NOTE: The pressure specified must remain within
NOTE: A compressed air supply that does not meet the
FIGURE 4.1:
14
Services Required
Services
4.1.1 Specification of Compressed Air Supply
Pressure at system
Flow
Oil content
Dew point
Dust particles size
Connection to system
specification at maximum flow.
above specifications will increase the frequency of
carrying out the following maintenance actions:
Cleaning of the vacuumsystem and the micro-ejectors (venturi's) if the oil
content exceeds the permissible level.
Draining the water from the mist filter (MF600-04-A) if the water vapour
content exceeds the permissible level.
Cleaning of the air filter (F600-04-BG-A) if the dust content exceeds the
permissible level.
In such cases, the compressed air supplier must take measures by installing gas
filters, etc. as required.
Connections.
Pre-installation Manual CSM 66/84/84V III
PA 1306/40 /50 /60 /61
Required
5
> 5 x 10
Pa (abs)
40 Nl/minute
≤ 0.1 mg/m
3
5
< 2°C at 7.0 x 10
5
< 4°C at 7.2 x 10
< 10µm
via 20 mm diameter coupling, (see fig. 4.1)
Pa
Pa
5322 871 63303
96.00