Page 1
BS-6023 Series Instruction Manual This manual is written for BS-6023 metallographic measurement device. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the microscope, it is strongly recommended that you read this manual carefully before operating the microscope.
BestScope International Limited User Notice BS-6023 1. Safety Symbols symbol explanation Indicates that the surface becomes hot, and should not be touched with bare hands. Before use, carefully read the user manual .Improper use could result in personal injury to the user and/or damage to the equipment.
BestScope International Limited BS-6023 3. Maintenance and Care 1. All the lenses have been adjusted properly; do not dismount them by yourself please. 2. The nosepiece and coarse and fine focusing parts are so delicate that it is forbidden to disassemble them carelessly by yourself.
BestScope International Limited 2.Assembly BS-6023 2-1 Assembly Diagram The following picture shows the sequence of the modules in which the number stand for the steps of assemble. Before assembling, make sure that every module is free of dusts and dirt. Do not scratch any modules or ...
BestScope International Limited 2-2Assembly Steps 2-2-1 Install L-plate (pic.1) Fix the L-plate and the fine-coarse movement component together by 4 matched bolts and washers. Pic.1 2-2-2 Install supporting component (pic.2) Insert the supporting component to the corresponding holes on the L-plate and...
Page 8
BestScope International Limited BS-6023 ○ 2-2-4 Install connecting plate (pic.4) Put the connecting plate on the L-plate and ○ fix it by 4 matched bolts. Make sure the locating ○ ○ surface 1 is adjacent with the surface...
Page 9
BestScope International Limited BS-6023 2-2-7 Install viewing head (pic.7) 1. Put trinocular viewing head onto the swallow-tail of the connecting-plate and ○ ○ screw bolt 1 to fix the viewing head. 2. Insert the eyepiece to the lens cone of the viewing head.
BestScope International Limited 3. Adjustment and Operation BS-6023 3-1Connect power source and regulate light (pic.1, pic.2) ○ 1. Connect the air plug of halogen lamp house with the aerial socket ○ of the power box, plug one head of power line to socket ○...
Page 11
BestScope International Limited BS-6023 ○ 3-3 Adjustment of coarse focus hand wheel (pic.4) The loose condition of coarse focus hand wheel ○ can be adjusted by turning the adjusting ring The coarse focus hand wheel is in a loose state when delivered, a heavy force is needed to rotate the hand wheel.
Page 12
BestScope International Limited BS-6023 3-6 3-gear light path switch (pic.7) ○ 1. When not using cameras, totally push in the light path switch rod and observing by eyepiece. Use the 10× objective to focus, first descend the focusing component and find the image in the 10×...
Page 13
BestScope International Limited BS-6023 3-7 Usage of color filter (pic.8, pic.9) Plug needed color filter in position1 (position 2 when use LED), make sure the filter is well settled. ○ Pic.8 ○ Color filters Function Ground glass Adjust the brightness of...
BestScope International Limited 6.Trouble Shooting BS-6023 6.1 Optical system Problem Cause Solution The nosepiece is not located in the Adjust it into the right position The edge of the field required position of view has shadow or The filament is not in center...
BestScope International Limited 6.2 Mechanical System Problem Cause Solution The sample moves not smoothly Slide is not tighten reliable Tighten it reliable The left and right fields of view is The interpupillary distance is Adjust interpupillary distance not coincided not correct...