C700 | C720 Series Operating Manual
Press
[MENU] to bring you back to the measurement screen, and press the
view the
TREND
chart of the saved data, 400 sets of data records per screen.
Under the current mode, press the
to accelerate the displacement.
When the bottom icon
direction, and press to switch the display range (enlarge/shrink).
Item
Display
Terminal Current
Tuning
System
Relay Test
No Conductivity
Module Required
Sensor 1
Data Logging Trend Chart
2021 -11-09 12:53:17 1200 µS/cm 25.0 ° C
20
16
12
8
4
0
50
Conductivity / TDS / Salinity / Resistivity
[ENT] key to display the left and right displacement data line (red), and long press the
[ENT] is in green, it is the movement direction of the key. Press
Selection
Current 1: 4mA
Current 1: 20mA
Current 2: 4mA
Current 2: 20mA
Relay 1 — Testing
Relay 2 — Testing
4-20mA + RS 485
Power
(100-230 VAC)
Sensor 2
(with C720 Model)
[TREND] button in the measurement mode to directly
2023-01-09 12:53:17
10 0
15 0
200
250
300
350
R
2023-01-09
12:53:17
83.01
µS/cm
25.0
°C
SP1
SP2
SP3
C1
14.51 mA
Conductivity / TDS / Salinity / Resistivity
M
C
E
E
ENU
AL
SC
NT
I
I
NFO
NFO
400
to switch the displacement
Sub Menu
(The positive and negative ends of the
multi-meter are connected to the current
1 or current 2 output terminals of the
instrument key respectively, press
key to adjust the current 4mA or 20mA ,
press
[ENT] to confirm.)
Press
[ENT] to test
Relay
Alarm/Cleaning
Relay 2
Pump/Valve
Relay 1
Pump/Valve
◀ ◀
13
|
[ENT] key