Modification of the actuator is not al- lowed in any way. Any modification will void the warranty and could create a safety hazard. Ewellix disclaims any and all liability or responsibility for the modified product and for...
Ewellix product. in medical and care sectors. We reserve the right to make changes which are in the inter- Other uses must be discussed with Ewellix Liestal. est of technical progress. 2.3 Ambient conditions Please read the Technical Instructions carefully and, above Operation: all, pay careful attention to the Safety Instructions.
Ec o m a g 3.0 Function ECOMAG linear actuators are DC actuators with worm gears and spindle units. The gears convert the motor's rotational movement into a linear movement through the threaded spindle's rotation being transferred to a pushing tube via a special nut.
Risk of accident! If there is a risk of hitting a mechanical limit stop during a stroke movement, Ewellix recommends that mechanical an- ti-pinching protection or a current cutoff device be used. When the rapid adjustment option is used, it must be en- sured that the entire stroke movement can be completed.
Ec o m a g 4.2.4 Fastening the control unit 4.3 Startup externally 4.3.1 Opening the control box cover The control unit must be installed at the installation slots provided (⮑ Fig. 4), and must be installed so that it is free Gently lift the nose with a screwdriver, and at the same time from mechanical forces and vibrations.
4 .0 I n s t a l l a ti o n a n d s t a r t u p 4.3.3 Connecting / disconnecting the 4.3.4 Closing the control box cover control device • Move the control box cover in the direction of the arrow (⮑...
Ec o m a g 5.0 Operation 5.1 Controlling an actuator 5.2 Locking device The EHE handswitch is used to actuate one or more As an option, it is also possible to lock the individual actua- ECOMAG linear actuators. A control panel is provided on tors on the EHE handswitch.
"EMERGENCY lowering" function is used. Damaged, spent batteries must be replaced anyway. WARNING Batteries must be disposed of correctly or returned to Ewellix. They must not be disposed of in the general household waste! WARNING Protection class IPX4 is only guaranteed if the marking on the cover...
Ec o m a g 5.4.2 Usage • Open the red clamping lever (⮑ Fig. 16, (1)). While doing so, hold the head of the bed firmly. Depending on the bed design, the head or foot can lower quickly. Fig. 16 •...
The user must undertake to monitor the products vigilantly. The actuator may be cleaned with a damp cloth and water. The actuators are greased with Ewellix-approved lubricants, Add a little isopropyl alcohol if necessary and do not require any maintenance for the following service...
Service Department or if the unit has not been operated in accordance with the specified application. Ewellix shall not be liable for any damage resulting from fail- ure to observe these instructions. These instructions shall not be regarded as an extension of the warranty and liability terms set out in the Conditions of Sale and Supply applied by Ewellix.
Return the control unit to the Ewellix Service Department Control device faulty Replace the control device Poor connector contact Fit the plug correctly. Control device not working Poor connector contact Fit the plug correctly If you are unable to eliminate a fault, contact the Ewellix Service Department.
Ec o m a g 7.0 Appendix Technical data Please refer to the following document: Ecomag datasheet PUB IL-07004-EN-October 2019 For further technical information please contact Ewellix...
Page 17
A p p e n d i x Ecomag Linear actuator Benefits • Compact • Cost effective • Silent operation • Control unit can be mounted to the actuator Technical data Unit ECO 20/40 ECO 60/80 ECO 30/50 ECO 70/90 Rated push load 2 000 6 000...
Page 18
Ec o m a g Dimensional drawing Ecomag 2-3-6-7 Ø30 +0,7 +0,5 +0,7 +0,5 Ø30 L = S+160 Ecomag 4 -5-8-9 Label position Label position Ø30 +0,7 +0,7 +0,5 +0,5 Ø30 22,4 L = S+160 83,4 Front attachment Rear attachment Rear attachment position push and pull version push and pull version...
Page 19
A p p e n d i x Performance diagrams Speed-load diagram Current-load diagram Speed [mm/s] Current consumption [A] 2 000 4 000 6 000 8 000 2 000 4 000 6 000 8 000 Load [N] Load [N] ECO 20 / 30 / 40 / 50 ECO 60 / 70 / 80 / 90 Safety factor load conditions Push load reduction safety factor = 4...
Page 20
Back-up nut (available only for drives with 6 kN, mandatory for IEC 60601-1) – only for ECO 7/9, L = +23 mm Customization Options shown in yellow are only available on request. Contact Ewellix for more information on minimum quantities and additional costs.