Index of refraction
range
Auto turn off
Memory for settings
and test results
Factory calibration
interval
Output wavelengths
Laser classification
Reflectance threshold
Loss threshold
Distance to the first
reflective open
Shortest distance to
an open APC
connection
Maximum distance
Distance accuracy
(0 m to 23,000 m or
0 ft to 7,549 ft)
Testing speed
Connector
Fiber types tested
1.45 to 1.5 (factory default is 1.468)
Automatically turns off after 5 minutes if
no keys are pressed. Backlight turns off
first.
Non-volatile memory. Saves the results
from up to 99 tests.
None
1550 nm ± 25 nm
Class 1 CDRH Complies to EN 60825-2
User-selectable from -24 dB to -62 dB
User-selectable from 0.5 dB to 6.1 dB
2 m
15 m
23 km (14.3 miles or 75,459 ft)
±1 m for reflective incidents ±1 % of
length
±3 m for non-reflective incidents ±1 % of
length
< 10 seconds typical
Removable/cleanable SC adapter, UPC
polish
9/125 µm singlemode
Specifications