only one conductor. Single-phase (2-wire) or
three-phase (3-wire) cables clamped together will not produce any
reading.
OK
Connect the BNC connector.
BNC connector grooves
Connector
guide
Clamp the conductor.
Position the clamp with the current
direction indicator pointing toward the
load side. (If installed in the opposite
direction,the phase deviates 180
degrees.)
jaws
Lever
1. Unlock the lever, if locked.
2. Engage the BNC connector
grooves with the connector-
Lock
guide projections, and turn
the connector clockwise to
lock the components.
3. Pull the lever toward you
while pressing downward,
and open the jaws.
Electric
4. Hold only one conductor at
conductor
the jaws center with the cur-
rent direction indictor point-
ing toward the load side.
5. Make sure the jaws are
closed.
Current direction
6. Slide the lock knob to lock
indicator
the jaws ("LOCK" is dis-
played).
To remove the BNC connec-
tor, turn the connector counter-
clockwise and pull it out.
Current direc-
tion indicator
Lock knob

Advertisement

loading