Factory Default Current Outputs; Calibration - ICON ProCon C700 Series Operating Manual

Table of Contents

Advertisement

C700 | C720 Series Operating Manual
Factory Defaul t Current Outputs

Calibration

Press
[MENU] to enter the calibration menu and select
Standard Calibration
Calibration
Field Calibration
Calibration of Standard Solution
The sensor has been factory calibrated before delivery and users can use it directly.
Note: Calibration if required can be made using at least one of the five calibration points. Not all points are
required for sensor calibration
If calibration is required, prepare a suitable standard reference liquid with known range value. Press the
Setting Mode — select the calibration point.
Enter the corresponding reference calibration value, then press
(If the instrument has been calibrated, the screen shows the calibration status. press the
calibration point. If re-calibration is required in this mode, press the
In the standard solution calibration mode, press the
If prompted to enter the calibration security password, press the
then press the
[ENT] key to confirm the calibration security password.
Calibration
Cell
Models
Cell 0.01
C250
Cell 0.01
C250
Cell 0.1
C350
Cell 0.01
C450, C650, C950
Cell 0.1
C350
Cell 0.1
C450, C650, C950
Cell 0.1
C350
Cell 1.0
C450, C550, C650, C950
Cell 10
C450, C650, C950
Cell 0.4
C554, C850
Toroidal
C750
Calibration Point 1
Calibration Point 2
Calibration Point 3
Calibration Point 4
Calibration Point 5
Field Calibration
Offset Adjustment
Slope Adjustment
Conductivity / TDS / Salinity / Resistivity
P oint 1
Calibrating
Range
0 – 2µS
0 – 10µS
0 – 20µS
0 – 100µS
0 – 200µS
0 – 1,000µS
0 – 2,000µS
0 – 10,000µS
0 – 200,000µS
0 – 500,000µS
0 – 2,000mS
Enter given standard liquid value: Defa
Enter given standard liquid value (Example: 84μS/cm)
Enter given standard liquid value (Example: 1413μS/cm)
Enter given standard liquid value (Example: 12.88mS/cm)
Enter given standard liquid value (Example: 111mS/cm)
Enter Value
Enter Value
Enter Value
[ENT] key to confirm and enter the calibration screen.
[ENT] key again to enter the re-calibration.)
key to switch the calibration points, press the
key or
key to set the calibration security password, and
V oltage:
-5 mV
0
µS/cm
Done
◀ ◀
|
[MENU] button to enter the
key to switch the calibration state of the
[ENT] key to start calibrating.
Sensor Voltage
Successful Calibration
11