STATIM
Operator's Manual should be consulted before using this autoclave.

BASIC OPERATION

1
Power unit ON.
3
Load configurations.
4
Close unit.
6
Open unit.

TOUCHSCREEN

Solid / Unwrapped:
IUSS* cycle for
unwrapped solid
instruments
Solid / Wrapped:
Wrapped Solid,
non-lumened
instruments
Universal:
Wrapped and
unwrapped solid and
hollow instruments,
cassettes, textiles
Drawer
lock status
*Immediate Use Steam Sterilization

MAINTENANCE SCHEDULE

Wipe the drawer seal.
Daily
Clean external surfaces.
Clean the chamber and trays.
Drain water reservoir.
Weekly
Disinfect external surfaces.
Inspect and clean the chamber filter
(or replace if necessary).
Every
2 Years
A complete maintenance of the autoclave including testing of the pressure relief valve and the power failure mode
(by a SciCan-approved technician) is recommended.
or
2,000 Cycles
www.scican.com
95-116554 EU EN R1. STATIM is a trademark of SciCan Ltd. A Coltene Group Company. Manufactured by: SciCan Ltd. 1440 Don Mills Road, Toronto ON M3B 3P9 / CANADA
B G4
+
®
Quick Reference Guide
2
Fill with demineralised/distilled water.
5
Run cycle.
A
Select a cycle.
C
Allow cycle to run.
7
Unload instruments.
Cycle records
Connectivity
Settings
Monthly
or
160 Cycles
Every
6 Months
or
1,000 Cycles
NEVER USE
TAP WATER
B
Start a cycle.
D
End of cycle.
Rubber & Plastic:
Unwrapped and wrapped
solid or hollow instruments
constructed of metal, rubber
and plastic
Helix/BD Test
Vacuum Test
Custom:
For devices that require different
cycle parameters than those
available. Set according to
the instrument manufacturer's
instructions.
Inspect the water reservoir filter
(clean or replace).
Clean the external fill water tank.
Perform all monthly cleaning tasks.
Replace bacteriological air filter.
Replace drawer seal.
+49 (0) 7561 98 343 - 0

WARM-UP MODE

STAND-BY LOW
For low to average use: 70°C / 158°F
STAND-BY HIGH
For high use: 120°C / 248°F
STAND-BY OFF (default) – For infrequent use. The wait time
will be longer (up to 12 minutes from a cold start)

INSTRUCTIONAL VIDEOS

LED COLOUR

WHITE
Unit is idle;
ready for
operation
BLUE
Cycle
complete

LOCK ICON

Drawer is locked;
the chamber is still
under pressure
Drawer is safe to
open

Spare Parts and Accessories

01-116292S
Air Filter
01-116293S
Drawer Seal, 6L
01-116294S
Tray, 6L
01-116296S
Exhaust Tube
01-116297S
Extended Exhaust Tube
01-116298S
Direct-to-Drain
01-116299S
Water Filling
01-116300S
Waste Bottle
01-116301S
USB Memory Stick
01-116302S
Water Reservoir Cap
SCI-BDSK134V
Bowie-Dick Test Kit
97902001
Helix Test Kit with 100 Chemical Indicators
97902002
Helix Text Kit with 400 Chemical Indicators
01-110282S
Power Cord EU 16A/250V
01-110285S
Power Cord UK IND plug 16A/250V
01-116496
StatMat Silicone Mat STATIM B
01-116497S
Reservoir Chamber Filter
01-116498S
Pouch Rack Kit
IMPORTANT!
Users of this device must be sure to select the most appropriate
sterilisation cycle according to the recommendations of the
leading infection control authorities and local regulatory
guidelines / recommendations.
RED
Cycle in
operation
ORANGE
Cycle
stopped or
cycle fault
condition
Drawer remains
locked;
not safe to open
Drawer remains
unlocked;
unsafe to start cycle

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SciCan STATIM B G4+ and is the answer not in the manual?

Questions and answers

Summary of Contents for Coltene SciCan STATIM B G4+

  • Page 1 (by a SciCan-approved technician) is recommended. guidelines / recommendations. 2,000 Cycles www.scican.com 95-116554 EU EN R1. STATIM is a trademark of SciCan Ltd. A Coltene Group Company. Manufactured by: SciCan Ltd. 1440 Don Mills Road, Toronto ON M3B 3P9 / CANADA...

Table of Contents