Chapter 7 Pre-Installation Conditions Check List - Philips CSM 66 III Preinstallation Manual

Table of Contents

Advertisement

CHAPTER 7 7 Pre-installation
CHAPTER
SUBJECT
MINIMUM DOOR OPENING
- width
- height
MINIMUM CEILING HEIGHT
MINIMUM FLOOR SPACE
FLOOR LOADING
- average
- per foot
LIGHTING
ENVIRONMENTAL CONDITIONS
- temperature
- relative humidity
- ambient air pressure
- dust class
COMPRESSED AIR SUPPLY
- pressure
- flow
- oil content
- dew point
- dust particles size
5322 871 63303
96.00
Pre-installation Conditions
Pre-installation
Conditions Check
Pre-installation Conditions
REQUIREMENTS
1600 mm
1810 mm
2200 mm
3600 x 4140 mm
4
2
> 55.6 10
N/m
4
2
> 13.9 10
N/m
+20 to +28 ˚C
80%
800 to 1100 mbar
100,000
> 5 x 10^5 Pa (abs)
650 Nl/minute
≤ 0.1 mg/m
3
5
< 2˚C at 7 x 10
Pa
5
< 4˚C at 7.2 x 10
< 10 µm
Pre-installation Manual CSM 66/84/84V III
PA 1306/40 /50 /60 /61
Check List
Conditions Check
Check List
ACTUAL (MEASURED)
VALUE
Pa
List
List
REMA-
RKS
Sect. 2.3
Sect. 2.4
Fig. 2.1
Sect. 2.4
Sect. 2.1.2
Sect. 3.1
Sect. 4.1
21