Download Print this page
Niigata seiki GDCS-150DR Instruction Manual

Niigata seiki GDCS-150DR Instruction Manual

Digital depth gauge

Advertisement

Quick Links

INSTRUCTION MANUAL
Precision Measuring Tool For Depth Measurement
DIGITAL DEPTH GAUGE ( One Side Base )
Thank you for purchasing Niigata Seiki Digital Depth Gauge ( One Side Base ) .
This product is a digital instrument for measuring depth.
●For safe and proper use of this product, please read this instruction manual before use and follow the procedures described.
●Keep this manual with the instrument if transferred or leased to a third party.
●For inquiries about this product, please contact distributor or place of purchase.
■PARTS IDENTIFICATION & FUNCTION
④Lock Screw
③LCD Display
②Depth Meas.Surface
①Base
①Base
.....................
Reference surface placed on face of
workpiece.
②Depth Meas.Surface
...
Insert into groove to measure depth.
③LCD Display
............
Displays digital measurement.
④Lock Screw
............
Tighten to prevent the slide and display
from moving.
⑤ [ HOLD ] Button
.........
When pressed, "HOLD"Icon appears on
Display and measurement will be fixed.
SAFETY PRECAUTIONS
To prevent harm to yourself or others, and to prevent damage to
property, always follow the procedures marked by the following
symbols.
WARNING:
Indicates risk of personal injury or property damage if not followed.
Read the manual and follow the directions.
・Use of product other than as described in this manual
may cause accident.
Use only for measuring.
・Use for any purpose other than measuring may  
damage or wear the instrument. Improper use may
also cause accident.
Handle with Care.
・Do not drop or subject instrument to excessive shock.
 Do not place under heavy objects. Improper handling
may cause damage or poor accuracy.
・Do not scratch instrument, for example by writing ID number.
Avoid using in high electric fields, such as near
fluorescent lights or switching power supplies.
・Incorrect value may be displayed on LCD.
PREPARATION FOR USE
●Wipe off oil, rust proofing, and dirt from measuring
surfaces and slide.
Contamination of surfaces may cause measurement error.
Model No.
⑤ [ HOLD ] Button
⑥Output Connector
⑩Scale
mm
⑦Battery Cover
⑧ [ ZERO ] Button
⑨ [ ON/OFF ] Button
⑥Output Connector
...
Connect to optional Bluetooth Box to
send measurements to PC.
⑦Battery Cover
.........
Open to access battery.
⑧ [ ZERO ] Button
.........
When pressed, Display will read "0.00"
⑨ [ ON/OFF ] Button
......
For turning power ON / OFF.
⑩Scale
.....................
Supports LCD Display Unit.
Please Observe
Denotes a requirement
-You MUST do
Use and store only in environment meeting the
following conditions:
●Dry location protected from rain, water, and oils.
●Location that is not excessively hot.
●Location protected from use by children and
unauthorized people.
・Use in places contrary to the above may cause  
damage to the product resulting in accident or injury.
Do not disassemble or modify.
・Disassembly may damage product or cause accident.
For repair, please contact distributor or place of
purchase.
Do not write on instrument with electric engraver.
・Do not engrave instrument body or jaw, such as an
ID number with engraver or electric pen.
Before using, please follow the steps below
●Allow the gauge and object to be measured time to reach
the same temperature.
A temperature difference between the object to be measured
and the gauge may cause measurement error. Allow time for
the temperatures to equilibrate.
INSERTING・REPLACING BATTERY
GDCS-150DR
①Slide cover in direction of arrow.
②Insert battery with (+) side outward.
③Replace battery cover.
Notice
Use only CR2032 type
battery.
●Use care to avoid damaging
terminals when inserting
battery.
●Terminals may be broken if battery is pressed straight down.
●Insert battery at angle, underneath the small hook, as shown.
Terminals
TROUBLESHOOTING
Q. Nothing is displayed on LCD.
A. The battery may be running low, please replace battery
with a new one.
Q. With power on, the number on the LCD does not change
as Display Unit is moved.
A. Please reset by removing battery for about 30sec., and
then replacing.
Q. Nothing happens when buttons are pressed.
A. Make sure battery is properly inserted.
Q. The characters "MIN", "MAX", or "△" are displayed on LCD.
A. Cycle power, or apply several long presses to [ HOLD ]
Button. Use as normal once character is off.
Q. The correct value is not displayed.
A. High humidity may cause dampness. Please dry off the
Denotes a prohibition
instrument and use only in dry location.
-You MUST NOT do
SERVICE
●If Instrument is not working properly, or if you have any
questions, please contact distributor or place of purchase.
●Please note, manufacturer is unable to respond to inquires or
provide service directly. Please contact distributor or place of
purchase.
SPECIFICATIONS
●Measurement Range
●Resolution
........................
● Maximum Permissible Error
●Display
...........................
●Auto Off
...........................
●Maximum response speed
●Operating Temp. Range
●Storage Temp. Range
●Power
...........................
●Battery Life
.....................
●Weight
...........................
●Accessory
........................
HOW TO USE
■Power ON/OFF
①To turn on, press the [ ON/OFF ] Button.
②To turn off, press the [ ON/OFF ] Button.
③With power on, press the [ HOLD ] Button to fix the value on the
display.
■Absolute Measurement
①Turn on power, and loosen the Lock Screw.
Battery
②Place the Base on a flat surface, and push the Scale down.
Cover
When the Measurement Surface contacts the surface, press
the [ ZERO ] Button. ( Fig.1 )
The Display will read "0.00mm" and the gauge is ready for
measurements.
③When placed on workpiece with Measurement Surface
lowered into groove, the depth of the groove will be displayed.
( Fig.2 )
( Fig.1 )
Battery
Hook
Terminals
Battery
■Comparative Measurement
①Turn on power, and loosen the Lock Screw.
②Place the Base on the Reference Part and push the Scale
down. When the Measurement Surface is in contact with the
bottom of the groove, press the [ ZERO ] Button. ( Fig.3 )
The Display will read "0.00mm" and the gauge is ready for
measurements.
③When placed on workpiece with Measurement Surface
lowered into groove, the difference the difference in depth
compared to the Reference Part will be displayed. ( Fig.4 )
( Fig.3 )
AFTER USE CARE ・ STORAGE
●After use, wipe off any dirt or oil from measuring
.........
surfaces, sliding surfaces, and exterior using a dry cloth
0-150mm
and perform rust prevention treatment.
0.01mm
For rust prevention, lightly apply anti-rust oil, or place in rust
...
±0.03mm
preventive bag.
※Not including quantization error ( ±1 count )
●Store with Lock Screw loosened.
5 Digit LCD
If stored with Lock Screw tight, expansion due to temperature
approx. 20 min
changes may cause excessive force which will cause inaccuracy.
●Store in a cool, dry, and dark location in the provided case.
...
Slide in pos. dir....1.5m/s
Slide in neg. dir....0.75m/s
Keep out of direct sunlight and moisture, and please keep
......
  0° C〜40° C
secure from unauthorized personnel.
●Remove battery if Instrument will not be used for a long
.........
−10° C〜60° C
period of time.
( 1x ) CR2032 Lithium Battery
Batteries may leak if left in unused instrument for an
approx. 2.5 years in normal use
extended period.
175g
CR2032 Lithium Battery ( for test )
( Fig.2 )
Scale
Scale
[ ZERO ]
Button
Depth Meas.
Surface
Base
Workpiece
Workpiece
Flat Surface
( Fig.4 )
Difference in
Depth
compared to
[ ZERO ]
Reference Part
Button
Reference Part
Reference Part
Workpiece
Workpiece
Niigata seiki Co., Ltd.
5-3-14, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-33-5522 Fax. : +81-256-33-5518
MAIL intl.sales@niigataseiki.co.jp
URL
http://www.niigataseiki.co.jp

