2.5 Connections
1
Probe
2
LAN
3
USB B
4
USB A
5
HDMI
6
In 24 V
7
Trigger in / out
8
Pat. Resp.
9
Right
10
Left
11
Contra
2.6 Calibration cavities
You can use the 0.2 ml, 0.5 ml, 2.0ml and 5 ml for validity check of the probe calibration.
To perform a calibration check, select a protocol the measures a tympanogram.
Do not use an ear tip! Place the probe tip completely into the cavity. Perform the measurement. Check the
volume that was measured.
The allowed tolerance in the volume measurement is ±0.1 ml for cavities up to 2 ml and ±5% for larger
cavities. These tolerances are applicable for all probe tone frequencies.
We strongly recommend calibrating probe and contra phone at least once a year.
D-0133834-A – 2022/09
AT235 - Instructions for Use - EN
Dedicated probe connection.
LAN (Not used).
For printer, mouse, keyboard, Memory stick .
For PC connection.
For external monitor or projector.
Use only specified power supply unit UES65-240250SPA3
type.
Cochlear implant trigger option.
Patient response button.
Audiometry Right output.
Audiometry Left output.
Contra transducer output.
Page 8