Field Calibration; Calibration - ICON ProCon C700 Series Operating Manual

Table of Contents

Advertisement

C700 | C720 Series Operating Manual
Calibration point 1: After entering the calibration mode, the current measurement value will be displayed. This is known as the standard
solution value.
Place the electrode into the standard solution of the corresponding value, and the corresponding voltage mV value and calibration state
will be displayed on the screen.
After completion of calibration,
If you want to calibrate the next point, press the
If only one point calibration is required, press the
During the calibration process, the calibration status will be displayed on the right side of the screen.
Calibrating
indicates in calibration, and
Field Calibration — Select on-site calibration methods:

Field Calibration

When the data from laboratory or portable instrument are input
into this item, the instrument will automatically correct the data.
Offset adjustment
Compare the data of laboratory or portable instrument with the data of instrument measurement,if there are errors, the error data can be
modified by this function.
Linear adjustment
Linear values after "field calibration" are stored in this item, with factory data of 1.00. The main value of the instrument displays the known
standard liquid value point.
Place the electrode into the standard solution of the corresponding value, and the corresponding voltage mV value and calibration state
will be displayed on the left side of the screen.
After completion of calibration,
the calibration point.
If only one point calibration is needed, after the calibration is completed, press
standard solution is wrong, the screen will show an Error.

Calibration

Press
[MENU] to enter the setting mode,set the recording interval, and the instrument will
Graphic Trend
(Trend Chart)
Data Log
Data Query
Interval Enter
Frequency of Data
Collection
Done
will be displayed on the right side of the screen.
key to switch the calibration point.
[MENU] key to return to the superior menu after calibration.
Error
indicates a sensor or a calibration issue.
[Linear calibration]
Done
will be displayed on the right side of the screen. If the next point is calibrated, press
Interval/point
1h/point
12h/point
24h/point
Year/Month/Day
7.5s
90s
|
1.5 minutes
180s
|
3 minutes
,
[Offset adjustment]
Conductivity / TDS / Salinity / Resistivity
Field Calibration
pH / ORP
µS/cm: 07.00
Vol:
T emp: 25.0 °C
Calibrating
[MENU] to exit. During the calibration process, when the
400 points per screen,displays the most recent data trend graph
according to interval Settings
400 points per screen, display trend chart of the last 16 days of data
400 points per screen, display trend chart of the last 200 days of data
400 points per screen, display trend chart of the last 400 days of data
Year/month/day time: minute/second value unit
Store data every 7.5 seconds
Store data every 90 seconds
Store data every 180 seconds
Done
indicates successful calibration,
,
[linear
adjustment].
Slope: 00.984
-002
Done
|
1.5 Minutes
|
3 Minutes
◀ ◀
12
|
to switch