Serial Port
Use this test to determine if the serial port is functioning properly.
The test sends data from the serial port, through a connector, and back into the
same port. The data sent and received should be identical.
1. Obtain a loopback connector, either by making your own or by ordering
from your local technical support provider.
The connector is a serial 9-pin male connector on which pins 2 and 3 are
connected and pins 4 and 6 are connected.
2. Select
Serial port
3. Connect the loopback connector into the serial port on the back of the
analyzer.
4. To begin the test, select the
The test continues until you exit the screen.
5. When finished, select
Touch Screen
Use this test to determine if the touch screen is functioning properly.
1. Select
Touch screen.
A screen displays that is filled with small boxes.
2. Select the center of each box.
As each box is selected, a check mark should display. It disappears when
the same box is selected again.
3. Select
Previous Screen
Barcode Reader
This test is identical to the Test bar code option described in Setup Menu 8.
If your handheld barcode reader is not reading your labels, you must determine
whether the problem is with the labels you are using or with the reader.
The package containing your barcode reader includes 2 sheets of barcoded
labels that are printed to the minimum specifications of the barcode reader. If
these labels cannot be read, the problem is probably with your reader. If they
read properly, the labels you are using may not be acceptable.
Clinitek Advantus Operator's Guide
to display the test screen.
Loopback
Previous Screen
key when finished.
key.
.
109