Advertisement

loading

Summary of Contents for Niigata seiki GDCS-150DR

  • Page 1 ③Replace battery cover. ③With power on, press the [ HOLD ] Button to fix the value on the display. Thank you for purchasing Niigata Seiki Digital Depth Gauge ( One Side Base ) . ■Absolute Measurement Notice Use only CR2032 type This product is a digital instrument for measuring depth.
  • Page 2 電池のセッ ト ・ 交換 使用方法 取扱説明書 GDCS-150DR 品番 デジタルデプスゲージ (片側ベース) ■電源のON/OFF 深さ測定用 ①電池蓋を矢印の向きに開けます。 精密測定工具 ②電池を、必ず (+) 側を表にしてセッ ト ①電源をONにするには、 “ON/OFF”スイッチを押してください。 してください。 ②電源をOFFにするには、 “ON/OFF”スイッチを押してください。 この度は「デジタルデプスゲージ (片側ベース) 」をお買上げいただきありがとうございます。 ③電池蓋を元通りに装着してください。 ③電源ONの状態から“HOLD”スイッチを押すと数値が固定されます。 この商品は、深さの測定を行うデジタル測定器です。 電池は必ずCR2032をお使 ご注意 ■絶対測定 ●正しく安全にお使いいただくため、ご使用の前に本取扱説明書を必ず読み、記載の手順に従ってご使用ください。 いください。 電池蓋 お読みになった後は、お使いになる方がいつでも見ることができる場所に大切に保管してください。 ①止めねじをゆるめ、 電源を入れてください。 ●電池をセットする際は、 端子の...